Buy Medical glove water-impermeability tester

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

- GB 10213 - 2006 Impermeability Test in Appendix A of Medical Rubber Examination Gloves

- GB 7543 - 2006 Impermeability Test in Appendix A of Sterile Rubber Surgical Gloves

- ASTM D5151 Standard Test Method for Detection of Holes in Medical Gloves, etc.

Features

1. Independently designed and original, with a beautiful and elegant appearance.

2. The bucket is made of high-grade acrylic, which will never yellow.

3. Aluminum alloy frame and stainless steel fittings ensure sturdiness, durability, and resistance to deformation.

4. Equipped with lockable swivel casters for easy movement.

Technical Parameters

Serial No.Parameter ItemParameter Details
1 Capacity At least 1000mL; ensures water filling and keeps gloves in a vertical state during testing
2 Number of Stations Equipped with multiple stations (e.g., 12 stations) to improve testing efficiency; each station has an automatic clamping function to ensure gloves' stability and accuracy during testing
3 Liquid-adding Function Equipped with automatic liquid-adding function; liquid-adding time can be set, which simplifies the testing process and improves testing accuracy
4 Timing Function Equipped with an independent timer to record test time; ensures the accuracy and repeatability of test results
5 Operation Interface Equipped with a large-screen touch screen; supports operation mode with Chinese and English menus, facilitating user operation
6 Transparent Cylindrical Tube Outer diameter: 60mm; inner diameter: 50mm; capacity: 1000mL of water; used to place gloves for testing

Accessoriess

1. Suspension device: The gloves are secured through a suspension system to prevent displacement during testing, thus avoiding deviations in the results.

2. Special sample clamp: Ensures a tight seal between the glove and the testing system, preventing liquid leakage that could affect the test results.

3. Pressure monitoring: Some models are equipped with a precision pressure gauge that displays the air pressure value in real time, with an accuracy of ±1kPa, ensuring testing stability.

Test Procedures

1. Preparation Before Operation

Equipment and Accessory Check:

Confirm that the instrument's transparent cylindrical tube (outer diameter 60mm, inner diameter 50mm, capacity 1000mL), multi-station (e.g., 12-station) automatic clamping function, automatic liquid filling, and timing module are functioning correctly; check that the equipment power supply and operating interface (Chinese and English touch screen) are stable.

Sample Preparation:

Select the medical gloves to be tested (e.g., rubber examination gloves, sterile surgical gloves), ensuring that the gloves are free of damage and stains, and conform to the product type corresponding to the test standard.

Standard Matching:

Clearly identify the standard to be followed for the test, and confirm the corresponding test duration requirements.

2. Equipment Parameter Settings

Start the instrument and select the matching test standard via the Chinese and English touch screen;

Set core parameters:

Liquid Filling Time: Set the automatic liquid filling duration according to the transparent cylindrical tube capacity (1000mL);

Test Time: Set the independent timer according to the duration required by the corresponding standard ;

Station Selection: Enable the required number of stations to improve testing efficiency.

3. Sample Installation

Place each medical glove onto the transparent cylindrical tube port of each station;

Activate the automatic clamping function of each station to fix the gloves and ensure they are in a vertical position, avoiding wrinkles or looseness (to prevent leakage during the test that is not due to the glove itself).

4. Automatic Liquid Filling and Test Start

Activate the automatic liquid filling function to fill each transparent cylindrical tube with water to 1000mL, ensuring the gloves are completely covered with water;

After confirming that the liquid filling is complete and there is no initial leakage, start the timer to begin the water tightness test.

5. Test Process Monitoring

During the test, observe the status of the gloves at each station: whether there is water leakage or penetration marks;

Maintain a stable testing environment (e.g., room temperature 20±5℃), avoiding external forces interfering with the equipment or gloves.

6. Test Completion and Result Determination

After the test duration is complete, stop the timer and drain the water from the cylindrical tube;

Remove the gloves and determine the results one by one: gloves without leakage are considered "qualified", and those with leakage are considered "unqualified";

Generate a test report using the equipment, recording sample information, test standards, and results for each test station.

7. Equipment Post-Maintenance

Clean any remaining water stains from the transparent cylindrical tube and the test station fixtures;

Check the operating status of the automatic liquid filling and clamping functions, organize and archive the test report, and ensure the stability of the equipment for future use.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

Automatic European Standard Washer for Colorfastness and Shrinkage Testing
Automatic European Standard Washer for Colorfastness and Shrinkage Testing Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Feature (1) Multi-layer design with flange plate for stress distribution and...
Gas Explosion Limit Tester,UL9540A
Gas Explosion Limit Tester,UL9540A Whats/App: +86 156 1874 6768 Web: standard-groups.com Introduction The explosion limit tester can realize the explosion limit measurement of com...
Fabric Flame Retardant Tester
Fabric Flame Retardant Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com A method is used to determine the damaged area and the length of the damage when the textil...
Model Coin-cell Swelling System
Model Coin-cell Swelling System Whats/App: +86 156 1874 6768 Web: standard-groups.com The Model Coin-Cell Swelling System is designed for in-situ measurement of thickness and vo...
Automatic Catheter Hydraulic Leakage Tester device
Automatic Catheter Hydraulic Leakage Tester device Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 15812.1: Disposable sterile catheters - Part 1: General requirements Featu...
All supplier products

Same products

ER4047 Aluminum Mig Welding Wire
ER4047 Aluminum Mig Welding Wire Seller: 868776 Hangzhou Kunli Welding Materials Co., Ltd. specializes in manufacturing reliable welding material...
ER4043 Silicon Aluminum Welding Wire
ER4043 Silicon Aluminum Welding Wire Seller: 868776 Hangzhou Kunli Welding Materials Co., Ltd. specializes in manufacturing high-quality welding mate...
Choosing the Right Equipment for Heavy-Duty Cleaning
Choosing the Right Equipment for Heavy-Duty Cleaning Seller: Taizhou Feike Mechanical Technology Co., Ltd. FEIKE’s High Pressure Washer is designed for industrial cleaning, equipment maintenance, ve...
ER4043 Silicon Aluminum Welding Wire
ER4043 Silicon Aluminum Welding Wire Seller: 868776 Hangzhou Kunli Welding Materials Co., Ltd. offers ER4043 Silicon Aluminum Welding Wire for alumin...
HDPE High Speed Water Cooled Double Wall Corrugated Pipe Extrusion Line
HDPE High Speed Water Cooled Double Wall Corrugated Pipe Extrusion Line Seller: Shandong HUASU Plastic Technology Co.,Ltd. The molds adopt the double flow channel structure and nitriding polishing processed die head. Thi...