Average Particle Size Analyzer
The Average Particle Size Analyzer has a simple, fast and reliable technique for measuring average particle size. This measurement can not be easily achieved by any other technique. Combined with the innovation of automatic viscosity flow controller for precisely regulating the flow permeability, the instrument can be used for measuring A wide range of powder, including surface area with only a few grams of material per square meter.
Operation Principle
The instrument accurately measures flow rate of gas through the sample as a function of deferential pressure, calculates the envelope surface area, and uses these results to compute average fiber diameter. The envelope surface area is the external particle surface area that sees flow of gas through the sample. The envelope surface area is obtained from the flow rate and pressure drop using the Carman-Kozeny relation (Gerard Kraus, J.W. Ross and L.A. Girifalco, Surface Area Analysis by Means of Gas Flow Methods. I. Steady State Flow in Porous Media, Phys. Chem., V330-333). The average fiber diameter is obtained from the envelope surface area using the following equation: D = 6 / S ρ
where D is the average fiber diameter, S is the envelope surface area per unit mass of the powder, and ρ is the true density of the powder.
Application
Providing the user with an average particle diameter measurement in less than five minutes, the main application is quality control. Samples tested include pharmaceutical powders, electrode components, ceramic powders, chemical powders, metallic powders, and other porous substances.
The instrument is utilized for this function in several industries, such as:
·Chemical and Mineral • Fuel Cell • Battery
·Powder Metallurgy • Pharmaceuticals • Ceramic
Features
·Uses nitrogen or other noncorrosive gases; no expensive gas mixtures required
·Windows-based software handles all control, measurement, data collection, and report generation; manual control also possible
·Compatible with Windows 95 or higher
·Real-time graphical test display depicts testing status and results throughout operation
·Nondestructive testing
·Length of test approximately 5 minutes
·Wide range of acceptable sample types and sizes
·Minimal maintenance required
Technical Parameters
Items |
Parameters |
Sample size |
1~10cc |
Pressure range |
0~500psi |
Sealing gas |
Nitrogen or any non-corrosive gas |
Pressure transducer range |
0~500psi |
Mass flow controller range |
10cc/minute~100cc/minute |
Power supply |
110/220VAC,50/60Hz |
Dimension |
7" H x 17" W x 17" D |
Weight |
30lbs |
Другие товары поставщика
|
Тестер проникновения синтетической крови для медицинской маски для лица |
Этотпредназначен для проверки устойчивости медицинских масок к проникновению синтетической крови в определенных условиях тестирования.
Особеннос... |
|
Chrysler Flex Fold Tester |
Chrysler Flex Fold Tester
Standards:
Chrysler LP-463LB-9-01, Chrysler LP-463KB-13-01, Chrysler LP-463KB-10-01, IFAI J1531, Commercial Standard CS-2... |
|
fabric textile random tumble pilling tester |
The Tester consists of individually lit aluminum chambers containing stainless steel impellors continuously rotating the test samples against cor... |
|
Qinsun Auto Fabric Air Permeability Test Apparatus |
Testing Standards: BS 5636, JIS L1096-A, DIN 53887, ASTM D737, ASTM D3574, EN ISO 9237, GB/T 5453, EDANA 140.2, TAPPI T251, EDANA 140.1, ASTM D737,... |
|
Vibration Fiber Fineness Tester |
TRANSLATE with x
English
Arabic
Hebrew
Polish
Bulgarian
Hindi
Portuguese
Catalan
Hmong Daw
Romanian
... |
Все товары поставщика
Похожие товары