Parameters
| Item | Specification |
|---|---|
| Tube furnace heating length | 600 mm |
| Temperature range | 0–1100 °C |
| Temperature accuracy | ±0.1 °C |
| Quartz tube | Heat-resistant quartz, thickness ≥ 2 mm |
| SO₂ measurement range | 0–1000 ppm |
| SO₂ resolution | 1 ppm |
| SO₂ accuracy | ±30 ppm |
| NOₓ measurement range | 0–3000 ppm |
| NOₓ resolution | 1 ppm |
| NOₓ accuracy | ±30 ppm |
| Gas sampling bag | PTFE-coated, 80 L capacity, with imported manual sampler |
| CO measurement range | 0–10 % |
| CO resolution | 0.01 % |
| CO accuracy | Absolute ±0.1 % or relative ±3 % |
| CO₂ measurement range | 0–20 % |
| CO₂ resolution | 0.01 % |
| CO₂ accuracy | Absolute ±0.2 % or relative ±3 % |
| Spectral analysis device | Spectrometer at 500 nm wavelength, membrane thickness 10 mm |
| Absorption solution | 0.1 N NaOH |
| Titration reagents | Picric acid aqueous solution (3 g/dm³) and sodium carbonate aqueous solution (50 g/dm³) |
Features
Sealed chamber design with the smoke generation system installed internally, providing stable operation and effective protection of core components
Large observation window allowing clear visual monitoring of test status and specimen movement
Integrated industrial design with panel-mounted structure, offering intuitive operation and ergonomic user interface
Tube furnace protected by perforated outer housing to enhance heat dissipation and ensure uniform temperature rise to the target value
High-temperature K-type thermocouple for accurate measurement of specimen zone temperature
Ceramic combustion boat driven by strong magnetic coupling and transmission mechanism for stable and repeatable specimen delivery into the furnace core
Precision lead-screw transmission system enabling accurate positioning and reliable sealing of the quartz tube, reducing the risk of damage due to manual operation
Dual filtration columns filled with activated carbon and color-changing silica gel for effective carrier gas purification and drying
Branded computer with dedicated control software for automatic test process control
Software-based temperature regulation and stabilization with real-time curve recording, automatic data display, storage, and report generation
Test Procedures
1. Pre-Test Preparation
Ensure the tester is installed in a well-ventilated laboratory environment. Verify that the tube furnace, quartz tube, and filtration system are clean and properly assembled. Check that the absorption solution (0.1 N NaOH) and titration reagents are prepared according to specified concentrations. Confirm normal operation of the temperature control system, gas analysis instruments, and data acquisition software.
2. Specimen Loading
Place the prepared test specimen into the ceramic combustion boat. Using the magnetic drive and transmission mechanism, steadily convey the boat into the center of the tube furnace. The lead-screw positioning system ensures accurate alignment and reliable sealing at the quartz tube interface, minimizing gas leakage during the test.
3. Heating and Smoke Generation
Set the target temperature according to the selected standard. The furnace temperature is automatically controlled by the software, with uniform heating achieved through regulated power output. As the specimen undergoes thermal decomposition or combustion, the generated smoke is carried by a controlled gas flow toward the sampling and absorption system.
4. Gas Collection and Analysis
The combustion gases are either introduced into the NaOH absorption solution via bubbling or collected in PTFE-coated gas sampling bags for subsequent analysis. Online infrared gas analyzers continuously measure CO and CO₂ concentrations, while SO₂ and NOₓ are determined using dedicated measurement channels or chemical analysis methods. All test data are automatically recorded and stored by the control software.
在线联系供应商
Other supplier products
| ISO 4020 Fuel Water Separation Filter Test Rig | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Fuel Water Separator Test Bench primarily follows ISO/TS 16332-2006 "Dies... | |
| Filter Air Tightness Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Filter Air Tightness Tester is a dedicated device designed to evaluate the sealing perf... | |
| high low temperature gravel impact tester for coatings | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application This tester is widely applied in durability and damage resistance evaluation f... | |
| Cleveland Open Cup Auto Flash Point Analyser, ISO2592,auto ignition flash point equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com Products: The instrument complies with ISO2592 "petroleum products of flash and fire assay... | |
| Diaper impurity testing machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Diaper Impurity Testing Machine is designed for visual inspection of disposable diapers... |
Same products
| Laser Cutting Machines | 卖方: Chanxan (Changshu) Laser Technology Co., Ltd. | Laser Cutting Machines PrecisionLaser Cutting Systemsfor Film, Glass & Industrial Materials ... | |
| Precision Laser Micromachining Machine | 卖方: Chanxan (Changshu) Laser Technology Co., Ltd. | Precision Laser Micromachining Machine| Ultrafast Laser Micromachining Systems Chanxan provides ... | |
| Ultrafast Laser Systems | 卖方: Chanxan (Changshu) Laser Technology Co., Ltd. | Ultrafast Laser Systems Femtosecond Laser System Ultra-precision femtosecond laser systemfor hi... | |
| Industrial Laser Equipment Manufacturer | 卖方: Chanxan (Changshu) Laser Technology Co., Ltd. | Industrial Laser Equipment Manufacturer for Precision Cutting, Marking & Micromachining Chan... | |
| Common Rail Injector Valve F00R J02 012 | 卖方: 868983 | Common Rail Injector Valve F00R J02 012 David Chen |