WDet-5000 Ammonia-Nitrogen Online Automatic Analyzer
Overview Of WDet-5000 Ammonia-Nitrogen Online Automatic Analyzer Wastewater
WDet-5000 is a self-developed ammoniacal nitrogen water qualityonline monitoring analyzer wastewater, based on the global initiative magnetic conductivity measurement platform, combining with constant optical fiber and digestion colorimetric integration technology. The ammonia-nitrogen automatic online monitoring analyzer wastewater has the features, including high precision, low detection limit, high stability and low maintenance frequency.
This ammonia analyzer wastewaterworks in accordance with the technical requirements of HJ/T 536-2009 Water Quality Ammonia-Nitrogen Measurement-Salicylic Acid Method, fully meets the pollution sources, municipal wastewater, industrial processes in the application areas, and has passed the certification of CMC and CCEP.
FeaturesTechnical Specification
Features Of WDet-5000 Ammonia-Nitrogen Online Automatic Analyzer Wastewater
High accuracy and reliability, strong capability of anti-interference
● Magnetic conductivity measurement technology used on the ammonia-nitrogen analyzer wastewater will be not affected by chromaticity, suspended solids, and bubbles, especially suitable for the monitoring of inlet water
● The ammonia-nitrogen analyzer wastewater has a special design of flow and reagent ratio, high precision, low detection limit(0.01mg/L)
● The ammonia-nitrogen analyzer wastewater adopts constant optical fiber and light intensity compensation algorithm, small drift for long-term operation
Low maintenance and cost
● The ammonia-nitrogen analyzer wastewater has a key self-check function, easy to debug, maintain, comparison test
● The ammonia-nitrogen analyzer wastewater has a self-diagnosis function, assisting maintainer to analyze the failure
● This type of water analyzer instrument importes single-channel high integration valve island which is simple and easy to maintain, dismantle and clean
● The ammonia-nitrogen analyzer wastewater has all-day networking for real-time inspection of instrument running state
High intelligent level
● Scan code function: rapidly obtain running/maintenance information of the the ammonia-nitrogen analyzer wastewater
● Alarm function: the ammonia-nitrogen analyzer wastewater provides intuitive and accurate information for grounding self-check, reagent insufficiency and etc
● Quality control: the ammonia-nitrogen analyzer wastewater is equipped with a parallel sample test, standard sample verification, and standard recovery rate detection
● Reverse control function: the ammonia-nitrogen analyzer wastewater starts automatically by flow and PH
Technical Specification Of WDet-5000 Ammonia-Nitrogen Online Automatic Analyzer Wastewater
Method |
Water Quality Ammonia-Nitrogen Measurement-Salicylic Acid Method |
Range |
|
Zero Drift |
<0.02mg/L |
Span Drift |
<1% |
Indication Error |
±10% |
Repeatability |
≤3% |
Memory Effect |
20% FS ± 0.3mg / L |
80% F.S. ±0.2mg/L |
|
Interval |
Hourly and interval measurement, and the instrument can also be started |
by serial port |
|
Maintenance |
<2 hour/month |
Quantitative |
|
Lower Limit |
|
Quality Control |
±10% |
Sample Comparison |
|
Water Sample |
>2mg/L±10% |
Comparison |
≤2mg/L±0.2mg/L |
Reagent |
1 set reagents is able for about 250 times of measurement (can be |
customized according to field needs) |
|
External |
1×4-20mA input/output (expandable to 2 ) |
Interface |
RS485/RS232/USB interface/WIFI/ Bluetooth (optional) |
Working Temp |
5℃~40℃ |
Power |
200W (220VAC 50Hz), regardless of the pump |
We also have types of portable gas analyzerfor sale, if you have needs, please contact us.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
Overview Of EM-5 Continuous Flue Gas Analyzer | HANGZHOU CHUNLAI (ZETIAN) TECHNOLOGY was established in 2008, and has national wide impact in China now as a fast-growing company specialized in en... | |
ERMS-2000-M Mobile Vehicle Exhaust Telemetry System | Overview The ERMS2000-M movable vehicle exhaust remote sensing system, developed by our company, is suitable for mobile enforcement of city roads a... | |
AM-5100 SO2 Analyzer | Are you looking for air analyzer equipment? Overview AM-5100 SO2 air pollution analyzeradopts ultraviolet fluorescence technique and combines advan... | |
EM-5 Flue Gas Analyzer | Overview EM-5 series flue gas analyzer is a series of self-developed gas analysis products for online gas analysisof domestic as well as foreign en... | |
DMS-100/MS-1000 Dust Monitor | Overview Of DMS-100/MS-1000 DMS100/MS-1000 is a self-developed online dust analysis product and a dust particle measuring device, adopting laser b... |
Похожие товары
GF302D1 PORTABLE THREE-PHASE kWH METER TEST DEVICE | Продавец: 782364 | GF302D1便携式三相电能表测试装置是一种误差测量仪器。GF302D1 具有 0.05% 的高精度、便携设计和对环境的高耐受性。特点: 1.量程:电压:0-500V;电流:0-120A;2.可... | |
WDet-7000 Integration Water Quality Online Analyzer | Продавец: Ханчжоу Zetian Technology Co., Ltd. | WDet-7000 multi parameter water analyzeris a high-level integration of a multi parameter water an... | |