Automatic catheter strength instrument

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

The instrument is compliant with the following standard:

(1) YY 0325-2016 – Single-use Sterile Urinary Catheters, specifying requirements for mechanical and physical strength testing

(2) ISO 10555-1:2013 – Intravascular catheters — Sterile single-use catheters — Part 1: General requirements

(3) ASTM F2081-15 – Standard Test Method for Tensile Strength of Catheter Tubing

(4) EN ISO 10555-2 – Sterile single-use urinary catheters — Part 2: Materials, tensile strength, and elongation testing

Features

(1) High-precision testing: load measurement from 0.01 to 120 N with ±0.001 N accuracy

(2) Adjustable loading speed: 0–500 mm/min ±0.5 mm/min

(3) Multi-language interface: Simplified Chinese and English

(4) Integrated data management: stores up to 100 test records per cycle including deviation values

(5) Real-time monitoring: movement control (up/down), timing, and calibration system

(6) High-resolution touchscreen: 7-inch K600+ core, 800×480 resolution, 65k colors

(7) Built-in thermal printer for immediate test report generation

(8) Compact and lightweight: net weight 15 kg, easy to move and operate

Technical Parameters

ParameterSpecification
Interface Language Simplified Chinese / English
Nominal Specifications Supports all catheter sizes
Load Range 0.01–120 N
Accuracy ±0.001 N
Loading Speed 0–500 mm/min ±0.5 mm/min
Printing System Stores 100 test records per cycle (data + deviation)
Net Weight 15 kg
Dimensions (L×W×H) 430 × 290 × 250 mm
Power Supply AC 110–220 V, 50 Hz (auto-adaptive)
Power Consumption 70 W
Control System PLC + ARM programmable controller
Touchscreen 7-inch K600+ core, 800×480 resolution, 65k colors
Sensor High-precision force transducer
Force Measurement Built-in load cell system
Printing Device Onboard thermal printer

Accessories

(1) Standard catheter clamps and fixtures

(2) Power cable

(3) Touchscreen interface manual

(4) Thermal printer paper rolls

(5) Calibration certificate

Test Procedures

(1) Mount the catheter sample in the provided clamps

(2) Set the nominal specification and span for the test

(3) Configure the load range and desired loading speed on the touchscreen

(4) Start the automated test, controlling movement up and down as needed

(5) Monitor load, deformation, and timing in real time

(6) Upon test completion, review results on the display

(7) Print test report using the onboard thermal printer

(8) Repeat for additional samples as required

Maintenance Information

(1) Keep the instrument clean and free from debris after each test

(2) Inspect clamps, sensors, and load cells regularly for wear or damage

(3) Calibrate the force transducer periodically to maintain measurement accuracy

(4) Avoid exposure to liquids and extreme temperatures

(5) Ensure proper grounding of the power supply to protect electronic components

(6) Maintain logs of calibration, testing, and maintenance for compliance purposes


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Automatic Coating Adhesion Scratch Tester | ISO 1518 Scratch Tester
Automatic Coating Adhesion Scratch Tester | ISO 1518 Scratch Tester Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Automatic Coating Adhesion Scratch Tester is a high-precision instrument ...
Nano Indenter,Nanoindentation tester
Nano Indenter,Nanoindentation tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Nano Indenter / Scratch Tester is a comprehensive micro- and nano-mechanical testing sy...
Rotary reciprocating friction and wear tester
Rotary reciprocating friction and wear tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Reciprocating-Rotary Friction and Wear Tester (Type 1) is designed for evaluating the t...
Laboratory 3D Surface Topography Profiler
Laboratory 3D Surface Topography Profiler Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system’s analysis capabilities and software design support surface met...
Contact Coolness Tester STD-QM
Contact Coolness Tester STD-QM Whats/App: +86 156 1874 6768 Web: standard-groups.com Test standard: JIS L 1927 Textiles-Measurement method of cool touch feeling property GB/T 3...
供应产品

Same products

MMA-120 Handheld Arc Welder
MMA-120 Handheld Arc Welder 卖方: 871183 The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con...
SHX Series Circulating Fluidized Bed Steam Boiler
SHX Series Circulating Fluidized Bed Steam Boiler 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi...
YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler
YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ...
DHL Series Coal-Fired Steam Boilers
DHL Series Coal-Fired Steam Boilers 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b...
SZL Series Biomass Steam Boiler
SZL Series Biomass Steam Boiler 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi...