GF6018A CLAMP TYPE MULTIMETER CALIBRATOR
This GF6018A high precision multi-product calibrator is compliance with national verification regulation: JJG124-2005 "Ammeter, voltmeter, power meter and resistance meter verification rules" and related national standards. 0.05% single-phase standard AC/DC voltage and current source, can verify level 0.2 and the following AC/DC voltmeter, ammeter, frequency meter, resistance meter & clamp meter. The source signals by using DSP and 16-bit high-speed A/D converters of controllable sine wave, the distorted wave signal source. Multifunction calibrator GF6018A is designed as universal calibration tool for electrical calibration laboratories.
The GF6018A clamp type multimeter calibrator with RS-232 port, it can be connected with PC to control, become automatic verification system. It has precise interface and multi-functions including verification, storage and query. The GF6018A multimeter calibrator adopts color LCD screen, chart character display is clear, with high precision, stable and reliable, convenient operation and flexible characteristics. It fully meets ISO17025 laboratory standards and is a good ideal test equipment for electrical engineers.
PARAMETERS
| Electrical parameters | ||
|---|---|---|
| Power supply | Single phase AC 220V±10% or 110V±10%, 50/60HzAC | |
| AC Voltage output | ||
| range | 200mV, 400mV, 2000mV, 4000mV, 20V, 100V, 200V, 500V, 1000V; max 1050V |
|
| Range 20V-1000V | Accuracy | ±(0.03%RD+0.02%FS) |
| Stability | ≤0.01% FS /60s | |
| Range 200mV-4000mV | Accuracy | ±(0.3%RD+0.2%FS) |
| Stability | ≤0.04% FS /60s | |
| Load capacity | 20VA | |
| Waveform distortion | ≤0.3% | |
| Adjustable Range | 0-120% FS (1000V not included) | |
| Fineness | 5×10-5 | |
| AC Current Output | ||
| Range | 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A; max 24A | |
| Range 0.2A-20A | Accuracy | ±(0.03%RD+0.02%FS) |
| Stability | ≤0.01% FS /60s | |
| Range 2mA-40mA | Accuracy | ±(0.3%RD+0.2%FS) |
| Stability | ≤0.04% FS /60s | |
| Load capacity | 20VA | |
| Waveform distortion | ≤0.3% | |
| Adjustable Range | 0-120% FS | |
| Fineness | 5×10-5 | |
| Frequency Output | ||
| Range | 45-65Hz | |
| Fineness | 0.001Hz | |
| Accuracy | 0.002Hz | |
| Harmonic Wave | ||
| Times | 2-31 | |
| Extent | 0-20% | |
| Each harmonic phase fineness | 0.01·N (N is harmonic order) | |
| DC voltage output | ||
| Range | 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V, 200V, 400V, 1000V; max 1050V |
|
| Load capacity | 20W | |
| Adjustable range | 0-110% FS (1000V not included) | |
| Adjustable fineness | 5×10-5 | |
| Stability | ≤0.01% FS /60s (peak-peak value) | |
| Accuracy | ±(0.03%RD+0.02%FS) | |
| Ripple wave | ≤1% FS | |
| DC current output | ||
| Range | 20uA, 40uA, 200uA, 400uA, 2mA, 4mA, 20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A; max 22A |
|
| Load capacity | 20W | |
| Adjustable Range | 0-110% FS (1000V not included) | |
| Adjustable Fineness | 5×10-5 | |
| Stability | ≤0.01% FS /60s (peak-peak value) | |
| Accuracy | ±(0.06%RD+0.04%FS) | |
| Ripple wave | ≤1% FS | |
| Resistance box | ||
| Range | 10Ω, 20Ω, 50Ω, 100Ω, 200Ω, 500Ω, 1kΩ, 2kΩ, 5kΩ, 10 kΩ, 20kΩ, 50KΩ, 100kΩ, 200kΩ, 500kΩ, 1MΩ, 2MΩ, 5MΩ, 10MΩ, 20MΩ | |
| Fineness | 0.02% | |
| Adjustable Range | 0-100%FS | |
| Accuracy | ±(0.3%RD+0.2%FS) | |
| Clamp meter range | ||
| 200A | Accuracy | ±0.2% |
| 600A | Accuracy | ±0.5% |
| 1000A | Accuracy | ±1% |
| Dot frequency square wave output | ||
| Range | 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz | |
| Square wave amplitude | 4.5V | |
| Setting accuracy | 5×10-5 | |
| Communication Port | ||
| Communication Port | RS232, USB | |
| Standard | ||
| Standard | JJG124-2005; JJF1587-2016; IEC61010, IEC 61000, IEC 61326 | |
| Safety | ||
| Isolation protection | IEC 61010-1:2001 | |
| Measurement Category | 300 V CAT III, 600 V CAT II | |
| Degree of protection | IP20 | |
| Declaration of conformity | CE, CNAS certified | |
| Mechanical parameters | ||
| Dimension (L×W×H) (mm) | 440x360x160 | |
| Weight (kg) | 12 | |
| Environmental conditions | ||
| Operating temperature | 20°C±10°C | |
| Storage conditions | -30°C to 60°C | |
| Relative humidity | ≤85%RH | |
| (1) FS means range, the same as below; (2) RD means the setted harmonic content, harmonic can be a single output, also multiple output. |
||
Send product request
Other supplier products
| FQ-RCT02:6000A flexible ROGOWSK coil current sensor | The FQ-RCT02 flexible current sensor is an AC current sensor composed of a flexible Rogowski coil. It is suitable for large current measuring. They... | |
| GF6018A CLAMP TYPE MULTIMETER CALIBRATOR | The GF6018A clamp type multimeter calibrator with RS-232 port, it can be connected with PC to control, become automatic verification system. It has... | |
| GF1000 MULTI-POSITIOS 1-PH ENERGY METER TEST BENCH | The GF1000 series multi-positions single phase kWh meter test bench can measure mechanical meters, electronic mechanical meters and electricity met... | |
| FQ-RCTA02: ROGOWSKI coil 3000A flexible current sensor with integrator | The FQ-RCTA02 rogowski coil flexible current sensor is designed for easy installed in tight spaces and without the need for dismantling the primary... | |
| XQ13: high performance 100A AC AMP current probe clamp | The model XQ13 100A current probe clamp is suitable for 0A-100A cable online measurement, it belongs to high performance transducer field. Frequenc... |
Same products
| Vacuum Conveying System. | Seller: Greentech International (Zhangqiu) Co., Ltd | This is a material pressure pneumatic system, which adopt Roots blower as generator, rotary feede... | |
| HD Industrial Videoscope | Seller: 869233 | HD Industrial Videoscope Coantec C60 can be widely used in aerospace, energy & power, pe... | |
| Variable Angle LCP Two-Column Volar Distal Radius Plate | Seller: ATOM Medical Devices Co., Ltd | The Variable-Angle Two-Column Volar Distal Radius Locking Plateis designed for the anatomical rec... | |
| Pneumatic Conveying System/PVC Powder Batch Dosing Mixing System. | Seller: Greentech Industry (Jinan) Co., Ltd | This is a material pressure pneumatic system, which adopt Roots blower as generator, rotary feede... | |
| Truck Paint Booth With Side Draft | Seller: Hwabu Tech Co., Ltd. | Description: Side draft truck paint booth for large vehicle painting and drying, designed with ... |