Retroreflectometer analyzer
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The system supports compliance with the following standards:
GB/T 28468-2012 – Method for Measuring Retro-Reflective Coefficient of School Safety Uniforms (Appendix A)
GB 20653-2020 – Method for Measuring Retro-Reflective Coefficient of High-Visibility Protective Clothing (Appendix C)
GB/T 18833-2012 – Road Traffic Reflective Film
GB/T 26377-2010 – Retro-Reflective Measurement Instruments
JT/T 612-2004 – Retro-Reflective Measurement Instruments
ASTM E808 – Standard Guide for Description of Retro-Reflective Materials
ASTM E809 – Standard Test Method for Measurement of Retro-Reflector Optical Performance
ASTM E810 – Standard Test Method for Retro-Reflective Coefficient of Sheeting Materials
DIN 67520 – Minimum Luminance Requirements for Traffic Safety Reflective Materials
Parameters
| Parameter | Specification |
|---|---|
| Measurement Range | 0–3000 cd/(lx·m²) |
| Measurement Accuracy | ±0.1 cd/(lx·m²) using microphotometric acquisition system |
| Observation Angle | 0°–2°, adjustable, ±0.002° |
| Incidence Angle | -45° to 45°, adjustable, ±1° |
| Light Source | Standard illuminant A, color temperature 2856 K, uniform vertical illuminance |
| Detector | Silicon photodiode with CIE-corrected spectral filter, V(λ) compliance |
| Sample Reference Center | Aperture subtended angle 12′, vertical illuminance uniformity ≤5% |
| Detector Movement | Vertically adjustable, observation angle 12′–120′ |
| Calibration Angles | Up to 12 angle pairs can be calibrated at once |
| Data Storage | Up to 10.000 test records |
| Screen | 4.5" touchscreen, 854 × 480 resolution |
| Storage Method | TF card |
| Sample Fixture | Quick-change structure for easy mounting |
| Calibration Plates | One black and one white calibration plate included |
| Power Supply | AC 220 V, 50 Hz, 500 W |
| Weight | Approx. 10 kg |
Features
High-precision retro-reflection measurement with microphotometric acquisition system
Adjustable observation and incidence angles for versatile testing conditions
Standardized illuminant and calibrated detector for accurate, repeatable results
Quick-change sample fixture allows efficient testing of multiple materials
Touchscreen interface for easy operation and data visualization
Built-in storage for up to 10.000 test records
Automatic calibration using included black and white plates
Test Procedures
Sample Preparation
Mount the reflective material or garment on the sample platform using the quick-change fixture.
Ensure the sample area corresponds to the testing aperture and standard requirements.
Calibration
Perform calibration using the included black and white plates to ensure measurement accuracy.
Measurement
Adjust the observation and incidence angles according to the testing standard.
Illuminate the sample with standard illuminant A and record the retro-reflective coefficient via the detector.
Repeat measurements for multiple points or samples as required.
Data Recording
Store test results on the TF card or view on the touchscreen interface.
Generate reports summarizing retro-reflection performance for quality assessment.
Maintenance Information
Regularly clean the detector and light source to maintain measurement accuracy
Inspect and tighten the sample fixture to prevent misalignment during tests
Calibrate the instrument periodically using the included plates
Ensure the touchscreen and data storage system function correctly
Store the instrument in a dry, dust-free environment to prevent sensor degradation
Send product request
Other supplier products
| Residual Ethylene Oxide Gas Chromatography System | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Residual Ethylene Oxide Gas Chromatography System is a professional analytical instrume... | |
| Laboratory Hot-Wire Thermal Conductivity Tester for Liquids | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The analyzer is designed and tested in accordance with internationally recognize... | |
| Absorbable Implant Plate Mechanical Strength Tester ASTM/ISO | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards (1)ASTM F2119: Standard test method for constant load testing of b... | |
| Rain Test Chamber | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Information: This instrument can simulate rain horizontally spraying to the test... | |
| Solar Simulator and Test Bench Manufacturer | IEC 60904-9 PV Testing System | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards (1) IEC 60904-9 – Photovoltaic Devices: Requirements for Sol... |
Same products
| HDPE High Speed Water Cooled Double Wall Corrugated Pipe Extrusion Line | Seller: Shandong HUASU Plastic Technology Co.,Ltd. | The molds adopt the double flow channel structure and nitriding polishing processed die head. Thi... | |
| PanEVAP & MVR system | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | A multiple-effect evaporator is a device that efficiently uses steam heat to evaporate water. Wat... | |
| PanGAP wide gap welded plate heat exchanger | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | PanGAP wide gap welded plate heat exchangerconsists of a welded plate pack with different channel... | |
| PanBloc welded plate heat exchanger | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | PanBloc welded plate heat exchangeris compact and efficient, although fully welded, is totally ac... | |
| PanBox welded plate heat exchanger | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | PanBox welded plate heat exchangerconsists of a laser welded plate pack with optimized corrugated... |