Walk-In Thermal Comfort Environmental Chambe
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
Internal compliance with laboratory safety and electrical standards.
Temperature and humidity control performance aligned with industrial environmental simulation requirements.
Customization may include air exchange compliance per client specifications.
Parameters
| No. | Item | Technical Specifications |
|---|---|---|
| 1 | Chamber Dimensions | External: 4 m (L) × 4 m (W) × 3.2 m (H) Internal: 3.8 m (L) × 3.8 m (W) × 2.6 m (H) Standard insulated door: 0.8 m × 2.0 m with double-glass observation window Internal lighting: 1 daylight lamp; power outlet: 2.5 kW Digital display panel for temperature and humidity readings Optional window installation available upon request |
| 2 | Temperature Range | 10–43°C ±0.5°C adjustable |
| 3 | Humidity Range | 35–90% ±5% adjustable |
| 4 | Air Change Rate | 4–12 ACH; optional single-pass fresh air supply available (additional installation required) |
| 5 | Electrical Control System | Centralized control panel; Supply: AC 380V/50Hz, max power ≤15 kW; Equipped with short-circuit and over-temperature protection |
| 6 | Working Environment | Standard laboratory conditions |
| 7 | Other Requirements | Refrigeration unit piping and electrical lines default 5 m; site preparation including leveling, wall openings, rain shelter, water supply, drainage, and main power line to control box to be provided by user; chamber panels: 100 mm PU double-color steel |
| 8 | Special Notes | Default specifications are factory standards; custom requirements exceeding these parameters require additional consultation and pricing |
Features
High-Precision Temperature and Humidity Control: Maintains stable, uniform environmental conditions with ±0.5°C and ±5% RH accuracy.
Flexible Chamber Design: Spacious interior allows large-scale experiments and accommodates multiple test subjects or equipment simultaneously.
Observation and Control: Equipped with observation window, internal lighting, and digital panel for real-time monitoring of chamber conditions.
Customizable Air Exchange: Air change rates can be adjusted; optional fresh air integration ensures specific ventilation needs.
Safety and Reliability: Electrical protection, short-circuit, and over-temperature safeguards ensure operational safety.
Modular Installation Support: Refrigeration, water supply, drainage, and electrical preparation allow for seamless on-site setup.
Accessories
Digital display panel for temperature and humidity monitoring.
Standard insulated chamber door with double-glass observation window.
Internal lighting system (daylight lamp).
Optional fresh air supply system.
Test Procedures
Chamber Preparation: Verify power, water, and drainage connections; ensure refrigeration system is operational (~10 min).
Set Environmental Conditions: Program desired temperature (10–43°C) and humidity (35–90%) on control panel or software interface (~5 min).
Load Samples/Subjects: Place equipment, test materials, or participants inside the chamber; ensure proper positioning and spacing (~10 min).
Run Test: Start test program; monitor chamber conditions in real-time via display panel or software interface; adjust parameters if needed (~duration dependent on experiment).
Data Recording and Export: Temperature, humidity, and other environmental data can be recorded, saved, and exported for analysis (~5 min).
Completion and Retrieval: End program, allow conditions to return to safe levels, and remove samples or subjects safely (~10 min).
Maintenance Information
Regularly clean observation windows, internal lighting, and control panels.
Inspect refrigeration unit, piping, and electrical connections for leaks, damage, or wear.
Verify sensor calibration periodically to ensure accurate temperature and humidity readings.
Maintain chamber in a clean, dust-free, and temperature-stable environment.
在线联系供应商
Other supplier products
| Coatings Shear Fatigue Test device | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0988.13-2017 (China): Specifies shear fatigue testing protocols for coatings... | |
| Laboratory Burner for Insulation Fire Performance | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system is compatible with the following international aerospace fire testing... | |
| Solid Auto - ignition Point Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction This tester determines the self - ignition characteristics of solid c... | |
| Micro Rapid Closed Cup Flash Point Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Reference Standards: SH/T 0768. SN/T 3077.1. DL/T 1354. ASTM D6450 SN/T 3077.2. ASTM D7094... | |
| Direcct flame impingement Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Introduction The testerisdesignedaccording to the BS476-12:1991.Method of test for ignit... |
Same products
| Walk-In Thermal Comfort Environmental Chambe | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards Internal compliance with labor... | |
| Laboratory Manikin for Simulating Human Coughing and Airflow | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 14296-2005 — Indoor... | |
| Breathing simulator manikin | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 14296-2005 — Indoor... | |
| Seat Comfort Thermal Manikin | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 18830-2009: Test methods ... | |
| Elmendorf Tearing Strength Testers ASTM D1424 | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com G008 Manual Tearing Tester, The ... |