AFC-133 Filter Material Test Bench
3. Dust capacity test. The dust capacity test can test over a wide range of dust concentrations and surface wind speed. The dust capacity returns to zero through the filter pressure trigger and backflush system. It also record test flow and other air parameters (temperature, humidity).
4.Differential pressure flow test. Evaluate the relationship between the resistance of the filter and the flow.
5.Pulse cleaning performance test. Do the aging test of pulse cleaning according to ISO11057 and VDI3926 to evaluate the filter aging performance.
| Items | Parameters |
| Air flow | 0.5~65m3/h±1% |
| Differential pressure | 10~2500Pa±0.1% |
| Cross section | 176cm2(diameter 150mm) |
| Cylinder pressure | 0.2~0.6MPa |
| Agitated time | 50~150 ms |
| Optical particle counter | LAP322 particle size spectrometer, 0.2~40μm(0.2~5μm, 0.7~40μm), 128 channel(test concentration<104 /cm3) |
| Test aerosol | Dust (AC fine, AC coarse, soot, calcite, limestone).Droplets (DEHS, DOP, paraffin, emery 3004).Salts (sodium chloride solution, ammonium sulfate solution). PSL small ball. |
| Power | 3×230 V, 16A |
AFC-133 Filter Material Test Bench
◆ Test the filter materials according to the standards (air filter EN779, air filters for passenger compartments ISO11155-1 ISO11057, VDI3926).
◆ Develop and optimize filter materials.
◆ Good quality assurance.
◆ Smart design makes its application with greater scalability and extensibility.
◆ The test bench can be operated and controlled by user-friendly PAFWin3.0 software.
◆ Aerosols are available in a wide range (gas, dust and droplets).
◆ The optimal combination of instruments makes it the most cost-effective.
2. Weight filtration efficiency. By measuring the end of the high efficiency filter and the weight of the filter materials to test the filter efficiency (ISO-5011).
AFC-133 filter materials test bench mainly test channel (including filter fixture), flow control unit, dust diffuser, optical particle counter and other components and control and data acquisition software. Horizontal device test dust pipe is in line with VDI-3926 with diameter 100mm of 1.4301 stainless steel tube. The flow control unit has designed a mass flow controller. The optical particle counter with switching valve (switching upstream and downstream aerosol samples) can measure the grading filtration efficiency of 0.2 ~ 40μm. At the same time, the particle counter can be purged by purging the air that up and down the sample. Moreover, Time can be selected by software.
The high efficiency filter installed at the end of the air purifying channel is used to measure the downstream aerosol concentration of filter in test state. The weighting method is the way to get filtration efficiency by measuring the weight of the backup filter and the filter under test. The filter clamp and the backup filter are free to be placed and removed from the test channel, making it easy to test and clean.
Control and data acquisition software PAFWin is the professional software based on filter materials and filter test and development. This software controls the operation of the test bench and test data records, which can be easily controlled and operated in Windows3.x / 95 / NT operation System. It not only can automatically conduct an experimental test according to the issued standard, but also can prompt the maintenance needs of manual control, calibration experiments and research and other information.
Send product request
Other supplier products
| electronic yarn twist tester | Standards: ASTM D1422/1423,BS 2085,DIN 53832,ISO 2061,IWTO 25-70,GB/T 2543.1-2001/2543.2-2001 Applications: TheElectric Twist Testeris profess... | |
| Static Tensile Tester | Static tensile testerApplicationUsed for static tensile test of elastic fabric performance, adopts aviation aluminum frame design, 304 stainless st... | |
| High Quality Testing Instrument for Mask burning tester | Introduction: The flame-retardant tester of medical mask is used to test the flame-retardant performance of the mask when it is accidentally contac... | |
| Dry Wet rubbing fastness Tester | Testing Standards:ASTM F1319, AATCC 8/165, BS 1006 D02, GB/T 3920, ISO 105 D02/X12, JIS L0849, M&S C8/C8A, NEXT TM6Applications:Electronic Croc... | |
| Medical Face Mask Bacterial Filtration Efficiency (BFE)Tester | Standards: · NIOSH 42 CFR Part 84 NIOSH Respiratory User Notices 42 CFR Part 84 Respiratory Protective Devices. · EN 134 Respirato... |
Same products
| C500-1.7 Series High Flow Multi-Stage Centrifugal Blower /Centrifugal Blower | Seller: Greentech International (Zhangqiu) Co., Ltd | Greentech International (Zhangqiu) Co., Ltdis the professional blower supplier.Multistage Centrif... | |
| Common Rail injector control valve F00VC01303 | Seller: China Lutong Part Plant | Common Rail injector control valve F00VC01303 Tina #Stainless Steel Cylindrical Pin# #Denso Comm... | |
| Common Rail injector control valve F00VC01200 | Seller: China Lutong Part Plant | Common Rail injector control valve F00VC01200 Tina #Stainless Steel Cylindrical Pin# #Denso Comm... | |
| Injector Control Valve Plate 24# | Seller: China Lutong Part Plant | Injector Control Valve Plate 24# Tina #Stainless Steel Cylindrical Pin# #Denso Common Rail Injec... | |
| Injector Control Valve Plate 12# | Seller: China Lutong Part Plant | Injector Control Valve Plate 12# Tina #Stainless Steel Cylindrical Pin# #Denso Common Rail Injec... |