ISO 22612 Dry microbial penetration resistance tester
The resistance to dry microbial penetration test (ISO 22612) evaluates resistance characteristics of protective equipment to dry bacteria-carrying particles.Testing is most often performed on equipment such as surgical gowns, drapes, and clean air suits when used as medical devices. This test is part of a test suite required by the EN 13795 guidance, which determines performance requirements and performance levels for gowns, drapes, and clean air suits when used as medical devices. It is also included in the test suite required by EN 14126, for personal protective equipment. Testing in compliance to EN 13795 is part of following the European Union Medical Device Directive 93/42/EEC, and is important when seeking a CE mark for products marketed in Europe. Nelson Laboratories personnel are experts in barrier testing, and we offer a one-stop shop for barrier tests required for marketing gowns and drapes in both the US and Europe.
Standards:
ISO 22612; YY / T 0506.5-2009
Test principle:
When testing, the samples were seperately fixed on a container. In these containers, five containers carrying the bacillus subtilis and 1 container carrying uncontaminated talc powder used as control. Insert 1 petri dish at the bottom of each container and beneathe the bottom of the specimen.
The supporting vessel's equipment is oscillated by a gas ball type oscillator, and all of the talcum powder passing through the sample falls onto the petri dish. Remove the petri dish for culture. The experiment results were evaluated by counting the colonies.
Scope of application:
Dry microbial penetration resistance performance tester is professionaly used to determine the bacterial penetration resistance performance of materials to dry particulates in the size range of human body dandruff. Whenn testing, the samples were fixed on a container respectively. In these containers, five containers containing subtilisin talc, one container added with Not stained talcum powder was used as a control. A Petri dish was inserted at the bottom of each container at a distance from the bottom of the timer.
The equipment supporting the container is oscillated by a Gas ball oscillator, and all the talc penetrating the sample Falled on the culture dish , cultured without the culture vessel. The test results were evaluated by counting the growing colonies.
Main parameters:
Vibration form: pneumatic ball vibrator
Vibration frequency: 20800 times / min
Vibration force: 650N
Worktable size: 40cm × 40cm × 10mm
Workbench media: marble plate
Experimental container: stainless steel test container
Work stations: 6
Cabinet negative pressure range: -50 Pa ~ -200 Pa
High efficiency filter filtration efficiency: better than 99.99%
Negative pressure cabinet ventilation flow: ≥ 5m3 / min
Data storage capacity: 100,000 group
Working voltage: 220V 50HZ
Features:
Reliable design, stable test;
Front switch glass doors, easy to observe the operation of the experimental staff;
Removable bracket, height adjustable bracket;
Support, mobile dual casters;
Negative pressure experimental system, with the fan exhaust system and the introduction of high efficiency filter to ensure the safety of operators;
Embedded high - speed industrial microcomputer control;
Dedicated operating software, with software parameters calibration, fault detection automatically
Other supplier products
|
Thermal Radiation Droplet Tester |
Application: Thermal Radiation Droplet Tester, to determine the resistance to the thermal radiation through up a radiant heat source placed on the ... |
|
Rain Tester and test machine |
Testing Standards:AATCC 35, BS EN ISO 22958Applications:Rain Tester is designed for determining the water-repellent property of fabrics or composit... |
|
Surgical face masks bacterial Filtration Efficiency (BFE)Tester |
Standards EN 14683 Medical face masks-Requirements and test methods. ASTM F2101 Standard Test Method for Evaluating the Bacterial Filtration Effici... |
|
Radiant Protective Performance Test Device |
Standardized sample clamping program, strictly meet the standards. While taking into account the operation, is simple and generous. The calorimeter... |
|
Perspiration Tester |
Perspiration Tester
Testing Standards:
ISO 105 E01/E02/E04, BS 1006, AATCC 15/106/107, DIN 54005, M&S C6/C7, NEXT TM4, DIN EN ISO 105 E01/E02/E... |
All supplier products
Same products