Filter Media Test System,ISO 5011 filter media performance tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Test System for multiple parameter testing at filter media, space-saving design by vertical-orientated flow channel. Performance of filter media is essentially defined by parameters such as differential pressure, gravimetric filtration efficiency and dust holding capacity.
This test bench has been designed as universal testing equipment. It combines a simple and robust setup with a manual and therefore cost efficient handling and operation. Because of its vertical test channel setup it is especially suitable for the testing of filter media according to ISO 5011.
Applications
Quality assurance of flat sheet filter media in terms of a final clearance and a receiving inspection
performance testing of filter materials during research and development stage.
Standards
GB/T 6165 — High Efficiency Air Filters — Performance Test Methods
GB 2626 — Respiratory Protection — Non-powered Air-Purifying Particle Respirators
EN 13274-7 — Respiratory Protective Devices — Test Methods — Part 7: Determination of Particle Filter Penetration
ISO 5011 — Inlet Air Cleaning Equipment for Internal Combustion Engines and Compressors — Performance Testing
ISO 29463-3 — High-Efficiency Filters and Filter Media for Removing Particulate Matter in Air — Part 3: Testing Flat Sheet Filter Media
ISO 29463-5 — High-Efficiency Filters and Filter Media for Removing Particulate Matter in Air — Part 5: Test Method for Filter Elements
Technical data
| Parameter | Unit | Value |
|---|---|---|
| Test flow rate | m³/h | 1.5-18 |
| Measuring range, differential pressure | Pa | 5000 |
| Compressed air supply (operating gas/air) | bar | 6 |
| Power supply | – | 110 … 230 VAC, 50/60 Hz |
| Operating medium (aerosol) | – | DEHS, ISO 12103-1 A2 |
Features
Reliable timesaving testing of gravimetric filter efficiency, dust holding capacity, differential pressure and fractional filtration efficiencyholders for test filter and backup filter separately removable, weighable without filter removal
Backup filter for protection of the flow rate unit in case of a filter breakthrough
Integrated Solid Aerosol Generator SAG 410 for feeding rates from approx. 1.5 - 20 g/h
Principle
- Flow rate unit: injector pump with a set of critical nozzles to adjust different flow rates
- Intake test air is directed through a special intake section for homogenous mixing with the test aerosol
- Test filter holding manually operable
- Solid aerosol generator exchangeable by different models of the series
- Control of the Solid Aerosol Generator integrated to the test rig
- Test dust dispersion by using an injector nozzle
- Suitable especially for test dusts ISO 12103-2 A2/A4
Accessories
Differential pressure measurement
Fractional efficiency measurement:
- Aerosol Spectrometer
- Aerosol Dilution System
- Sample Switching, controllable by means of software of aerosol spectrometer or via serial interface
Test Procedures
The flow rate is adjusted by a flow rate unit consisting of an injector pump with a set of critical nozzles;
The intake test air is guided through a dedicated intake section to ensure homogeneous mixing with the test aerosol;
The test filter is mounted in a manually operable filter holder;
The solid aerosol generator can be exchanged with other models of the same series and is fully integrated into the test rig control;
Test dust is dispersed into the airflow using an injector nozzle, making the system particularly suitable for ISO 12103-2 A2/A4 test dusts;
Under defined test conditions, filtration efficiency, differential pressure, and dust holding capacity are measured and recorded.
Maintenance Information
(1) Regularly inspect and calibrate the flow rate unit and differential pressure measurement devices
(2) Keep the aerosol generator and injector nozzles clean to prevent dust blockage
(3) Ensure the compressed air quality meets system requirements
(4) Clean the test channel and filter holders after each test
(5) Store the system in a clean and dry environment during long-term shutdown
在线联系供应商
Other supplier products
| Automated Coating Adhesion and Scratch Test Device | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards JB/T 8554-1997 Automatic Scratch Test for Coating Adhesion (Mechanical Industry ... | |
| Automotive Wires Ozone Aging Chamber | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO 6722: Road vehicles – Electrical cables – Dimensions and test me... | |
| Particle Strength Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Particle Strength Tester is a high-precision instrument designed to measure the mechani... | |
| Torque - to - Failure Tester for Metal Bone - connecting Screws | Standard YY/T 0662-2008(Chinese standard for orthopedic implant testing) ISO 5832-1(Metallic materials for surgical implants – Wrought stai... | |
| Advanced Vehicle and Aircraft Air Intake Filter Test System | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system complies with: (1) ISO 5011 – Air filter test for combustion e... |
Same products
| Reciprocating Tribometer | 卖方: Standard Groups | The Reciprocating Tribometer is a value-engineered tribological testing instrument designed for l... | |
| High Frequency Friction Machine | 卖方: Standard Groups | The High Frequency Friction Machine is a reciprocating tribological testing instrument designed t... | |
| Dry Friction and Wear Tribometer (Value Engineered Tribometer) | 卖方: Standard Groups | The Dry Friction and Wear Tribometer is a value-engineered, multi-functional testing instrument d... | |
| Four-Ball Tribometer(Value Engineered Tribometer) | 卖方: Standard Groups | The Four-Ball Tribometer is a high-precision laboratory instrument designed to evaluate the wear ... | |
| Rotary Tribometer ASTM D2266 DIN 50324 ISO 11008 | 卖方: Standard Groups | The Rotary Tribometer is a multifunctional friction and wear testing instrument designed for comp... |