Oxidizing Solids Tester

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Conforms to UN Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria 34.4.1 Test O.1: Test for oxidizing solids, 34.4.3 Test O.3: Test for oxidizing solids by weight and GB/T 21617-2008 Test Methods for Oxidizing Solids of Dangerous Goods.

Applications

This testing instrument is used to determine the potential of a sample to increase the combustion rate or intensity of a certain combustible substance when mixed with it. The combustion rate can be determined by the combustion time test or gravimetric analysis, thereby confirming its packaging category. It is applicable to solid substances.

Standards

GB/T 21617-2023:Dangerous goods - Flammable solids - Test method for autoignition temperature

GB/T 21617-2008:Dangerous goods - Flammable solids - Test method for autoignition temperature

Features

1. It adopts an embedded processor and the Windows CE operating system.

2. With an 8 - inch liquid - crystal touch screen, it can display the test status in real - time.

3. It is compatible with both O.1 standard and O.3 standard.

4. It uses full - automatic photoelectric detection technology to reliably detect combustion phenomena.

5. It has a full - automatic weighing detection by balance, which automatically records the mass loss during the combustion process.

6. It is equipped with a mold made of standard metal wire.

7. It can automatically detect the breakage of metal wires and provide audible and visual alarms.

8. It has a high - precision temperature and humidity detection function to effectively collect environmental temperature and humidity data.

9. During the test process, it displays the running time and test animation images in real - time.

10. It supports remote - controlled ignition to ensure the safety of personnel during the experiment.

11. The test process is fully automated, requiring no manual intervention. It supports functions such as parameter setting, data storage, and report viewing.

12. It is equipped with a temperature and humidity module to monitor the temperature and humidity of the test environment in real - time.

Technical Parameters

Working Environment (5~40)℃, <85%RH
Heating Method Standard shape inert metal wire heating
Heating Wire Power 150W±2W
Detection Method Infrared sensor array detection method or balance weighing detection method
Balance Weighing Accuracy ±0.1g
Ambient Temperature Measurement Range -10℃~+60℃
Ambient Temperature Measurement Accuracy ±0.5℃
Ambient Humidity Measurement Range 5%RH~95%RH
Ambient Humidity Measurement Accuracy ±5.0%RH
Timing Display Resolution 1s
Timing Accuracy ±7ms/Min
Test Plate Made of corundum, 150mm long, 150mm wide, 6mm thick
Sample Making Molds Made of PTFE, inner diameter 70mm, angle 60° cone

Accessoriess

1. Core Operating Components

Test Carrier Components: Standard metal wire mold (compatible with UN O.1/O.3 and GB/T 21617 standards, used for loading oxidizing solid samples and reference materials);

Detection Components:

Fully automatic photoelectric detection device (reliably identifies combustion phenomena);

Metal wire breakage detection sensor (in conjunction with an audible and visual alarm device);

High-precision temperature and humidity detection sensor (collects ambient temperature and humidity data);

Weighing Unit: Integrated high-precision balance (automatically records mass loss during combustion);

Control and Display Components: 8-inch LCD touchscreen (equipped with Windows CE system, displays test status in real time);

Safety Components: Audible and visual alarm device (triggers warnings in case of metal wire breakage or abnormalities).

2. Auxiliary Components

Sample Processing Components: Sample spoon (for quantitatively taking oxidizing solid samples), standard reference materials (e.g., dried cellulose, matching the O.1/O.3 standard control sample);

Calibration and Maintenance Components: Balance calibration weights (calibrates weighing accuracy), metal wire calibration components (verifies breakage detection function);

Data Management Components: USB flash drive (exports test data and reports).

Test Procedures

1. Sample and Control Preparation

Grind and sieve the oxidizing solid sample to be tested (ensuring the particle size meets the standard requirements);

Prepare the control (e.g., dried cellulose) according to the test standard (O.1 or O.3), and weigh the sample and control in the specified proportions (O.1 is a sample-cellulose mixture, O.3 is a sample prepared by weight ratio).

