.

Humidifying Bottle Performance Tester

The Humidifying Bottle Performance Tester is used to test items such as leakage, pressure resistance, humidification output, and overflow of humidifying bottles.

Applications

The humidifier bottle performance tester is mainly used to detect various properties of humidifier bottles, including but not limited to leakage, pressure - resistance, humidification output, and overflow. This instrument can ensure the safety and effectiveness of humidifier bottles during use, and it is an important tool for the quality inspection of medical oxygen humidifiers.

Standards

YY/T 1610 - 2018:Anaesthetic and respiratory equipment - Oxygen humidifiers for medical use

Features

• Supports leakage, pressure resistance, humidification output, and overflow testing functions for humidification bottles.

• 7-inch color touchscreen control and intuitive interface significantly enhance the user experience.

• The device is equipped with a miniature printer for real-time printing of test results.

• Equipped with a high-precision pressure sensor, allowing for stable control of output pressure via a rotary pressure knob.

• Equipped with a high-precision flow meter for accurate measurement of flow rate through the tubing.

• Features a 20° tilt design device to accommodate observation at various angles for different specifications.

• Utilizes a universally designed conical test interface to meet the testing requirements of most products on the market.

• The device has a powerful data storage function, capable of storing up to 50.000 data entries.

• Miniature printer and universal USB data interface facilitate data output and transfer.

• Intelligent features such as overload protection and power-off memory effectively ensure user safety.

• Optional GMP "Computerized Systems" compliant testing software is available, featuring user management, permission management, and data audit trail functions.

Technical Parameters

ParameterSpecification
Working Range Usually between 0 and 1000KPa; some models may offer higher pressure options (e.g., 2000KPa or 2500KPa)
Flow Accuracy ±1%, ensuring the accuracy of testing
Power Supply Generally 220V, 5A, 50Hz
Instrument Operating Voltage Usually 220V
Air Source Requirements Adjustable from 0 to 2MPa; the air source needs to be provided by the user
Main Unit Dimensions Vary depending on the model; e.g., some models: 570×400×200mm (Length×Width×Height)
Other Functions Some advanced-model humidifier bottle performance testers may have features like micro-computer control, large touch-screen display, Chinese menu settings, digital preset test flow, and printing function (making operation more intuitive and convenient)

Accessoriess

Standard configuration: electronic control unit, testing unit, printer

Test Procedures

1. Pre-test Preparation

Check that the instrument's power supply (220V, 5A, 50Hz) is functioning correctly, and confirm that the external gas source (0-2MPa) is connected and adjustable;

Calibrate the flow sensor (ensuring an accuracy of ±1%), clean the instrument interfaces and tubing to ensure the test pathway is unobstructed.

2. Humidifier Bottle Sample Installation

Remove the humidifier bottle to be tested, check its appearance for damage, and connect the humidifier bottle to the instrument's test tubing according to the interface specifications;

Seal the connection between the humidifier bottle and the instrument, ensuring there is no risk of leakage at the interface (low-pressure gas can be pre-applied to initially verify the seal).

3. Test Parameter Settings

Select the test item (e.g., leak test, pressure test, humidification output test) via the instrument's touchscreen;

According to the YY/T 1610-2018 standard, set the corresponding test parameters: such as pressure test pressure, pressure holding time, test flow rate, etc., and save the parameters after confirmation.

4. Starting the Test

After confirming that the humidifier bottle connection and parameter settings are correct, click the "Start" button, and the instrument will automatically execute the corresponding test procedure (e.g., the leak test will monitor pressure changes, and the pressure test will stably maintain the set pressure);

During the test, the instrument will display pressure, flow rate, and other data in real time.

5. Test Process Monitoring

Observe whether the data on the instrument interface is stable. For leak tests, pay close attention to whether the pressure drops abnormally; for pressure tests, confirm that the humidifier bottle is not deformed or ruptured;

If abnormal data or equipment alarms occur, pause the test immediately and troubleshoot the problem (e.g., interface leakage, unstable gas source).

6. Test Completion and Cleanup

After the test is completed, the instrument will automatically shut down, and the test results will be recorded (e.g., whether the leakage is qualified, whether the pressure resistance value meets the standard);

Disconnect the gas source, loosen the humidifier bottle interface, remove the sample, clean the instrument tubing and interfaces, turn off the equipment power, and prepare the test report.


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Conveyor Belt Drum Friction Testing Machine
Conveyor Belt Drum Friction Testing Machine Main Purposes of Conveyor Belt Drum Friction Testing Machine The conveyor belt drum friction testing machine has extensive application value in mu...
Safety Helmet High Temperature Pre-Treatment Chamber
Safety Helmet High Temperature Pre-Treatment Chamber The safety helmet high-temperature pre-treatment chamber is manufactured according to the GB/T2812-2019 standard for safety helmets and is primaril...
Electric rolling wheel peeling testing machine
Electric rolling wheel peeling testing machine Electronic pressure roller testing machine (automatic pressure roller machine) specializing in intaglio printing process for the production of plas...
Room Corner Test Machine, ISO 9705. En 14390
Room Corner Test Machine, ISO 9705. En 14390 Introduction: According to ISO 9705 of building materials fire prevention performance standards, this machine control platform to provide independ...
Stapler Blade Sharpness Tester
Stapler Blade Sharpness Tester It features a 7-inch K600+ core/dot-matrix 65-color 800*480 resolution touchscreen display with English and Chinese menu options. Test parameters, ...
供应产品

Same products

Full-Body Garment Stress Tester,FZ/T 70015
Full-Body Garment Stress Tester,FZ/T 70015 卖方: Standard International Group (HK) Limited TheFull-Body Garment Stress Testeris a precision instrument designed to quantify the pressure exe...
Industrial Dry Cleaning Machine,industrial garment steamer
Industrial Dry Cleaning Machine,industrial garment steamer 卖方: Standard International Group (HK) Limited TheIndustrial Dry Cleaning and Energy-Saving Dryer Machinesare designed for high-performance, rel...
MST Medical Compression Stocking Tester,compression sock tester
MST Medical Compression Stocking Tester,compression sock tester 卖方: Standard International Group (HK) Limited The MST Medical Compression Stocking Tester is a professional instrument designed to evaluate the...
MMT Liquid Water Separation Tester,MMT Liquid Water Analysis Tool
MMT Liquid Water Separation Tester,MMT Liquid Water Analysis Tool 卖方: Standard International Group (HK) Limited Applications Textile Manufacturing:Testing waterproof fabrics for outdoor gear R&D:Developi...
Dental material color stability tester
Dental material color stability tester 卖方: Standard International Group (HK) Limited Application The applications of color stability testers for dental materials are mainly reflecte...