Spray Aerosol Ignition Distance Test machine
Conforms to UN Recommendations on the Transportation of Dangerous Goods - Manual of Tests and Criteria 31.4 Ignition Distance Test for Spray Aerosols and GB/T 21630-2008 Test Method for Ignition Distance for Dangerous Goods Spray Aerosols.
Standards
GB/T 21630-2008
GAT536
NY/T 1860
Product Features
1. It adopts an embedded processor and the Windows CE operating system.
2. With an 8 - inch liquid - crystal touch screen, it can display the test status in real - time.
3. It integrates two working modes: the "standard mode" and the "custom mode".
4. The aerosol sprays automatically, and the aerosol can clamping device can automatically make the can stand upright or turn it upside - down.
5. The nanosecond - level photosensitive sensor detection module automatically detects whether the spray is burning.
6. Driven by a hybrid stepping motor and equipped with an optical encoder, it enables precise positioning of the measurement position.
7. It uses a spark lighter as the ignition source, producing a blue - colored non - glowing flame with a height of 4 - 5 cm, and the flame height is adjustable.
8. The test process is fully automated, requiring no manual intervention. It supports functions such as parameter setting, data storage, and report viewing.
9. The dedicated fire - proof accordion cover can effectively prevent equipment contamination.
10. It is equipped with a temperature and humidity module to monitor the temperature and humidity of the test environment in real - time.
Technical parameters
|
Operating Environment |
(5~40)℃, <85%RH |
|
Ambient Temperature Measurement Range |
(-10~+60)℃ |
|
Ambient Temperature Measurement Accuracy |
±0.5℃ |
|
Ambient Humidity Measurement Range |
(5~95)%RH |
|
Ambient Humidity Measurement Accuracy |
±5.0%RH |
|
Flame Detection Response Time |
≤6ms |
|
Flame Detection Distance |
(15~90)cm |
|
Positioning Accuracy |
±1mm |
|
Minimum Pressing Duration |
5s |
|
Ignition Method |
Gas Igniter |
|
Aerosol Tank Positioning |
Aerosol canisters can be turned over automatically |
|
Compatible Spray Tank Dimensions |
Height (120~400)mm, Diameter (30~80)mm |
Send product request
Other supplier products
| Spring torsion tester | Main uses: Spring torsion testing machine is suitable for all kinds of large spring torsion Angle, torsion moment test, add the corresponding acce... | |
| Solids Spontaneously Igniting Tester | Product Description Conforms to the UN Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria 33.3.1.4 Test N.2: Test ... | |
| Liquid Sustained Combustion Tester | Product Introduction This product complies with "Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria" 32.5.2 Test L... | |
| Respirator Fit meter | PortaCount Respirator Fit Tester: Qinsun assesses the overall integrity of masks by performing an aerosol leak test utilizing the PortaCount Respi... | |
| Rupture strength tester | Standard IEC 61730-2: Photovoltaic (PV) Module Safety Qualification – Requirements for Testing UL 61730: Safety Standard for Photovoltaic... |
Same products
| FH-77 Dissolved hydrogen meter | Seller: Kelilong Electron Co.,Ltd | Pen type dissolved hydrogen detector / dissolved hydrogen meter: used to measure the dissolved hy... | |
| Xenon Arc Weather Ometer | Seller: Standard International Group (HK) Limited | Get a quote: Whats/App: +86 181 0160 0864 SI4000Xenon Arc Weather Ometer utilizes a xenon lamp ... | |
| Xenotest | Light Fastness Tester Si220 (SunTest) | Seller: Standard International Group (HK) Limited | Get a quote: Whats/App:+86 180 0160 0864 A Light Fastness Tester, or Xenon Tester, Accelerated ... | |
| Ozone Aging Chamber | Seller: Standard International Group (HK) Limited | Get a quote: Whats/App:+86 156 1874 6768 The Ozone Aging Chamber is a precision testing instrum... | |
| Plastic Smoke Density Tester F313 | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com The F313 Plastic Smoke Density Tester is ... |