ACREL MEDICAL IT ISOLATED POWER SYSTEM
Acrel medical isolated power supplysystems are mainly used in operating rooms, ICU/CCU intensive care units, premature infant care units, delivery rooms, hemodialysis centres, and emergency rooms. The hospital isolated power systemsprovide safe, reliable and continuous power distribution for important equipment in these places. They can realize the conditions of insulation, load and isolation transformer temperature of IT system. Real-time monitoring with functions such as system insulation fault circuit location and centralized monitoring of multiple systems.
Medical Isolated Power Supply System For Sale
Hospital isolated power systems provide safety, continuous and reliable power supply, real-time detection of the insulation state to the ground to ensure the life safety of patients. Insulation fault location, convenient maintenance.
Hospital Isolated Power Systems Data Sheet
|
Technical Parameters of AITR Series of Medical Isolation Transformer |
|||||
|
Type |
AITR10000 |
AITR8000 |
AITR6300 |
AITR5000 |
AITR3150 |
|
Insulation class |
H |
H |
H |
H |
H |
|
Protection class |
IP00 |
IP00 |
IP00 |
IP00 |
IP00 |
|
Power/voltage/ current |
|||||
|
Rated power |
10000VA |
8000VA |
6300VA |
5000VA |
3150VA |
|
Rated frequency |
50/60Hz |
50/60Hz |
50/60Hz |
50/60Hz |
50/60Hz |
|
Rated input voltage |
AC230V |
AC230V |
AC230V |
AC230V |
AC230V |
|
Rated input current |
36A |
||||
|
Rated output voltage |
AC230V/115V |
AC230V/115V |
AC230V/115V |
AC230V/115V |
AC230V/115V |
|
Rated output current |
|||||
|
Inrush current |
<12In |
<12In |
<12In |
<12In |
<12In |
|
Leakage current |
<200μA |
<200μA |
<200μA |
<200μA |
<200μA |
|
No load input current |
|||||
|
No load output voltage |
235V±3% |
235V±3% |
235V±3% |
235V±3% |
235V±3% |
|
Short circuit voltage |
<6.9V |
<6.9V |
<6.9V |
<6.9V |
<7.5V |
|
General parameters |
|||||
|
Fuse wire |
80A |
63A |
50A |
35A |
25A |
|
Primary winding resistance |
<55mΩ |
<64mΩ |
<80mΩ |
<131 mΩ |
<245mΩ |
|
Secondary winding resistance |
<45mΩ |
<64mΩ |
<80mΩ |
<116 mΩ |
<228mΩ |
|
Iron loss |
<150W |
<105W |
<107W |
<77W |
<55W |
|
Copper loss |
<230W |
<200W |
<170W |
<125W |
<120W |
|
Efficiency |
>96% |
>96% |
>96% |
>96% |
>95% |
|
Maximum ambient temperature |
<40℃ |
<40℃ |
<40℃ |
<40℃ |
<40℃ |
|
No-load temperature rise |
<36℃ |
<33℃ |
<31℃ |
<26℃ |
<22℃ |
|
Full load temperature rise |
<65℃ |
<76℃ |
<67℃ |
<62℃ |
<55℃ |
|
Noise grade |
<40dB |
<40dB |
<40dB |
<40dB |
<40dB |
Isolated Power System Purpose
Medical isolated power panelsprovide safe, reliable and continuous power distribution for the important equipment at these locations:
Used for the transformation of TN system into IT system (ungrounded system)
Functions of real-time monitoring and fault alarming of the ground insulation resistance, transformer load current and transformer winding temperature of the monitored IT system
When the monitored IT system has an insulation fault, the medical isolation panel, hospital isolation panelcan initiate and generate the fault location signal to the system, and realize fault locating function with the help of the fault locator
Acrel isolation panel for operating roomscan remotely monitor the real-time operating conditions of up to 16 pieces of systems, and the main interface intuitively displays whether the access system communication is intact
Medica/Hospital Isolated Power Systems FAQs
What is hospital isolated power system?
It is an isolated power system operating roomthat conducts insulation monitoring and fault determination of the power distribution system, to further provide safe, reliable and continuous power distribution.
What are functions of medical isolated power panels?
What's the consists of medical isolated power supply system?
What are the applications of hospital isolated power systems?
If you want to know more kinds of split current transformer, please visit our website.
在线联系供应商
Other supplier products
| MICROGRID ENERGY MANAGEMENT SYSTEM | Under the threat of global energy shortage and the urgency to curb environmental pollution, the demand for energy conservation and emission reducti... | |
| ISOLATED POWER SYSTEMS | Acrel Isolated power systems are mainly used in the medical and industrial sectors. The medical IT system is mainly designed for the large threat t... | |
| ENERGY MANAGEMENT DEVICES | As a specialized energy meter supplier, Acrel provides various programmable energy metertypes for sale. Acrel Power Meters and Energy Management sy... | |
| ACREL ADW SERIES IOT POWER METER | Acrel ADW series energy meter using iotcompared with smart energy meter has the biggest advantage is that it can be controlled by wireless communic... | |
| AKH-0.66 / K Split Core Current Transformer | Acrel AKH-0.66/K series split core current transformerare mainly used in reconstructing projects of urban and rural power grids. The split core cur... |
Same products
| HM B200/B210 Series | 卖方: BEIJING HIGHMESH INFORMATION TECHNOLOGY CO.,LTD. | HM B200/B210 Series USRP B200/B210 Product Overview The B200 and B210 hardwarecovers RF frequen... | |
| HM-B210(Single-Board) | 卖方: BEIJING HIGHMESH INFORMATION TECHNOLOGY CO.,LTD. | HM-B210(Single-Board) Description TheB210is a highly integrated Universal Software Radio Periph... | |
| Busbar systems | 卖方: Jinan Dema Electric Co., Ltd. | Busbar systems Busbar Systems for Plant Reliable Industrial Power Distribution Solutions for Fa... | |
| Aluminium Cable Festoon System | 卖方: Jinan Dema Electric Co., Ltd. | Aluminium Cable Festoon System The company's newly developed DM-CL Aluminium Cable Festoon Syste... | |
| Open copperhead conductor Bar | 卖方: Jinan Dema Electric Co., Ltd. | Open copperhead conductor Bar DMGH Series Open Copperhead Conductor Bar Reliable Power Supply S... |