Automatic catheter strength instrument
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The instrument is compliant with the following standard:
(1) YY 0325-2016 – Single-use Sterile Urinary Catheters, specifying requirements for mechanical and physical strength testing
(2) ISO 10555-1:2013 – Intravascular catheters — Sterile single-use catheters — Part 1: General requirements
(3) ASTM F2081-15 – Standard Test Method for Tensile Strength of Catheter Tubing
(4) EN ISO 10555-2 – Sterile single-use urinary catheters — Part 2: Materials, tensile strength, and elongation testing
Features
(1) High-precision testing: load measurement from 0.01 to 120 N with ±0.001 N accuracy
(2) Adjustable loading speed: 0–500 mm/min ±0.5 mm/min
(3) Multi-language interface: Simplified Chinese and English
(4) Integrated data management: stores up to 100 test records per cycle including deviation values
(5) Real-time monitoring: movement control (up/down), timing, and calibration system
(6) High-resolution touchscreen: 7-inch K600+ core, 800×480 resolution, 65k colors
(7) Built-in thermal printer for immediate test report generation
(8) Compact and lightweight: net weight 15 kg, easy to move and operate
Technical Parameters
| Parameter | Specification |
|---|---|
| Interface Language | Simplified Chinese / English |
| Nominal Specifications | Supports all catheter sizes |
| Load Range | 0.01–120 N |
| Accuracy | ±0.001 N |
| Loading Speed | 0–500 mm/min ±0.5 mm/min |
| Printing System | Stores 100 test records per cycle (data + deviation) |
| Net Weight | 15 kg |
| Dimensions (L×W×H) | 430 × 290 × 250 mm |
| Power Supply | AC 110–220 V, 50 Hz (auto-adaptive) |
| Power Consumption | 70 W |
| Control System | PLC + ARM programmable controller |
| Touchscreen | 7-inch K600+ core, 800×480 resolution, 65k colors |
| Sensor | High-precision force transducer |
| Force Measurement | Built-in load cell system |
| Printing Device | Onboard thermal printer |
Accessories
(1) Standard catheter clamps and fixtures
(2) Power cable
(3) Touchscreen interface manual
(4) Thermal printer paper rolls
(5) Calibration certificate
Test Procedures
(1) Mount the catheter sample in the provided clamps
(2) Set the nominal specification and span for the test
(3) Configure the load range and desired loading speed on the touchscreen
(4) Start the automated test, controlling movement up and down as needed
(5) Monitor load, deformation, and timing in real time
(6) Upon test completion, review results on the display
(7) Print test report using the onboard thermal printer
(8) Repeat for additional samples as required
Maintenance Information
(1) Keep the instrument clean and free from debris after each test
(2) Inspect clamps, sensors, and load cells regularly for wear or damage
(3) Calibrate the force transducer periodically to maintain measurement accuracy
(4) Avoid exposure to liquids and extreme temperatures
(5) Ensure proper grounding of the power supply to protect electronic components
(6) Maintain logs of calibration, testing, and maintenance for compliance purposes
Send product request
Other supplier products
| Melt Mass-Flow Rate Test Machine, ASTM D1238-98 | Whats/App: +86 156 1874 6768 Web: standard-groups.com Introduction: This machine is deisgned according to ISO1133. ASTM D1238-98 ,ISO1133; Melt ... | |
| Multi-function drum and reciprocating tribology tester factory | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application This testing machine is suitable for friction and wear evaluation under both l... | |
| Particle Strength Test machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Particle Strength Tester is a high-precision instrument designed to measure the mechani... | |
| High frequency reciprocating tribometer | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Friction and wear testing of metallic and non-metallic materials. Evaluation ... | |
| AATCC Light Fastness Tester Xenon Arc Si220 Factory Supplier | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard Compliance: AATCC TM16-2004:Colorfastness to Light AATCC TM16.3:Co... |