Point-to-point Resistance Tester EN 1149?1?1995 GB/T 1410?2006

Standards

GB12014-2009. GB/T 1410-2006. ASTM D257-99. GB/T 10581?2006. GB/T 1692?2008. GB/T 2439?2001. GB/T 12703.4?2010. GB/T 10064?2006. GB/T 22042?2008. EN 1149?1?1995. GB/T 1410?2006.

Introduction

The point-to-point resistance tester is used to detect the point-to-point resistance rate of anti-static work clothes, which is suitable for class A and b-class anti-static work clothes. The principle is to add test voltage 100 V to test samples for total 15 s. The weak current flowing through the sample is sampled and amplified by the standard resistance and read on the high impedance meter. The figures show the resistance value directly, with high accuracy, quick display, good stability and easy reading.

Operation Principle

According to ohm's law, the tested resistance Rx is equal to the applied voltage V divided by the current I. The traditional high impedance meter works by measuring the fixed voltage V, and the resistance value is get by measuring the current I that flows through the sampling resistor. As can be seen from the ohm's law, since the current (I) is inversely proportional to the resistance rather than directly proportional, so the display value of resistance is nonlinear. Namely when the resistance is infinite, the current is zero, and the zero level at the header is ∞. The scale interval is very close with a very low resolution, so the whole scale is nonlinear. The voltage V also changes due to the measurement of different resistance, so the normal high resistance meter is low in accuracy with low resolution.

The point-to-point resistance tester measures the voltage V at the two ends of the resistance and the current I flowing through the resistor at the same time. Through the large scale integrated internal circuit, the calculation of voltage divided by the current is completed. Then, the results obtained through A/D conversion shows digital resistance value. Even though the voltage V at the two ends of the resistance and the current I flowing through the resistor change at the same time, the displayed resistance value will not change like the ordinary high resistance meter subject to the change of the tested voltage V and the current I. So, even if the measured voltage, the measured resistance and power voltage change, it has a little influence on the results. The accuracy is very high (patent), and theoretically the error can be zero, and the actual error can achieve parts per thousand or parts per ten thousand.

Technical Parameters

Items Parameters
Range limit 105Ω-1013Ω
High resistance meter indication error ±1%
Voltage 100V DC
Display mode Digital LCD
Operation environment Temperature: 0-40℃; Relative humidity: 80%
Power supply AC 220V 50Hz, power consumption about 5W
Dimension 252mm*265mm*100mm
Net weight About 5W

Отправить запрос, связаться с поставщиком

Кому: Шанхайская компания Qianshi Precision Electromechanical Technology Co., Ltd
Ваш E-mail:
Текст письма:


Send to other suppliers

Другие товары поставщика

Fat End Type of Thread EndSuction Machine
Fat End Type of Thread EndSuction Machine lt is applicable to clean up large-sizegarment, such as: surgical gown, robeetc.,thread end and fibre dust cutdown from carpet and rugs (no statice...
G678W-L cold resistance testing machine for shoes
G678W-L cold resistance testing machine for shoes Shoes cold resistance testing machine is used to test finished shoes, rubber, sole, synthetic leather, plastic and other test requirements under va...
Standard Helmet Visual Field Tester
Standard Helmet Visual Field Tester Standard EN 1078:2023: Helmets for Pedal Cyclists, Skateboarders, and Roller Skaters CE Certification(Conformité Européenne) Equiv...
Mask Visual Field Tester GB 2890-2009 GB 2626-2019
Mask Visual Field Tester GB 2890-2009 GB 2626-2019 Standards: GB 2890-2009 Respiratory protection Self-priming filter gas mask 6.8 Field of vision GB 2626-2019 Respiratory protective products Self...
Mask Respirator Breathing Resistance Tester
Mask Respirator Breathing Resistance Tester Standards: EN149-2009 section 7.15, 8.3.4, EN 13274-3, GB 19083, GB 2626, NIOSH 42 CFR Part 84, JIS/T 8151 Applications: Mask & Respir...
Все товары поставщика

Похожие товары

Plastic Smoke Density Tester F313
Plastic Smoke Density Tester F313 Продавец: Standard International Group (HK) Limited The F313 Plastic Smoke Density Tester is a precision instrument designed to measure the specific ...
Xenon Test Chamber
Xenon Test Chamber Продавец: Standard International Group (HK) Limited Features and Benefits 1.The use of xenon arc lamp as a light source, can simulate the full sunli...
Single Cable Vertical Flame Tester
Single Cable Vertical Flame Tester Продавец: Qinsun Instruments Co., LTD Introduction Single Cable Vertical Flame Tester is used for the vertical flame propagation of a ...
Safety Helmet Flammability Tester
Safety Helmet Flammability Tester Продавец: Qinsun Instruments Co., LTD Product Introduction: Safety hat flame retardant performance tester for safety hat flame retarda...
Furniture flammability tester
Furniture flammability tester Продавец: Qinsun Instruments Co., LTD Product details Furniture flammability tester is mainly used to test the combustion performance ...