PN-2000FB Respiratory Humidifier (Support Heated Wire)
Pigeon MedicalRespiratory Humidifier PN-2000FB offers an integrated solution to provide humidified and warmed gas to those patients in need of mechanical ventilation or other respiratory supports. With built in servo control feedback system and alarm system, it incorporates to monitor constantly and provide visible and audible alarm according to the proximal and distal temperature respectively.
In addition, the over-heat protection will automatically perform thermal cut-off once the temperature of the heating plate excess 95oC to avoid any unexpected circumstances. PN-2000FB also supports heated wire on both disposable and reusable breathing circuit, ensure to avoid temperature loss and water condensation while the medical gas travels through the circuit.
Working Principle of PN-2000FB Respiratory Humidifier
Pigeon Medical Respiratory Medical HumidifierPN-2000FB is designed to use in hospital environment which can be capable with mechanical/ non-mechanical ventilation and oxygen therapy for adults. Meanwhile, it supports invasive ventilation, infant resuscitation and bubble CPAP and oxygen therapy for neonatal care.
With optional heated wire function, it matches with both disposable/reusable and heated wire/non-heated wire breathing circuit.
Features of PN-2000FB Respiratory Humidifier
Heated or non-heated wire temperature control with 3 different phases independently
Match with all kind of ventilators
Provide different connectors for pediatrics and adults breathing circuit
Built-in alarm function
Provide high quality aluminium clamp and slot plate for the blender to safely hang on the trolley
Specification of PN-2000FB Respiratory Humidifier
Model |
PN-2000FB |
Power Source |
AC 220V, 50-60Hz (Optional: AC 110V, 50-60Hz) |
Heated Plate |
150W |
Thermal Cut Off |
95±6℃ |
Display |
Three digit, 0.56 inch, 7 segment Led |
Temperature Monitoring |
0-75℃ |
Max-temperature of Heated Plate |
85℃ |
Heated Wire |
AC 23V 60W, 50-60Hz |
Power Rate |
220W |
Alarm |
Yes |
Temperature Control Setting |
Heated Wired Mode:P1: 25-29℃, P2: 29-35℃,P3: 35-39℃ |
Non-heated Wired Mode:P1: 23-27℃, P2: 27-32℃,P3: 32-37℃ |
|
Range |
10-70oC, Accuracy: ±2℃ (In 25-45℃ temperature range) |
在线联系供应商
Other supplier products
Adult High Flow Nasal Cannula | Pigeon MedicalNasal Cannula are designed to allow for patients with independent breathing and through provid high-flow, warmed, humidified breathin... | |
Air Oxygen Blender | Pigeon Medicalair-oxygen blender is a device used by medical personnel to control the concentration and flow of oxygen for the rescue of patients w... | |
Oxygen Analyzer | Oxygen Analyzer is normally used to detect, monitor, and analyze the oxygen concentration of ambient air or mixed gas. It can work independently an... | |
PN-3000 60LPM Medical Air Compressor | Ergonomic outlook and structure, high quality material, stable performance. All these features will make you fall in love with Pigeon MedicalAir Co... | |
Ventilator Circuit Support Arm (Light Type) | Pigeon MedicalCircuit Support Arm (Light Type) hook can provide a secure connection for the Ventilator, CPAP and Anesthesia Machine breathing circu... |
Same products
Ventilator Circuit Support Arm (Heavy Type) | 卖方: Guangdong Pigeon Medical Apparatus Co.,Ltd | Ventilator Circuit Support Arm (Heavy Type) Pigeon Medical Circuit Support Arm (Heavy Type) hook... | |
Air Oxygen Blender | 卖方: Guangdong Pigeon Medical Apparatus Co.,Ltd | Pigeon Medicalair-oxygen blender is a device used by medical personnel to control the concentrati... | |
Adult High Flow Nasal Cannula | 卖方: Guangdong Pigeon Medical Apparatus Co.,Ltd | Adult High Pressure Nasal Cannula Pigeon MedicalNasal Cannula are designed to allow for patients... | |
AD-I Bubble CPAP | 卖方: Guangdong Pigeon Medical Apparatus Co.,Ltd | What Is Bubble CPAP? AD-I NCPAP(nasal Continuous Positive Airway Pressure) AD-I infant bubble C... | |
AD3000-TPA Infant Resuscitator | 卖方: Guangdong Pigeon Medical Apparatus Co.,Ltd | AD3000-TPA Infant Resuscitator With safe, precise and controllable PIP (Peak Inspiratory Pressur... |