Nasal Oxygen Cannula Tester

Whats/App: +86 156 1874 6768

Web: standard-groups.com

This device is used to test various performance characteristics of nasal cannulas, such as airflow resistance and resistance to kinking. It is also known as a nasal cannula airflow resistance and kinking resistance tester or nasal cannula testing equipment. This device is suitable for all nasal cannulas, whether they are standalone or used in conjunction with humidifiers or masks. It can test parameters such as airflow resistance, resistance to kinking, leakage, pressure resistance, and bending resistance (patency).

Applications

The nasal oxygen tube tester is mainly used to test key parameters of nasal oxygen tubes, such as oxygen content, flow rate, and pressure, as well as to evaluate their physical properties like pressure - resistance, crush - resistance, and air permeability. This tester plays an important role in various aspects, including the production, quality control, and pre - use inspection of nasal oxygen tubes, ensuring that nasal oxygen tubes can provide patients with a stable and effective oxygen supply. Specifically, its main uses are as follows:

- Production Quality Control: During the production of nasal oxygen tubes, manufacturers can use this tester to conduct sampling inspections on finished products, ensuring that each batch of nasal oxygen tubes complies with relevant standards and regulations. - Pre - use Inspection: Before using nasal oxygen tubes, medical institutions can use this tester to conduct quality inspections to ensure that the performance of nasal oxygen tubes is stable, safe, and reliable. - Research, Development and Improvement: The nasal oxygen tube tester can also be used in the research, development and improvement process of nasal oxygen tubes, helping R & D personnel understand the performance characteristics and deficiencies of nasal oxygen tubes, so as to make targeted improvements and optimizations.

Standards

YY/T 1543 - 2017:Nasal Oxygen Cannula

GB/T 1962.1-2005:Conical fittings with a 6%(Luer) taper for syringes, needles and certain other medical equipment -- Part 1: General requirement

Features

• Supports five major testing functions: airflow resistance, crush resistance, leak testing, and pressure resistance.

• 7-inch color touchscreen control with an intuitive interface for a significantly improved user experience.

• The device is equipped with a miniature printer for real-time printing of test results.

• Equipped with a high-precision pressure sensor, allowing for stable control of output pressure via a rotating pressure knob.

• Equipped with a high-precision flow meter for accurate measurement of flow through the tubing.

• Unique rotary design accommodates crush resistance testing of nasal oxygen tubes with various inner diameters.

• Features a universally designed conical test interface, meeting the testing requirements of most products on the market.

• The device has powerful built-in data storage capabilities.

• Miniature printer and universal USB data interface for convenient data output and transfer.

• Intelligent features such as overload protection and power-off memory effectively ensure user safety.

• Free software upgrade service is provided.

Technical Parameters

ParameterSpecification
Test Pressure Range Usually covers 0 - 600kPa or higher, to meet testing requirements of nasal oxygen tubes with different specifications and materials
Test Flow Range Simulates normal oxygen supply flow (e.g., not less than 15L/min), ensuring air permeability of nasal oxygen tubes in actual use
Accuracy High level of accuracy; e.g., pressure accuracy reaches ±1% or higher, ensuring test result accuracy and reliability
Data Collection and Display Equipped with data collection card and display screen, which collect and display real-time test data (pressure, flow rate, time, etc.) for easy observation and recording
Clamping Device Fixes nasal oxygen tubes to keep their position stable (no displacement/shaking during testing); adaptable to nasal oxygen tubes of different specifications and shapes
Operation Interface Simple and clear for easy operation; some testers are equipped with touch-screen control and menu-style functions to improve operational convenience

Accessoriess

1 main unit;

1 set of testing software;

1 instruction manual;

1 certificate of conformity;

1 warranty card;

1 delivery receipt;

1 nameplate;

1 power cord;

1 set of wrenches;

Several brochures;

1 x 1/8 barbed connector;

1 x 1/8 Luer connector (male).

Test Procedures

1. Pre-test Preparation

Check that the instrument's power supply and pipeline connections are normal, start the instrument and preheat it to a stable working state.

Calibrate the pressure and flow sensors (refer to the instrument calibration guide) to ensure that the detection accuracy meets the requirements.

