High-Precision Material Wear Tester with Probe
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
GB/T 3960-2016 Test Methods for Sliding Friction and Wear of Plastics
ISO 20502:2016 Test Method for Coating Scratch Resistance
ASTM D7027 Coating Scratch Hardness and Wear Scar Measurement
JB/T 8554-1997 Automatic Scratch Measurement Method for Coating Adhesion (can be used with wear scar analysis)
Parameters
| Parameter | Technical Specification |
|---|---|
| Wear Scar Scanning Mode | Automatic |
| Probe Vertical Resolution | 0.02 μm |
| Wear Volume Test Force Range | ≤40 g |
| Probe Vertical Measurement Range | ±0.5 mm (customizable) |
| X-Axis Movement Range | 0–50 mm |
| Y-Axis Movement Range | 0–25 mm |
| Z-Axis Movement Range | 0–50 mm |
| Sample Plate Size | Φ76 mm |
| Instrument Main Unit Dimensions | 230 × 300 × 260 mm |
| Instrument Main Unit Weight | 10 kg |
| Power Supply | 210–235 V, 50–60 Hz, 2500 W |
| Workbench Size | 1500 × 750 × 800 mm (user-provided) |
Features
High-precision measurement: vertical resolution of 0.02 μm ensures accurate detection of fine wear scars.
Automatic scanning and data processing: measures scar width, depth, and wear volume, generating two-dimensional profile maps.
Wide measurement range: Z-axis ±0.5 mm, X/Y axes movable to accommodate different sample sizes.
Multifunctional application: can function as a step gauge for coating thickness and surface profile measurements.
Data storage and display: real-time graphical display for research analysis and experimental data archiving.
User-friendly operation: easy sample installation, suitable for laboratory and industrial use.
Accessories
| Accessory | Specification/Description |
|---|---|
| Automatic Scanning Probe | High-precision measurement of wear scars and profiles |
| Sample Fixing Fixture | Fits Φ76 mm sample plate, ensures measurement stability |
| Power Cable | Input 210–235 V, 50–60 Hz |
| Data Acquisition and Analysis Software | Automatically generates 2D profile maps and calculates wear volume |
| Workbench Protective Cover | Ensures safe operation and prevents measurement interference |
Test Principle
The instrument uses a probe to scan along the wear scar either transversely or longitudinally, measuring scar width and depth while recording wear volume. Minute vertical displacements of the probe are captured with high precision to form a two-dimensional wear scar profile, reflecting material surface wear. The device allows quantitative evaluation of coating or material abrasion resistance and can perform step or profile measurements, providing reliable data for material performance research.
Operation Instructions
Ensure proper connection of power, and that the probe and sample fixture are securely installed.
Select appropriate scanning parameters and measurement ranges based on scar size and depth.
Do not manually touch moving probes or samples during testing to ensure safety.
After the test, save data and generate 2D profile maps and wear volume reports via software.
Regularly inspect the probe for wear and replace as needed to maintain measurement accuracy.
Test Procedure
Sample Installation: Fix the sample onto the sample plate and adjust the fixture to ensure a flat surface.
Parameter Setting: Choose scanning mode, measurement range, and test force.
Start Scanning: The probe automatically moves along the wear scar surface, collecting depth and width data.
Data Acquisition and Display: Generate a 2D wear scar profile in real-time and calculate wear volume.
Result Saving: Save data and generate analysis reports.
Data Analysis: Evaluate material surface abrasion resistance and coating quality based on wear morphology and wear volume.
在线联系供应商
Other supplier products
| High Precision Body Part Test Dummy for Ergonomic Studies | Whats/App: +86 156 1874 6768 Web: standard-groups.com 5th Percentile Female SAE Butt Form The 5th Percentile Female SAE Butt Form is meticulousl... | |
| Battery Adiabatic Calorimeter | Battery adiabatic calorimeters play a crucial role in battery research and development, providing essential thermodynamic data for safety assessmen... | |
| BS5852 Flammability Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ◆UK Regulation British furniture (fire and safety) regulations test ◆BS 5852 Me... | |
| Circuit Integrity Under Fire Conditions Apparatus BS 6387,cable fire performance tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard: BS 6387 Feature Control cabinet parameters: 1.With double flow meters and pre... | |
| Advanced suture needle cutting force equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications Evaluating Cutting Performance: By measuring the cutting force required for s... |
Same products
| Metal Angular Fixator Testing Machine | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Metal Angular Fixator Testing Machine... | |
| Bone Plate Four-Point Bending Fatigue Testing Machine | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Bone Plate Four-Point Bending Fatigue... | |
| Bone Plates Static Four - point Bending Tester | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Bone Plates Static Four-Point Bending... | |
| Spinal Internal Fixation System Assembly Testing Machine | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Spinal Internal Fixation System Assem... | |
| Spinal Internal Fixation System Component Testing Machine | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Spinal Internal Fixation System Compo... |