| No. | Item | Technical Specifications |
|---|---|---|
| 1 | Chamber Dimensions | External: 5.1 m (L) × 5.6 m (W) × 3.2 m (H) Internal: 4.9 m (L) × 5.4 m (W) × 2.6 m (H) Standard insulated door: 0.8 m × 2.0 m with double-glass observation window Internal lighting: 1 daylight lamp; power outlet: 2.5 kW Digital display panel for temperature and humidity Optional additional window available at extra cost |
| 2 | Temperature Range | 10–40°C ±0.5°C adjustable |
| 3 | Humidity Range | 35–90% ±5% adjustable |
| 4 | Air Change Rate | 3–16 ACH; optional single-pass fresh air supply available (installation-dependent, extra cost) |
| 5 | Electrical Control System | Centralized control panel; Supply: AC 380V/50Hz, max power ≤28.5 kW; Equipped with short-circuit and over-temperature protection |
| 6 | Airflow Organization Modes | 9 modes: Mixed Ventilation:(1) Top diffuser with side-lower return, (2) Side supply-side return, (3) Side supply-top return Displacement Ventilation:(1) Top return, (2) Side-top return Floor Ventilation:(1) Top return, (2) Side-top return Stratified Ventilation:(1) Top return, (2) Side-lower return Optional quick-connect supply fittings available (extra cost) |
| 7 | Working Environment | Standard laboratory conditions |
| 8 | Other Requirements | Refrigeration unit piping and electrical lines default 5 m; site preparation including leveling, wall openings, rain shelter, water supply, drainage, and main power line to control box to be provided by user; chamber panels: 100 mm PU double-color steel |
| 9 | Special Notes | Default specifications are factory standards; custom requirements exceeding these parameters require additional consultation and pricing |
Features
High-Precision Environmental Control: Maintains stable temperature and humidity with ±0.5°C and ±5% RH accuracy.
Multiple Airflow Modes: Supports mixed, displacement, floor, and stratified ventilation with nine programmable configurations.
Observation and Monitoring: Equipped with observation window, internal lighting, and digital panel for real-time monitoring of chamber conditions.
Customizable Ventilation: Air change rates adjustable; optional quick-connect fittings allow flexible supply modifications.
Safety and Reliability: Electrical protections including short-circuit and over-temperature safeguards.
Modular Installation Support: Refrigeration, water, drainage, and power line preparation allow smooth on-site assembly.
Accessories
Digital display panel for temperature and humidity monitoring.
Standard insulated chamber door with double-glass observation window.
Internal lighting (daylight lamp).
Optional quick-connect supply fittings for airflow customization.
Test Procedures
Chamber Preparation: Confirm power, water, and drainage connections; ensure refrigeration system is operational (~10 min).
Set Environmental Conditions: Program desired temperature (10–40°C), humidity (35–90%), and airflow mode via control panel or software (~5 min).
Load Samples/Subjects: Place equipment, test materials, or participants inside chamber; ensure proper positioning (~10 min).
Run Test: Activate test program; monitor environmental conditions in real-time; adjust airflow mode if needed (~variable duration).
Data Recording and Export: Record chamber temperature, humidity, and airflow data; save and export for analysis (~5 min).
Completion and Retrieval: End program; allow chamber to stabilize at safe conditions; remove samples or subjects safely (~10 min).
Maintenance Information
Regularly clean observation window, internal lighting, and control panel.
Inspect refrigeration unit, piping, and electrical connections for wear or damage.
Verify airflow systems and sensors for calibration and accuracy.
Maintain chamber in a clean, dust-free, and temperature-stable environment.
Send product request
Other supplier products
| Automatic sauce packet pressure apparatus | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications The main purpose of the pressure resistance tester for sauce packets is to ev... | |
| Compression Tester STD-FB3-A,ASTM D903 compression resistance verification machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Compression Tester STD-FB3-A is a professional instrument designed for quantitative eva... | |
| Battery Isothermal Calorimeter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Description Battery Isothermal Calorimeter is a professional instrument developed ... | |
| Image test standard transmission light box | Standard CIE 15:2004: Adheres to colorimetry standards for spectral measurement. ISO 3664:2009: Compliant with viewing conditions for graphic art... | |
| Alternative to BBA steady state solar simulator tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The BBA Steady-State Solar Simulator is applicable in multiple scenarios: Fac... |
Same products
| Rubber Mechanical Properties Tester | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The Rubber Sleeve Rebound Test... | |
| Packaging Film Shrinkage Tester | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM D2732: Standard Test Meth... | |
| AI-powered spencer impact tester instrument | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM D3420:Standard Test Metho... | |
| G506 Filter Performance Tester | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM F2100:Standard Specificat... | |
| Automatic dual flat-panel thermal conductivity detector | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 10294-2008:Determination ... |