Anti Synthetic blood penetration tester
Standards:
GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood. YY/T 0691-2008 Clothing for protection against infectious agents-Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected) YY 0469-2011 Technical requirements of medical face masks– Blood penetration test equipment. ISO 22609:2004 Clothing for protection against infectious agents - Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected) ASTM F1862-07Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Know Velocity)
Applications:
This Anti Synthetic blood penetration tester is to test the resistance of medical face masks to the penetration of synthetic blood under specified test conditions.
Product Information:
Spraying distance: 300mm-305mm adjustable | Nozzle length: 12.7mm |
Space between front and rear plates: 0.6cm | Target plate bore diameter: 0.5cm |
Control panel: Touch-tone | Total weight: 35Kg |
Nozzle size: 0.84mm | Power supply: 220V 50HZ |
Spraying speed: 450cm/s, 550cm/s, 635cm/s | Appearance size: 1000mm*500mm* 650 mm |
Spraying pressure: Standard/10.6kPa, 16kPa,21.3kPa/adjustable |