Adhesion Tester for quality control

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

- Specific standards for different test objects, such as the adhesion test standards for medical gels, cosmetics and skin

- care products, food packaging materials, etc.

- General standards in the adhesive industry, such as GB/T 2792 Test Method for 180°Peeling Strength of Pressure

- Sensitive Adhesive Tapes, ASTM D3330 Standard Test Method for Peel Adhesion of Pressure - Sensitive Tape, etc. These standards specify in detail the test methods for the adhesion tester, specimen preparation, test conditions, etc., providing a unified standard for the use of the tester.

Features

1. 7-inch touchscreen control, stylish human-machine interface

2. Lead screw drive system, greatly improving the accuracy of transmission displacement and enabling stepless speed regulation.

3. Allows for parameter setting, printing, clearing, calibration, and other operations.

4. Calibration interface password protected to prevent errors in calibration force values due to accidental operation.

5. Features power-off memory, overload protection, and line position protection.

6. The instrument has applied for a patent for its sensor protection device function, patent number: ZL.8

7. Equipped with a standard USB interface and a printer interface for easy connection to a computer.

8. The instrument can complete tests independently, and the mini printer automatically prints the test results.

Technical Parameters

Technical ParameterSpecification
Measurement Range Usually covering 0 - 30000mN or wider (adapts to different materials). Some devices have a smaller range (e.g., 0.01N - 100N), suitable for materials with relatively low adhesion.
Measurement Accuracy Generally grade 0.5 or higher; ensures the accuracy of test results.
Stroke Usually 650mm or longer; meets requirements of different specimen sizes and test methods.
Pressing Roller Mass Usually 2000g±50g; applies pressure to the specimen to ensure tight adhesion between the specimen and the test plate.
Test Speed Adjustable range: generally 1 - 500mm/min (stepless speed regulation) to simulate different peeling speeds; some devices have a wider speed-adjustment range.
External Dimensions Usually about 880mm (length) × 320mm (width) × 400mm (height); facilitates placement and movement (dimensions vary by brand/model).
Power Supply Usually AC 220V 50Hz (ensures normal operation of the tester); some devices support other voltages and frequencies.

Accessoriess

Standard Configuration: Main unit, standard fixtures, mini printer

Optional Accessories: Professional software, communication cable, test plate, standard pressure roller, sampling knife, custom-made non-standard fixtures

Note: This machine can operate independently. If connection to a computer is required, the computer must be provided by the user. The mini printer is thermal.

Test Procedures

1. Equipment Preparation

Connect to AC 220V 50Hz power supply. Use a standard adhesion sample calibration device conforming to GB/T 2792 (ensuring measurement accuracy of 0.5 grade or higher). Confirm that the pressure roller (2000g±50g), clamps, and other components are in normal condition.

2. Sample Preparation

According to the test object (e.g., medical adhesive removal, food packaging materials),

cut the sample according to the corresponding standard (e.g., ASTM D3330) (e.g., for 180-degree peel tests, the sample size must match). Clean the sample and test plate surfaces (remove impurities to avoid affecting adhesion).

3. Parameter Settings

Select the compliant standard (e.g., GB/T 2792) through the instrument interface, set the measurement range (e.g., 0-30000mN), and test speed (1-500mm/min, select as needed). 4. Sample Adhesion and Clamping

Affix the sample to the test plate and roll it back and forth with the pressure roller a specified number of times (e.g., the number of rolls required by the standard) to ensure a tight fit between the sample and the test plate.

Then, fix the test plate/sample to the instrument fixture, aligning it with the force direction of the peel/adhesion test (e.g., the fixture angle for a 180-degree peel).

5. Start the Test

After confirming that the clamping and parameters are correct, start the test program. The instrument will automatically complete the adhesion force test at the set speed and record the force value data in real time.

6. Result Processing

After the test, export the data and determine whether the adhesion force meets the requirements by referring to the corresponding standard (e.g., ASTM D3330). If multiple tests are required, repeat the sample clamping and testing steps.

7. Post-processing

Remove the sample, clean the fixture, pressure roller, and test plate, and turn off the equipment power.


Отправить запрос, связаться с поставщиком

Кому: Standard International Group (HK) Limited
Ваш E-mail:
Текст письма:


Send to other suppliers

Другие товары поставщика

Aircraft Seat Cushion Fuel Burn Test System for Aviation Safety
Aircraft Seat Cushion Fuel Burn Test System for Aviation Safety Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system is designed to support testing in accordance with the following aeros...
High-Precision Cone Calorimeter for Flammability Testing
High-Precision Cone Calorimeter for Flammability Testing Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards This apparatus conforms to the following international and national standards: ...
Flooring VOC Emission Test Chamber
Flooring VOC Emission Test Chamber Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard ISO 16000-9(Indoor Air - Determination of Volatile Organic Compounds) ASTM D51...
Fabric Ultraviolet Tester
Fabric Ultraviolet Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards AATCC 183 – Determination of SPF for textiles AS/NZS 4399 – Sun Pro...
Cabin Air Filter Test System(Gas Adsorption),ISO 11155 Cabin Air Filter Testing Equipment
Cabin Air Filter Test System(Gas Adsorption),ISO 11155 Cabin Air Filter Testing Equipment Whats/App: +86 156 1874 6768 Web: standard-groups.com The Filter Test System for gas adsorption/desorption of air filters for passenger compartme...
Все товары поставщика

Похожие товары

Surface adhesion testing equipment factory price
Surface adhesion testing equipment factory price Продавец: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards JIS L 1066 (Testing Methods fo...
Continuous load scratch resistance tester manufacturer
Continuous load scratch resistance tester manufacturer Продавец: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO DIS 12137-2 (Plastics &mda...
High precision continuous load surface tester
High precision continuous load surface tester Продавец: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for...
Multi-function surface performance test system
Multi-function surface performance test system Продавец: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for...
Laboratory tribology research equipment
Laboratory tribology research equipment Продавец: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for...