The Rubber Sleeve Rebound Tester is designed to comply with the following standards:
(1) YY/T 1618-2018: Collection Needle Rubber Sleeve Resilience Test (China)
(2) DIN 53 512: Testing of Rubber – Rebound Resilience (Germany)
(3) ISO 4662: Rubber, Vulcanized or Thermoplastic – Determination of Rebound Resilience
(4) ASTM D7121: Standard Test Method for Rubber Compression Set
Features
(1) Accurate measurement of rubber sleeve resilience at multiple distances.
(2) Adjustable test speed and pressure for various sample requirements.
(3) Real-time display of test force and calculated results.
(4) Onboard printing for test reports, including average values and deviation.
(5) Supports testing of needles and collection tubes with minimal setup.
(6) Compact design suitable for laboratory and production environments.
Parameters
| Parameter | Specification |
|---|---|
| Pressure test range | 0 – 50 N, error ≤ ±1%, resolution 0.01 N |
| Test time | 1 – 999 s, error ≤ ±0.5 s |
| Test speed | 50 mm/min, 80 mm/min, 100 mm/min, error ≤ ±2% |
| Measuring distance | 0 – 50 mm, error ≤ ±0.1 mm |
| Rebound force range | 0.5 – 1.5 N |
| Printing capability | 8. 10. or 13 test data points, including average and deviation |
| Machine dimensions | Customizable depending on model and fixture |
| Power supply | AC220 V, 50 Hz (standard) |
Accessories
(1) Main test unit – 1 set
(2) Needle clamping fixtures – 1 set
(3) Printer paper roll – 1 piece
(4) Operation manual – 1 set
(5) USB storage (optional)
Test Procedures
(1) Insert the needle at the butt end into the die hole of the clamping fixture, 2 mm above the surface.
(2) Start the test device to compress the rubber sleeve.
(3) Measure the rubber sleeve resilience at distances of 5 mm and 15 mm between the needle tube and the needle base.
(4) The instrument displays and records the rebound force. Each test can print 8. 10. or 13 data points, including average and deviation.
(5) Confirm that the measured resilience falls within the standard range of 0.5 N ≤ resilience ≤ 1.5 N.
Maintenance Information
(1) Regularly clean the clamping fixtures and die holes to prevent debris buildup.
(2) Calibrate the force sensor periodically to maintain accuracy.
(3) Avoid prolonged exposure to high humidity or corrosive environments.
(4) Inspect electrical connections and printer components for wear or damage.
(5) Follow standard laboratory safety procedures when operating the device.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Automatic Nonwoven Liquid Loss Measurement System | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Nonwovens Liquid Loss Tester is a precision instrument designed to evalua... | |
| Catheter Flow Meter | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Catheter Flow Meter is specifically designed for measuring the water flow rate of medic... | |
| Aerosol Foams Flammability Tester,Automatic Reaction Calorimeter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction This product complies with "Recommendations on the Transport of Dange... | |
| Particle Extraction System for Cleanliness Analysis VDA 19 | ||
| Alternative to High temperature vacum friction & wear tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The working principle of a high-temperature vacuum friction and wear testing machine is to ... |
Похожие товары
| MMA-120 Handheld Arc Welder | Продавец: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... | |
| SHX Series Circulating Fluidized Bed Steam Boiler | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi... | |
| YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ... | |
| DHL Series Coal-Fired Steam Boilers | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b... | |
| SZL Series Biomass Steam Boiler | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi... |