Bacterial Filtration Efficiency Tester
Standards:
YY 0469-2004 Technical requirements for surgical mask BS EN 14683-2014 Medical face masks - Requirements and test methods ASTM F2100-2019 Standard Specification for Performance of Materials Used in Medical Face Masks ASTM F2101 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
Applications:
Mask Bacterial Filtration Efficiency Tester is applied to test the filtration efficiency of mask, can be used in metering calibration department, scientific research institution, mask manufacturer and other related departments.
Product Information:
Mask bacterial filtration efficiency (BFE) tester is mainly applicable to the measurement and verification departments, scientific research institutes, medical mask manufacturers and relevant testing departments for the performance of mask bacterial filtration efficiency.It is composed of aerosol generation system, aerosol chamber and aerosol transmission device, negative pressure cabinet and 28.3l /min sampler.The whole tester is controlled by the console.The control console adopts microcomputer control system to coordinate and control the work of aerosol generation system, transmission system, negative pressure cabinet and sampling system, and display the working status in real time. The whole measurement work is completed automatically.
| Main Parameter | Parameter range |
| vacuum pump | Oil-free rotary vane vacuum pump, maximum flow up to 57LPM, equipped with muffler, stable airflow |
| Aerosol generator flow control | The flow control of aerosol generator adopts mass flow control and the control is accurate |
| Channel I sampling flow | Anderson sampling flow of 28.3 L/min, resolution: 0.1 L/min |
| sampling flow | Anderson sampling flow of 28.3 L/min, resolution: 0.1 L/min |
| sampling | Equipped with mass flow controller |
| Pneumatic pressure | 0-5KPa |
| Spray flowmeter pressure | 0-300KPa |
| aerosol chamber negative pressure | 0-120KPa |
| Working Temperature | 0-50℃ |
| Cabinet negative pressure | 0-120Pa |
| Cabinet negative pressure | Data can be downloaded via usb flash drive |
| Data storage capacity | Data can be downloaded via usb flash drive |
| mass flow controller | 0-20L/min,resolution:0.1 L/min |
| Peristaltic pump flow | (0.01~3.0)mL/min |
| High efficiency air filter characteristics | The filtration efficiency of particles above 0.3um is≥99.99% |
| Median mass diameter of aerosol generator | The average particle diameter was(3.0±0.3)μm,Geometric standard deviation≤1.5 |
| Two-channel 6-stage Anderson sampler captures particle size | Ⅰclass>7μm,class(4.7~7)μm,Ⅲ class(3.3~4.7)μm,class(2.1~3.3)μm,class(1.1~2.1)μm,class(0.6~1.1)μm |
| Fog chamber specifications | (600×80×5)mm |
| Total number of positive quality control sampler particles | (2200±500)cfu |
| Ventilation flow of negative pressure system | 800m³/min |
| Operation room dimension | (740×940×440)mm |
| dimension | About 1200×650×2100mm |
| Instrument noise | <65Db(A) |
| weight | 120kg |
| power consumption | <1500W |
在线联系供应商
Other supplier products
| Oil Filter Hydrostatic Breakage Tester | Meet the standard: ISO 2941 SAE J1858 Product Introduction: The Oil Filter Hydrostatic Breakage Tester operates by using a liquid medium (typ... | |
| Filter Flow Resistance Tester | Meet the standard: ISO 3968 SAE J806 Product Introduction: The Filter Flow Resistance Tester operates by circulating a test fluid (such as hy... | |
| Wool Shrinkage Tester | Product Introduction: 1, YM116 type wool fold shrinkage tester is a precision instrument to determine the length of the towel under the standard t... | |
| Minimum Ignition Energy (MIE) – Dust Cloud Tester/ IEC 61241-2-3 Dust Ignition Tester/ MITC Dust Explosion Tester | Description: Dust Cloud Minimum Ignition Energy Tester disperses a certain mass of dust specimen to be tested into the Hartmann tube device thro... | |
| Anti Synthetic blood penetration tester | Standards: GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood. YY/T 0691-2008 Clothin... |
Same products
| Battery Diaphragm Gas Permeability Tester ISO | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Battery Diaphragm Gas Permea... | |
| Torsion Testing Tester ASTM, JIS, and DIN | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Universal Torsion Testing Ma... | |
| Wire Torsion and Winding Tester ASTM and JIS | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Wire Torsion and Winding Tes... | |
| Metal Wire Repeated Bending Tester ISO/DIS 7801 | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Metal Wire Repeated Bending ... | |
| Metal Wire Wrap Tester ISO 4101 | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Metal Wire Wrap Tester is de... |