Infusion Pump Flow Rate Tester
The Infusion Pump Flow Rate Tester is specifically designed for testing the flow rate of sodium chloride solution in infusion sets. It is suitable for use by medical device manufacturers, third-party testing institutions, medical device inspection and research institutes, and research and educational institutions.
Applications
The infusion set flow rate tester is an instrument specifically designed for measuring and monitoring the flow rate of infusion sets. Its main uses include: - Flow Rate Measurement: Precisely measure the flow rate of the infusion set, usually displayed in milliliters per hour (ml/h) or drops per minute (gtt/min). - Flow Monitoring: Monitor the changes in flow rate in real - time during the infusion process, and promptly detect abnormal flow rate situations, such as the flow rate being too fast, too slow, or blocked, to ensure the safety and effectiveness of the infusion process. - Alarm Function: When the infusion flow rate exceeds the set normal range, the tester can emit audible and visual alarm signals to remind medical staff to take timely action and avoid potential medical risks. - Data Recording and Storage: Record and store information such as flow rate data and time during the infusion process, which is convenient for medical staff to view and analyze, and helps track and evaluate the patient's infusion treatment.
Standards
GB 8368 - 2018 : Disposable Infusion Sets
Features
1. Industrial-grade capacitive touch screen with a menu-driven interface for easy and convenient operation;
2. The fixture is suitable for various types of catheters;
3. Equipped with a miniature printer for easy printing of test results.
Technical Parameters
| Parameter | Specification |
|---|---|
| Testing Range | Wide flow rate testing range (e.g., 1 - 2000ml for some testers), adapting to different types/specifications of infusion sets (varies by catheter specifications) |
| Testing Accuracy | High accuracy for flow rate measurement (e.g., ±5ml/h or ±0.1gtt/min, etc.) to ensure test result accuracy |
| Static Water Level Height | Adjustable range: usually 0 - 1000mm, meeting requirements under different testing conditions |
| Temperature Control Range | Usually 5 - 90℃, simulating infusion situations under different temperature conditions |
| Fixtures | Adapt to various catheters; customized according to catheter specifications to firmly hold infusion sets of different specifications during testing |
| Flow Calculation | Adopts methods like the weighing method for flow calculation, ensuring accurate flow measurement |
| Alarm Threshold | Allows users to set the alarm threshold; triggers an alarm when the flow rate exceeds the set range |
Accessoriess
Standard accessories include: specialized fixing clamps for different types of infusion sets (for securely holding various infusion sets), connecting tubing (for setting up the flow path to complete flow rate testing), equipment calibration auxiliary parts (to ensure testing accuracy), operating manual, and calibration certificate.
Test Procedures
Infusion Flow Rate
For infusion sets with a drip rate of 20 drops/mL, the infusion set should deliver at least 1000 mL of sodium chloride solution [mass concentration p(NaCl) = 9 g/L] within 10 minutes under a static pressure head of 1 meter.
For infusion sets with a drip rate of 60 drops/mL, the infusion set should deliver at least 1000 mL of sodium chloride solution [mass concentration p(NaCl) = 9 g/L] within 40 minutes under a static pressure head of 1 meter.
在线联系供应商
Other supplier products
| Dental Implants Fatigue Tester | Standard ISO 14801:2016: Specifies dynamic fatigue testing protocols for endosseous dental implants, including load ratios, frequency, and failure... | |
| Heel impact fatigue testing machine丨Heel impact fatigue tester | Main uses: Heel impact fatigue testing machine is mainly used to test the continuous impact tolerance index of women's high heels, mid-heels and o... | |
| Controll Four-ball Friction Tester | Applicable Standards: When conducting tests, the microcomputer - controlled four - ball friction testing machine needs to comply with the followin... | |
| Sliding Bearing Fatigue Testing chamber | Standards ASTM D2596 / ASTM D2783 / IP 239 EP, DIN 51350 / 1-5. ISO / CD 11008. VW 1444. VW 1454 Material Research and Development During the re... | |
| In-situ expansion analysis system | The In-situ Expansion Analysis System is designed for high-precision measurement of dimensional and volumetric changes in electrochemical cells and... |
Same products
| Full-Body Garment Stress Tester,FZ/T 70015 | 卖方: Standard International Group (HK) Limited | TheFull-Body Garment Stress Testeris a precision instrument designed to quantify the pressure exe... | |
| Industrial Dry Cleaning Machine,industrial garment steamer | 卖方: Standard International Group (HK) Limited | TheIndustrial Dry Cleaning and Energy-Saving Dryer Machinesare designed for high-performance, rel... | |
| MST Medical Compression Stocking Tester,compression sock tester | 卖方: Standard International Group (HK) Limited | The MST Medical Compression Stocking Tester is a professional instrument designed to evaluate the... | |
| MMT Liquid Water Separation Tester,MMT Liquid Water Analysis Tool | 卖方: Standard International Group (HK) Limited | Applications Textile Manufacturing:Testing waterproof fabrics for outdoor gear R&D:Developi... | |
| Dental material color stability tester | 卖方: Standard International Group (HK) Limited | Application The applications of color stability testers for dental materials are mainly reflecte... |
















