Rubber Mechanical Properties Tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The Rubber Sleeve Rebound Tester is designed to comply with the following standards:
(1) YY/T 1618-2018: Collection Needle Rubber Sleeve Resilience Test (China)
(2) DIN 53 512: Testing of Rubber – Rebound Resilience (Germany)
(3) ISO 4662: Rubber, Vulcanized or Thermoplastic – Determination of Rebound Resilience
(4) ASTM D7121: Standard Test Method for Rubber Compression Set
Features
(1) Accurate measurement of rubber sleeve resilience at multiple distances.
(2) Adjustable test speed and pressure for various sample requirements.
(3) Real-time display of test force and calculated results.
(4) Onboard printing for test reports, including average values and deviation.
(5) Supports testing of needles and collection tubes with minimal setup.
(6) Compact design suitable for laboratory and production environments.
Parameters
| Parameter | Specification |
|---|---|
| Pressure test range | 0 – 50 N, error ≤ ±1%, resolution 0.01 N |
| Test time | 1 – 999 s, error ≤ ±0.5 s |
| Test speed | 50 mm/min, 80 mm/min, 100 mm/min, error ≤ ±2% |
| Measuring distance | 0 – 50 mm, error ≤ ±0.1 mm |
| Rebound force range | 0.5 – 1.5 N |
| Printing capability | 8. 10. or 13 test data points, including average and deviation |
| Machine dimensions | Customizable depending on model and fixture |
| Power supply | AC220 V, 50 Hz (standard) |
Accessories
(1) Main test unit – 1 set
(2) Needle clamping fixtures – 1 set
(3) Printer paper roll – 1 piece
(4) Operation manual – 1 set
(5) USB storage (optional)
Test Procedures
(1) Insert the needle at the butt end into the die hole of the clamping fixture, 2 mm above the surface.
(2) Start the test device to compress the rubber sleeve.
(3) Measure the rubber sleeve resilience at distances of 5 mm and 15 mm between the needle tube and the needle base.
(4) The instrument displays and records the rebound force. Each test can print 8. 10. or 13 data points, including average and deviation.
(5) Confirm that the measured resilience falls within the standard range of 0.5 N ≤ resilience ≤ 1.5 N.
Maintenance Information
(1) Regularly clean the clamping fixtures and die holes to prevent debris buildup.
(2) Calibrate the force sensor periodically to maintain accuracy.
(3) Avoid prolonged exposure to high humidity or corrosive environments.
(4) Inspect electrical connections and printer components for wear or damage.
(5) Follow standard laboratory safety procedures when operating the device.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Industrial large walk-in aging test chamber supplier | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com TheLarge Walk-In Aging Chamberis designed for long-duration temperature and h... | |
| Wheel Load Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard EN 1569: Determines rolling load resistance of synthetic sports surfaces (e.g., P... | |
| FPC Flex Durability meter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard ASTM F392 Product features 1. the stepper motor drive, high-precision positioni... | |
| ASTM D2061 Zipper Cycling Tester | Zipper Endurance Test Equipment Supplier | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Features 1. The counter adopts LCD display, 0~ 999.999 continuous adjustable... | |
| Alcohol Abrasion Testing Machine (for Alcohol abrasion tester) | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction: This machine adopts variable convex shaft structure, diameter Japane... |