G506 Mask Automatic Filter Performance Tester,CE certified mask filtration tester
The latest G506 mask automatic filter performance tester uses the most advanced technology and design concepts, to meet a variety of test standards and the latest National standard GB / T 32610.Only relatively little operation is required tocomplete all the function and operation. Hardware performance of G506 mask automatic filter performance tester is stable and excellent in repeatability. Software design meet with people's thinking mode, the interface is user-friendly, and the functional and parameter settings are very reasonable. It is convenient to userforacademic research and product development.
Application
G506 Mask Automatic Filter Performance Tester is used fortestingprofessional filter material and filtration efficiencyof a wide range of air filters including HEPA and ULPA filter, respirator, filter cartridge and protective clothing test. It also can use salt or oil mist to test filter media. Rapid acquisition of filter media performance evaluation parameters, and accurate evaluation of filter material quality, performance, life and so on. It is convenientto userforacademic research and product developmentof new filter materials, product quality control, material performance verification.
Standards
ASTM F2100:Standard Specification for Performance of Materials Used in Medical Face Masks
EN149:Respiratory protective devices - Filtering half masks to protect against particles - Requirements, testing, marking
NIOSH-42 CFR84:NIOSH Guide to the Selection and Use of Particulate Respirators(或 42 CFR Part 84 - Approval of Respiratory Protective Devices)
Feature
Complies to more standards than any other filter tester
Salt and oil testing in one unit
Highly consistent test results
New user-interface and data recording options
Customizable fully automated testing sequences
Ergonomic design
Reduced cost of ownership with new serviceable photometers
Worldwide service and support
Parameters
| Items | Parameter |
| Apllication | Load and quality control |
| Max Efficiency | 99.999% |
| Aerosol Type | DOP/PAO/DEHS/Paraffi/Others Oil Aerosol、Salt |
| Aerosol Generation | Atomizing |
| CMD | 0.2μm(Oil)or 0.75um(NaCl) |
| GSD | <1.6(Oil)or <1.86(NaCl) |
| Flow Rate | 15-100L/min |
| Resistance | 0-150mm H2O(0-1470Pa) |
| Particle Detect | Scatter Photometer |
| Measurement Time | 10 Sec. |
| Data Report | LCD,RS-232 |
| Operation | Independent tewster、automatic production line |
| Standard compliant | US 42 CFR part 84.EN 143.JMOL,ISO 23328-1 ISO 16900-3 |
| Power Supply | 115VAC、60Hz,10A或230VAC,50Hz,5A |
| Compressor Air | 80psig,7scfm(550kPa,198std.L/min) |
| Dimension | 70cm x 70cm x 147cm(28in, x 28in.x 58in.) |
Accessoriess
Main Unit: As the core component of the testing instrument, it is responsible for overall control and data processing.
Aerosol Generator: Includes salt particulate aerosol generators and oil particulate aerosol generators, used to generate the aerosol particles required for the test.
Classifier: Utilizes the principle of inertia to collect unwanted large particles into a collection bottle, ensuring that the aerosol particles entering the testing system meet the requirements.
Particle Counter: Usually equipped with two units, used to measure the number or mass concentration of aerosols upstream and downstream of the filter media, and is a key device for calculating filtration efficiency. Some systems may use a laser scattering photometer instead of a particle counter.
Static Neutralization Device: Neutralizes the static electricity carried by the aerosols, reducing particle loss and the influence of static electricity on the test results.
Differential Pressure Sensor: Monitors the pressure drop across the filter media, assessing changes in resistance during the filtration process.
Flow Meter: Monitors the flow rate during the test, ensuring consistency of test conditions. Flow meters are typically 2.5 grade accurate, ranging from 10-100 L/min.
Vacuum pump: Provides the necessary negative pressure environment to ensure aerosol particles can smoothly enter the testing system.
Pneumatic clamps and protective devices: Employ pneumatic clamping, offering convenient operation and high safety, while also featuring protective devices to ensure safe use.
Programmable logic controller (PLC) and touchscreen: Enables automated control of the testing process, including control, monitoring, calculation, and data display functions. Some devices utilize a PLC control system and are equipped with a 7-inch color touchscreen.
Instruction manual and power cord: Provides an operation guide and power connection instructions to ensure users can correctly operate the equipment.
Test Procedures
Operating Procedures:
First: Prepare the accessories and test sample;
Turn on the instrument and enter the LCD screen operating interface;
Perform a no-load test first;
Press the dual-control pneumatic keys on both sides to lower the clamp;
Set the airflow rate and aerosol particle parameters;
After the data stabilizes, proceed to the next step;
Obtain the amount of aerosol upstream of the filter media;
Once the data stabilizes, a "Place Clamp" dialog box will pop up on the screen;
Place the prepared sample into the instrument clamp and clamp it securely;
Start the test;
Wait a few seconds to obtain the amount of aerosol downstream of the filter media and calculate the filtration efficiency;
The instrument will automatically print out the test report;
Remove the sample;
After the test, turn off the instrument and evaluate the filtration performance of the material based on the test data results.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Appliance Stability Tester | Product Introduction The device conforms to GB4706-1 Article 20.1. IEC60335-1 Article 20.1 and other standards. The device is composed of a workbe... | |
| Wet leakage current test system of photovoltaic module | The Wet Leakage Current Test System is designed for high-precision evaluation of insulation resistance and leakage current in photovoltaic modules ... | |
| HEPA Filter Element Quality Control Test SystemEN 1822-3 HEPA and ULPA filter testing System | The HEPA Filter Element Quality Control Test System is designed for quality control of small HEPA filter elements in production. It complies with E... | |
| Sauce Packet Pressure Testing Instrument | Applications The main purpose of the pressure resistance tester for sauce packets is to evaluate the ability of soft packaging materials like sauc... | |
| Flame Propagation Test for a Single Insulated Cable IEC / EN 60332-1-2,EN 50086 | Description: This testermeetthe standards IEC 60332-1-1:2004 in the provisions of technical indicators and requirements. Single cable combustion ... |
Похожие товары
| Узел управляющего клапана Common Rail F00ZC01307 | Продавец: Pioneer Diesel Parts | Pioneer Diesel Parts Whats/App: +86 152 6025 8100 Web: pioneerdieselparts.com Управляющий клап... | |
| Комплект клапанов управления топливными форсунками F00RJ01479 | Продавец: Pioneer Diesel Parts | Pioneer Diesel Parts Whats/App: +86 152 6025 8100 Web: pioneerdieselparts.com Комплект клапано... | |
| Комплект клапанов топливных форсунок F00RJ02213 | Продавец: Pioneer Diesel Parts | Pioneer Diesel Parts Whats/App: +86 152 6025 8100 Web: pioneerdieselparts.com Комплект клапано... | |
| Комплект клапанов форсунки Common Rail F00RJ00339 | Продавец: Pioneer Diesel Parts | Pioneer Diesel Parts Whats/App: +86 152 6025 8100 Web: pioneerdieselparts.com Комплект клапано... | |
| Топливный насос высокого давления 723945- 51320 для Yanmar WB93R, WB97R, WB97S, WB98A | Продавец: Pioneer Diesel Parts | Pioneer Diesel Parts Whats/App: +86 152 6025 8100 Web: pioneerdieselparts.com Топливный насос ... |