Automatic voltage internal resistance tester

The Automatic Voltage & Internal Resistance Tester is a precision electrical testing instrument designed for rapid and accurate measurement of open-circuit voltage (OCV) and DC internal resistance of lithium-ion battery cells. The system is widely used in battery research laboratories, production lines, and quality control departments to evaluate electrical consistency, detect defective cells, and support battery performance classification and safety assessment.

Application

(1) Open-circuit voltage (OCV) measurement of lithium-ion pouch cells, cylindrical cells, and coin cells.

(2) DC internal resistance measurement for cell grading and sorting.

(3) Incoming inspection of battery cells and semi-finished products.

(4) Quality control during cell production, formation, and aging processes.

(5) Electrical consistency evaluation for batch comparison and statistical analysis.

(6) Screening of abnormal cells with high internal resistance or voltage deviation.

(7) Laboratory research and process validation for battery development.

Standards

(1) IEC 62660-2 – Secondary Lithium-Ion Cells – Reliability and Abuse Testing (reference application).

(2) ASTM E83 – Verification of Linear Displacement Measuring Systems (instrument calibration reference).

(3) ISO 3611 – Verification of Linear Measuring Instruments (instrument calibration reference).

(4) RoHS – Restriction of Hazardous Substances Directive.

(5) CE Certification – Conformité Européenne.

Features

(1) Simultaneous measurement of voltage and internal resistance with high accuracy.

(2) Automated test sequence reduces operator influence and improves repeatability.

(3) Fast measurement suitable for production-line and batch testing.

(4) Wide compatibility with pouch, cylindrical, and coin-type battery cells.

(5) Stable contact design ensures reliable electrical connection.

(6) Built-in data storage and result classification functions.

(7) Digital communication interface for data export and system integration.

(8) Electrical safety protection with overload and short-circuit prevention.

Technical Parameters

ParameterSpecification
Voltage Measurement Range 0 – 10 V
Voltage Measurement Accuracy ±0.05% F.S
Internal Resistance Range 0.01 mΩ – 1,000 mΩ
Internal Resistance Accuracy ±0.5% F.S
Measurement Method DC pulse method
Measurement Time ≤ 1 s per cell
Data Storage ≥ 100,000 records
Communication Interface USB / RS232 / Ethernet (optional)
Power Supply AC 100–240 V, 50/60 Hz
Dimensions (W×D×H) 300 × 250 × 400 mm
Weight 15 kg (net) / 18 kg (gross)

Accessories

(1) Battery test fixture for pouch, cylindrical, and coin cells.

(2) Kelvin test leads and probe assemblies.

(3) Calibration resistor set for internal resistance verification.

(4) Data acquisition and analysis software.

(5) Power cables and grounding accessories.

(6) Operation and maintenance manual.

(7) Compliance and calibration certificates.

Test Procedure

(1) Power on the tester and allow system self-check.

(2) Perform calibration using certified voltage sources and standard resistors.

(3) Place the battery cell into the appropriate test fixture.

(4) Ensure correct polarity and stable electrical contact.

(5) Start the test sequence to measure voltage and internal resistance.

(6) Display, store, and export the test results as required.

Maintenance Information

(1) Keep test probes and fixtures clean to ensure stable contact resistance.

(2) Periodically verify measurement accuracy using standard resistors.

(3) Inspect cables and connectors for wear or damage.

(4) Avoid exposure to corrosive gases, electrolyte leakage, or high humidity.

(5) Store the instrument in a clean and temperature-controlled environment when not in use.


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Bending Tester
Bending Tester The Pure Bending Tester analyzes hand movements-referred to as "bending"-performed by artisans and professionals when judging a fabric's texture. T...
Building material flame retardancy test furnace
Building material flame retardancy test furnace Technical parameters Gas used: methane flow rate (35 ± 0.5)L/min, its purity is greater than 95%, air flow rate (17±0.2)L/min Speci...
Surgical Blade Sharpness Tester
Surgical Blade Sharpness Tester The Surgical Blade Sharpness Tester is a multi-functional testing device featuring a 7-inch LCD touchscreen display, suitable for testing the sharp...
Mechanical load testing machine for photovoltaic modules
Mechanical load testing machine for photovoltaic modules Standard IEC 61215: Terrestrial Photovoltaic (PV) Modules – Design Qualification and Type Approval IEC 61646: Thin-Film Terrestrial Photovo...
Reciprocating Wear Tester
Reciprocating Wear Tester Standards ASTM F2496: Scratch hardness and adhesion testing for prints/coatings. ISO 1518-1 & -2: Scratch resistance determination for paints...
供应产品

Same products

Air Permeability Tester STD- F8
Air Permeability Tester STD- F8 卖方: Standard International Group (HK) Limited The Air Permeability Tester is designed to quickly and accurately measure the breathability and p...
Surface Performance Tester STD-FB4 - A
Surface Performance Tester STD-FB4 - A 卖方: Standard International Group (HK) Limited Introduction TheSurface Tester analyzes finger movements-particularly, strokes-performed by arti...
Compression Tester STD - FB3 - A
Compression Tester STD - FB3 - A 卖方: Standard International Group (HK) Limited TheCompression Tester analyzes hand movements-particularly, pushing with a finger-performed by ar...
Single - Strand Bending Tester STD - FB2 - SH
Single - Strand Bending Tester STD - FB2 - SH 卖方: Standard International Group (HK) Limited Standards ASTM A370. ASTM A416. ISO6934.4 The Single Fiber Bending Tester makes it possible to ...
Bending Tester STD- FB2 - S
Bending Tester STD- FB2 - S 卖方: Standard International Group (HK) Limited Compliance Standards ASTM D790 (Plastics Flexural Properties) ISO 178 (Flexural Strength of Rig...