2. Instrument Preparation

Connect the power supply and start the instrument; the embedded processor will automatically complete a fault self-check;

Calibrate the integrated balance with standard weights to confirm that the temperature and humidity sensor and photoelectric detection device are functioning properly;

Select the test mode (“O.1 Standard Mode” or “O.3 Standard Mode”) via the 8-inch touchscreen.

3. Sample Loading and Adjustment

Load the prepared sample and control into the standard metal wire mold and fix the mold to the designated test position on the instrument;

Confirm that the metal wire is within the monitoring range of the fracture detection sensor, and close the test chamber.

4. Start Automatic Testing

Set the test parameters (e.g., combustion monitoring duration) on the touchscreen and click "Start"; The instrument automatically records the initial mass of the sample, then starts the test process, monitoring combustion phenomena, mass loss, and the state of the metal wire in real time.

5. Test Process and Judgment

During the test, if the metal wire breaks, the instrument triggers an audible and visual alarm and records the corresponding status; After the test, the instrument automatically calculates the combustion duration and mass loss data, and determines the oxidizing power level of the sample according to UN/GB standards.

6. Completion and Data Export

After the test chamber cools, remove the mold and clean the sample residue; Export the test data, ambient temperature and humidity records, and oxidizing power determination report to a USB flash drive via the touchscreen; Turn off the instrument power to complete the operation.


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

High-speed Impact Testing Apparatus
High-speed Impact Testing Apparatus Whats/App: +86 156 1874 6768 Web: standard-groups.com Main Purpose To test the impact - resistance performance of glasses, goggles and face shie...
Vacuum Friction and Wear Tester
Vacuum Friction and Wear Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Vacuum Friction and Wear Tester integrates a proven friction and wear testing mechanism...
Catheter Friction Tester
Catheter Friction Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Catheter Friction Tester is a precision instrument used to measure the surface friction...
Textile Far Infrared Emissivity Tester Manufacturer Direct Sale Factory Price
Textile Far Infrared Emissivity Tester Manufacturer Direct Sale Factory Price Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Quality control and performance evaluation of natural fibers suc...
Water Leakage Sensor
Water Leakage Sensor Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Description GB/T 21619-2008 Test Methods for Dangerous Goods Flammable Solids Emit...
供应产品

Same products

PH-2398  pH/EC/ORP/温度 多参数测试仪
PH-2398 pH/EC/ORP/温度 多参数测试仪 卖方: 科立龙电子股份有限公司 大显示,可同时进行水质PH、氧化还原电位、电导率、温度的测量; 带背光LCD显示清晰,功能齐全,精度高; 仪表采用微电脑测量与控制技术,具有自动校准功能,方便实用。 内置蓝牙,可以与电脑或...
ORP-2026S  防水型 土壤ORP&温度 二合一测试仪
ORP-2026S 防水型 土壤ORP&温度 二合一测试仪 卖方: 科立龙电子股份有限公司 范围 ORP -999~999mV Temp 0~50℃ 分辨率 ORP 1mV...
防水型 EC /TDS/温度三合一参数测试仪
防水型 EC /TDS/温度三合一参数测试仪 卖方: 科立龙电子股份有限公司 范围 EC ~19.9mS TDS Temp 0~50℃ 分辨率 ...
EC-98330 电导率营养测试棒
EC-98330 电导率营养测试棒 卖方: 科立龙电子股份有限公司 本营养测试仪适用于水培和农业应用,能够准确测量水中的营养液电导率(EC)单位为mS/cm,以及总溶解固体(TDS)转换为百万分之一(ppm)。 适合需要较高EC值的植物,该测试仪具有高达6.0...
PH-808防水笔式土壤PH计
PH-808防水笔式土壤PH计 卖方: 科立龙电子股份有限公司 两点按键校准,自动识别pH标准缓冲液 自动温度补偿确保全量程范围的精确测量。 自动电源关闭,有效节省电池能耗。 可替换型电极降低仪表维护与使用成 全防水性能保障在恶劣环境中对仪表的完整保...