Multifunctional radio-measuring instrument RST-430
A universal radio test set for complete maintenance and repair of FM and AM radio stations of the VHF band
RST-430 is a multipurpose radio measuring device for maintenance of communication equipment in indoor and field conditions.
Functions
- HF generator
- LF generator
- Frequency meter
- Power meter
- Transmitter test load meter
- Modulation meter
- Distortion factor meter
- AC voltmeter
Measured parameters
- Carrier-to-noise ratio
- Transmitter frequency
- Ringing frequency
- Frequency deviation
- Sensitivity of transmitter modulation input
- Distortion factor of transmitter modulation characteristics
- Receiver sensitivity, including the signal-to-noise and distortion method (SINAD)
- Receiver output voltage
- Selectivity
- Distortion factor
- Noise suppressor threshold
Standard scope of supply
- Radio test set RST-430
- AC220/12 V power adapter
- Instrument certificate, testing guidance, operation manual
- Set of cables
- Carrying case
Technical characteristics
1. HF generator
Baseband | 90... 210 MHz |
Allowed additional bands - by agreement with the Customer (frequency step 1 kHz) | 100kHz...470(527) MHz |
Carrier setting error | < 10-6* |
Deviation in FM mode | 0.1 ... 20 kHz(step 10 Hz) |
Deviation setting error | <10% |
Output signal level (step 1 dBm) | -60 (-80)... -120 dBm |
Level setting intrinsic error | <3 dB |
Auxiliary error | < 0,1 µV |
Amplitude modulation(option) | 0…100% |
QAM setting error | < 10% |
Modulation signal frequency (step 1 Hz) | 20 Hz ... 20 kHz |
2. LF generator
Frequency band (step1 Hz) | 20 Hz ... 20 kHz |
Frequency instability | < 0.2 Hz |
Output signal level (step 10 mV) | 0 ... 2 V |
Setting error | < (0.01+0.03*U)V |
Distortion factor | < 0.3 % |
3. Modulation meter
FM deviation | 0.1 ... 20 kHz |
Error | < (0.1+0.1*D) kHz |
Modulation rate | 0...100% |
Error | < 10% |
4. Frequency meter
LF | 10 Hz ...100 kHz |
HF | 40 MHz... 240 MHz (...1000 MHz) |
Reference generator instability | < 10-6* |
5. Distortion meter
Frequency band | 20 Hz ... 10 kHz |
Limits | 1... 50 % |
Error | < (5+5*Kmst)% |
6. DC voltmeter
Range | 0 ... 30 V |
Error | < (2+2*(Uk/Umst))% |
7. AC voltmeter
Range | 0.02 ...20 V |
Frequency band | 20 Hz ... 20 kHz |
Error | < (0.02+0.03*U) V |
8. Power meter
Level | 0,5 ... 20 W (up to 50 W - w/ external attenuator FAD-2, up to 100 W - w/ FAD-3, by agreement with the Customer) Note that when using the FAD type fix attenuators or analogs for measuring parameters of radio transmitter with the output power above 20 W, the power value measured by the device must be multiplied by the attenuation (fading) ratio of the attenuator by the formula Ratt=10 to the power of Ratt, dB/10) |
Error | < [10+(Pk/Pmst)] % |
* If necessary, a reference generator with stability of up to 5*10-7 can be added.
NVRAM can be used for storage of the test results.
The device can be supplied with a power source which provides not less than 4 hours offline
Also a solar cell battery can be used as a power supply.
Carrying case
Dimensions: 420 × 380 × 205 mm
在线联系供应商
Other supplier products
Multifunctional radio-measuring instrument RST-430 | A universal radio test set for complete maintenance and repair of FM and AM radio stations of the VHF band RST-430 is a multipurpose radio measuri... | |
Digital combo set SVG-5 | A digital combo set.Available in four modifications. A digital combo set for maintenance and repair of carrier channels and data transmission equi... |
Same products
manual pressure testing pump | 卖方: Hangzhou Evershining Machinery Co.,Ltd | Item:RP-50 Pressure: 0--50bar Flow volume:45ml/min Water tank volume:12L Meas: 51*18*29cm | |
Digital Rail Head Wear Gauge | 卖方: 武汉联科瑞际轨道设备有限公司 | Rail wear gauge(rail head loss gauge also call it rail profile gauge)is mainly used to measure th... | |
Indicators | 卖方: ZHEJIANG HAOYU INDUSTRY & TRADE CO.,LTD | Electronic Scales cannot be properly operated without suitable weighing indicator. Their design a... | |
HY001 Weighing Indicator | 卖方: ZHEJIANG HAOYU INDUSTRY & TRADE CO.,LTD | Application of HY 001multiple Funcation Weighing Indicator with Printer HY001 must suitable for ... | |
Diaphragm Pressure Gauge | 卖方: Lanso Instruments LLC. | Diaphragm pressure gauges are mainly used to measure the pressure of special media such as high-v... |