Intramedullary Nails Internal-fixation Pull-out Force Tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
The Intramedullary Nails Internal-fixation Pull-out Force Tester is a specialized device designed to measure the axial pull-out strength of intramedullary nails and internal fixation screws. It evaluates the mechanical anchoring performance of implants in simulated bone or test materials, ensuring their stability and reliability under clinical loading conditions. This tester is widely used in orthopedic research, implant development, and quality control to verify that internal fixation devices can resist dislodgement and provide safe, long-term fixation in patients.
Standard
(1)ASTM F543: Standard specification for metallic medical bone screws (dynamic loading protocols).
(2)ISO 12172: Implants for surgery—Partial and total hip joint prostheses (fatigue testing methodologies).
(3)YY/T 0342: Chinese standard for surgical implant testing (cutout resistance evaluation).
Adaptive Compliance: Customizable protocols for emerging regulatory frameworks.
Product Advantages
(1)Dynamic Load Simulation: Replicates physiological gait-cycle forces (0–1.500 N at 5 Hz) for real-world failure prediction.
(2)Dual-Axis Measurement: Simultaneous tracking of cutout force (N) and angular displacement (°) during cyclic testing.
(3)Failure Prediction Algorithm: Proprietary software identifies pre-failure indicators using machine learning.
(4)High-Speed Imaging Integration: Syncs with external cameras for real-time deformation visualization (optional upgrade).
(5)Ergonomic Workflow: Touchscreen interface with pre-loaded ASTM/ISO templates and customizable test sequences.
(6)Modular Design: Quick-swap fixtures for testing various nail designs (helical blades, lag screws, etc.).
Product Features
(1)Electrodynamic Actuator: Delivers sinusoidal, triangular, or custom waveforms up to 10 Hz.
(2)6-DoF Load Cell: Measures axial/torsional forces and moments with 0.1% FS accuracy.
(3)High-Resolution Motion System: ±5° angular resolution for tracking nail migration paths.
(4)Thermal Stability Chamber: Controls test environment (-20°C to +50°C) for material property analysis.
(5)AI-Powered Analytics: Auto-generates failure mode maps comparing historical/current implant designs.
(6)Remote Access: Cloud-connected system for real-time collaboration and audit trails.
Technical Parameters
| Parameter | Specification |
|---|---|
| Max. Dynamic Load | ±1.500 N (axial) / ±50 Nm (torsional) |
| Frequency Range | 0.1–10 Hz (sinusoidal) |
| Angle Measurement | ±5° (0.001° resolution) |
| Data Sampling Rate | 10 kHz (synchronized across 8 channels) |
| Test Duration | 1–10⁶ cycles (programmable) |
| Sample Compatibility | 8–16 mm diameter nails, 100–500 mm length |
| Power Requirements | 220V/110V ±10%, 50/60Hz, 1.200W |
| Dimensions (W×D×H) | 1.200 × 800 × 1.800 mm |
| Weight | 450 kg (including safety enclosure) |
Send product request
Other supplier products
| Retroreflectometer analyzer | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system supports compliance with the following standards: GB/T 28468-2012 &n... | |
| Medical Face Mask Synthetic Blood Penetration Test machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applicable Standards: GB 19083-2010 YY/T 0691-2008 YY 0469-2011 ISO 22609 ASTM F1862-0... | |
| High speed reciprocating friction wear tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications You can use this tester for: High-temperature friction and wear testing of m... | |
| Wheel Load Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard EN 1569: Determines rolling load resistance of synthetic sports surfaces (e.g., P... | |
| IEC 62716 Ammonia Corrosion Test Chamber Manufacturer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards IEC 62716: Photovoltaic (PV) modules—Ammonia corrosion testi... |
Same products
| MMA-120 Handheld Arc Welder | Seller: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... | |
| SHX Series Circulating Fluidized Bed Steam Boiler | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi... | |
| YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ... | |
| DHL Series Coal-Fired Steam Boilers | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b... | |
| SZL Series Biomass Steam Boiler | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi... |