High Precision Rotation Luminaire Goniospectroradiometer
LSG-1890BCCD/LSG-1800ACCD goniospectroradiometer is high precision automatic goniophotometric instrument for luminous intensity distribution measurement with facility for rotating the light source. The LSG-1890BCCD/LSG-1800ACCD can do the spatial CCT test and the intensity distribution test. It is for industrial laboratory photometric data measurement of luminaires such as LED luminaires, LED Plant Lighting, HID luminaires, fluorescent lamps and so on.
Standard:
Full meet theEN13032-1clause 6.1.1.3 goniophotometer type 1,COMMISSION DELEGATED REGULATION (EU) 2019/2015,CIE-70,CIE-121,IES LM-79,IES LM-75,IES LM-80and ect.
Measurement:
Luminous Intensity Data, Luminous Intensity Distribution, Zonal Luminous Flux, Luminaries Efficiency, Luminance Distribution(Option), Coefficient Of Utilization, Luminance Limitation Curves Glare, Maximum Ratio of Distance to Height, Equal Illuminance Diagrams, Curves of Luminaires VS Lighting Area, Isocandela Diagrams, Efficient Luminescence Angle, EEI, UGR, CCT, CRI, Color Temperature Distribution, Spectrum, (x, y)/(u, v), PAR, PPF, PPFD 3D distribution test and etc.
Specification:
• LSG-1890BCCD/LSG-1800ACCD Goniospectroradiometer=LSG-1890B/LSG-1800A Goniophotometer System+LPCE-2 Integrating sphere Spectroradiometer System
• The goniophotometer can do the B-β test with two pillars and the C-γ test with single pillar & multi-function jigs
• Test Max Luminaires size and weight: LSG-1890BCCD is 2000mm/60kg and LSG-1800ACCD is 1600mm/50kg
• The tested luminaries rotates around an angle of (γ)±180°(or 0-360°) and the tested luminaries rotates around itself with an angle of (C)±180°(or 0-360°)
• Luminosity Testing Range:Illuminance0.01lx~99,999lx;Light Intensity1.0cd~10^7cd (detector)
• Angle accuracy: LSG-1890BCCD is 0.1°, LSG-1800ACCD is 0.2°
• Work with high accuracy and quick CCD Spectroradiometer to measure spatial color parameters
• Accuracy of chromaticity coordinate: ±0.0015 or ±0.0005(under standard A lamp)
• Spectral Range Wavelength: 350nm ~800nm; Accuracy of wavelength: ±0.5nm
• The Hardware and Software can do the PAR, PPF and PPFD 3D distribution test, which can export IES/LDT files
•PhotometryAccuracy: CIE Class A (Class L is option)
• Testing Accuracy: 2%(Under Standard lamp); Stray Light: less than 0.1%
• English version software can run in Win7, Win8 or Win10, Win11
• LISUNgoniophotometersoftware can export CIE, IES, LDT and other format files. These kinds of format files can be transferred by other illumination and luminaire design software such as DiaLux
在线联系供应商
Other supplier products
High Precision Rotation Luminaire Goniophotometer | What is goniophotometer definition? LSG-1890B/LSG-1800Agoniophotometeris automatic goniophotometric instrument for luminous intensity distribution... | |
Automatic Electrical Safety Tester | Product Description The LS9955/LS9956Automatic Safety Test System(electrical safety analyzer) combined the following function in one set: With... | |
Electrodynamic Vibration Generator Test Table | Product Description LVD-100KG/LVD-100KG-6DElectrodynamic Vibration Generator Test Table is according toIEC 60068, IEC68-2-6,JJG189-97,GB/T1330... | |
Portable CCD LED Colorimeter Spectroradiometer | The Portable CCD LED Colorimeter Spectroradiometer has the following versions which can measure different parameters: LISUN Model Functio... | |
High Voltage Lightning Surge Generator | Basic Info. Model NO. SG-61000-5 After-sales Service Perfect Warranty 24months Trademark LISUN Origin China ... |