Bend catheter flexibility equipment
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
ISO 10555 -Intravascular catheters- Sterile and single-use catheters
ISO 11o70- Catheter introducers, dilators, and guidewires
ASTM F2394-Guidewire mechanical property evaluation
ASTM F2908- Stiffness and flexibility of guidewires
ISO 178:2oo1 -Flexural properties of plastics (applicable to catheter materials)
Features
7-inch touchscreen interface with programmable test parameters and intuitive controls
Real-time force vs. displacement data acquisition with automated reporting
Dual-system control for independent force and displacement parameter setting
High-precision ball screw drive and guide mechanism for smooth operation
O.5% FS force sensor accuracy with resolution up to O.01 N
Supports catheter diameters up to 3.3 mm (customization available)
Advanced displacement control with O.1 mm accuracy
Pneumatic clamping system designed for fast, secure catheter mounting
Stores up to 1oo,ooo test records, supports audit trails and data export
Includes Rs232 interface and micro printer for traceable output
Technical Parameters
| Parameter | Specification |
|---|---|
| Testing Range (Force Range) | Usually 0 - 10 kilograms or more, depending on the tester's model and specifications |
| Testing Stroke | Varies in vertical & horizontal directions; e.g., vertical stroke: 0 - 300mm, horizontal stroke: 0 - 700mm |
| Testing Speed | - Press-down Speed: Adjustable within 0 - 300mm/min- Wire-drawing Speed (for applicable testers): 0 - 500mm/min or faster |
| Catheter Clamping Method | Usually adopts pneumatic or mechanical clamping, ensuring the stability and accuracy of the catheter during testing |
| Control System | Adopts PLC control system (high precision & stability); operation interface may be equipped with a color touch screen for convenient setting and operation |
Accessoriess
Main unit, computer, horizontal traction module, pneumatic gripper, 7-inch color touch screen
Test Procedures
1. Insert the intervention catheter into the test pulley system and secure it, ensuring the correct spacing.
2. Adjust the vertical spacing of the test pulleys to maintain a constant test distance.
3. Connect one end of the catheter to the traction head, ensuring the traction head guides the catheter at a constant speed.
4. Monitor the force value change curve during the test using computer software.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Adhesion evaluation equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications The adhesion tester has extensive application value in multiple fields, inclu... | |
| Mask respiratory resistance tester | The Mask Respiratory Resistance Tester is designed to measure the inhalation and exhalation resistance of respirators and masks under controlled co... | |
| Bending Tester STD- FB2 - S | Whats/App: +86 156 1874 6768 Web: standard-groups.com Compliance Standards ASTM D790 (Plastics Flexural Properties) ISO 178 (Flexural Strength ... | |
| UV Accelerated Weathering Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard: ASTM D4329、ASTM D4587、ASTM D4799 、ASTM D5208、ASTM G151、ASTM G154 、EN 927-6、EN 12... | |
| Wear Track Profilometer | A wear profiler is a device used to measure the characteristics of worn surfaces, providing high-precision wear rate measurements. It provides reli... |
Похожие товары
| Bluetooth catheter flow meter | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 15812.1 - 2005 Non - Vasc... | |
| Automatic Catheter Hydraulic Leakage Tester device | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 15812.1: Disposable steri... | |
| Balloon Catheter Leak and Compliance Tester | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO 13988-1:2017: Intravascula... | |
| Advanced catheter residual vacuum tester instrument | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards YY/T 0339 - 2019: Suction Cath... | |
| Bend catheter flexibility equipment | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO 10555 -Intravascular cathe... |