Ultraviolet accelerated aging test chamber
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
(1)IEC 61215 Terrestrial Photovoltaic (PV) Modules – Design Qualification and Type Approval
(2)IEC 61646 Thin-Film Terrestrial Photovoltaic (PV) Modules – Design Qualification and Type Approval
Parameters
| Item | Technical Specification |
|---|---|
| Structural material | High UV-blocking, high-sealing, high-thermal-insulation specialized material |
| Total UV irradiation area | ≥2000 mm × 2000 mm |
| Maximum test capacity | Simultaneous testing of 2 photovoltaic modules (2 m × 1 m each) |
| Light source type | Xenon lamp |
| Lamp lifetime (single set) | Approx. 1000 h |
| Spare lamps | 1 set (standard configuration) |
| System cumulative service life | ≥2000 h |
| UVA / UVB energy ratio | UVB accounts for approx. 3%–10% of total energy |
| Irradiance intensity | Approx. 100 W/m² in the 280–385 nm range |
| 280–320 nm proportion | ≤10% |
| Irradiance uniformity | Better than ±15% |
| UVA probe | 320–385 nm, fiber-optic type |
| UVB probe | 280–320 nm, fiber-optic type |
| Module surface temperature | 60 ℃ ±5 ℃ |
| Temperature resolution | 0.1 ℃ |
| Temperature measurement accuracy | ±0.1 ℃ |
| Ozone generation | Negligible |
| Irradiation method | Top irradiation, light source vertically directed at module surface |
| Control mode | Touchscreen control with automatic monitoring and light source cut-off |
Features
Large irradiation area suitable for full-size photovoltaic module testing
Independent UVA and UVB dual-band monitoring compliant with IEC standard requirements
High irradiance intensity effectively shortens test duration
Precise irradiance uniformity control ensures consistency and comparability of test results
Real-time module surface temperature monitoring prevents non-UV-related thermal degradation
Extremely low ozone generation avoids additional material aging effects
Touchscreen human–machine interface enables intuitive parameter setting and traceable operation status
Accessories
(1)Xenon lamp set (standard configuration)
(2)UVA fiber-optic probe (320–385 nm)
(3)UVB fiber-optic probe (280–320 nm)
(4)High-precision surface temperature sensor
(5)Control system (touchscreen and control module)
(6)Dedicated power supply and connection components
Test Procedures
Confirm normal status of the light source, temperature sensors, and irradiance probes before testing
Install photovoltaic modules according to standard requirements, ensuring the irradiated surface is perpendicular to the light source
Set irradiance intensity, test duration, and module surface temperature
Start the ultraviolet aging test and monitor system parameters in real time
Keep the chamber door closed throughout the test to prevent ultraviolet leakage
If irradiance or temperature exceeds preset limits, the system automatically cuts off the light source
After test completion, allow modules to cool to a safe temperature before removal
Maintenance Information
Regularly inspect xenon lamps and replace them according to specified service life
Periodically calibrate UVA and UVB probes to ensure measurement accuracy
Check temperature sensors and control systems for stable operation
Keep the irradiation chamber clean to maintain irradiance uniformity
Verify sealing components to prevent UV leakage and ensure operator safety
在线联系供应商
Other supplier products
| Flammability Limits Tester Manufacturer | LEL UEL Gas Explosion Safety Testing Equipment | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com E-mail:info@standard-groups。com Parameters UnitSpecification Env... | |
| Large Cone Calorimeter with Variable Atmosphere,Cone calorimeter for fire safety testing | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Large Cone Calorimeter with Variable Atmosphere is designed to measure the heat release... | |
| High precision manikin respiration simulation system | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Manikin Breathing System is a high-precision experimental device designed... | |
| China Photovoltaic Acid Mist Test Chamber Manufacturer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards IEC 60068-2-42. IEC 60068-2-43. IEC 60068-2-60 GB/T 10125 Artific... | |
| CE certified mask fit tester | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Qinsun PortaCount Respirator Fit Tester assesses the overall integrity of... |
Same products
| PP Pickling Tank | 卖方: Hebei Annuo Automaiton Technology Co., Ltd. | PP Pickling Tank PP Pickling Tank — Polypropylene Pickling Tanks for Hot Dip Galvanizing ... | |
| Galvanizing Tank/Galvanizing Kettle/Zinc Kettle | 卖方: Hebei Annuo Automaiton Technology Co., Ltd. | Galvanizing Tank/Galvanizing Kettle/Zinc Kettle Zinc Kettle / Galvanized Kettle for Hot-Dip Galv... | |
| Galvanizing Furnace | 卖方: Hebei Annuo Automaiton Technology Co., Ltd. | Galvanizing Furnace Hot Dip Galvanizing Furnace Manufacturer— Engineered for Your Producti... | |
| Flue Gas Heat Economizer waste heat recovery | 卖方: Hebei Annuo Automaiton Technology Co., Ltd. | Flue Gas Heat Economizer waste heat recovery Flue Gas Heat Economizer – Waste Heat Recover... | |
| Granite Tanks | 卖方: Hebei Annuo Automaiton Technology Co., Ltd. | Granite Tanks Granite Tank — Granite Pickling Tank for Hot Dip Galvanizing The granite ta... |