Medical Surgical Masks Air Exchange Pressure Difference Tester
Standards:
YY 0469-2011 Surgical mask YY 0969-2013 Single-use medical face mask EN 14683:2014 Medical face masks - Requirements and test methods
Applications:
Product Information:
Medical Face Mask Air Exchange Pressure Difference Testermeet the standard requirements, which is YY0469-2004 "Medical surgical mask technical requirements" and YY 0969-2013 "Disposable medical masks", medical face mask air exchange pressure difference tester is measured by gas pressure difference on both sides of the mask (ie the damping from the mask exert to the airflow) to determine the breathability of the mask, and the instrument test results is accurate.
TheMedical Face Mask Air Exchange Pressure Difference Testeris used for the determination of gas exchange pressure difference of medical surgical masks and can also be used to determine the gas exchange pressure difference of other textile materials.
Standards:
YY 0469-2011Surgical mask
YY 0969-2013Single-use medical face mask
EN 14683:2014Medical face masks - Requirements and test methods
YY 0969-2013Single-use medical face mask
EN 14683:2014Medical face masks - Requirements and test methods
1, using the suction gas source as the instrument power source, free from test space constraint;
2, with a high precision differential pressure sensor, digital show the sample pressure of both sides;
3, special sample holder to ensure a good clamping sample.
2, with a high precision differential pressure sensor, digital show the sample pressure of both sides;
3, special sample holder to ensure a good clamping sample.
Technical Parameters
Items | Parameters |
Air supply | Suction type |
Air flow | 8L / min |
Seal type | O-ring seal |
Sample breathable diameter | Φ25mm |
Differential pressure sensor range | 0 ~ 500Pa |
Display mode | Digital display pressure difference |
Power supply | 220V, 50Hz |