Clean the fixtures and pipelines of impurities, prepare the nasal oxygen cannula samples to be tested, and record the basic information of the samples, such as specifications and materials.

2. Sample Installation

Connect the nasal oxygen cannula sample to the corresponding pipeline interface of the instrument, and use a special fixture to fix both ends of the sample, ensuring that the connection is sealed without leakage and that the sample is not bent or displaced.

3. Test Parameter Setting

Through the instrument's touch screen interface, select the target test item (such as airflow resistance, pressure resistance, leakage test, anti-flattening performance, etc.).

According to the corresponding standard for nasal oxygen cannulas (such as YY/T 1543-2017), set the corresponding test parameters (such as test flow rate, pressure threshold, pressure holding time, etc.), and save the parameters after confirmation.

4. Starting the Test

After confirming that the sample connection and parameter settings are correct, click the "Start" button. The instrument will automatically execute the test process, collecting and displaying data such as flow rate, pressure, and time in real time.

5. Test Process Monitoring

During the test, observe the data changes on the instrument interface and the sample status. If any abnormalities such as loose pipelines or leaks occur, pause the test immediately and handle the issue.

6. Test Completion and Cleanup

After the test is completed, the instrument will automatically shut down. Record or export the test data (such as airflow resistance value, pressure resistance results, etc.).

Loosen the fixture and remove the sample, clean the fixture and pipelines, turn off the instrument power, and prepare the test report.


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Shoe heat insulation meter -- sand bath
Shoe heat insulation meter -- sand bath Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard GB/T 20991: Determines thermal insulation of safety shoe soles under simulated ho...
Wet leakage current test system of photovoltaic module
Wet leakage current test system of photovoltaic module Whats/App: +86 156 1874 6768 Web: standard-groups.com The Wet Leakage Current Test System is designed for high-precision evaluation of insulation...
High-Precision Vertical Fire Test Furnace for Construction Elements
High-Precision Vertical Fire Test Furnace for Construction Elements Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The equipment complies with the following standards: (1) GB/T 9978.4-2008 &ndas...
Rubber Sleeve Rebound Tester
Rubber Sleeve Rebound Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The tester is applicable to a variety of medical and laboratory products, incl...
Ergonomics and Human Thermal Response Measurement
Ergonomics and Human Thermal Response Measurement Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The Porton-man Thermal Manikin is suitable for a wide range of applications, i...
供应产品

Same products

PH-2398  pH/EC/ORP/温度 多参数测试仪
PH-2398 pH/EC/ORP/温度 多参数测试仪 卖方: 科立龙电子股份有限公司 大显示,可同时进行水质PH、氧化还原电位、电导率、温度的测量; 带背光LCD显示清晰,功能齐全,精度高; 仪表采用微电脑测量与控制技术,具有自动校准功能,方便实用。 内置蓝牙,可以与电脑或...
ORP-2026S  防水型 土壤ORP&温度 二合一测试仪
ORP-2026S 防水型 土壤ORP&温度 二合一测试仪 卖方: 科立龙电子股份有限公司 范围 ORP -999~999mV Temp 0~50℃ 分辨率 ORP 1mV...
防水型 EC /TDS/温度三合一参数测试仪
防水型 EC /TDS/温度三合一参数测试仪 卖方: 科立龙电子股份有限公司 范围 EC ~19.9mS TDS Temp 0~50℃ 分辨率 ...
EC-98330 电导率营养测试棒
EC-98330 电导率营养测试棒 卖方: 科立龙电子股份有限公司 本营养测试仪适用于水培和农业应用,能够准确测量水中的营养液电导率(EC)单位为mS/cm,以及总溶解固体(TDS)转换为百万分之一(ppm)。 适合需要较高EC值的植物,该测试仪具有高达6.0...
PH-808防水笔式土壤PH计
PH-808防水笔式土壤PH计 卖方: 科立龙电子股份有限公司 两点按键校准,自动识别pH标准缓冲液 自动温度补偿确保全量程范围的精确测量。 自动电源关闭,有效节省电池能耗。 可替换型电极降低仪表维护与使用成 全防水性能保障在恶劣环境中对仪表的完整保...