ISO 11155 Filter Test System

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

The system complies with:

(1) ISO 11155-1 – Test method for particulate filtration performance of fuel cell air filters.

(2) ISO 11155-2 – Test method for adsorption performance of fuel cell air filters.

(3) DIN 71460-1 – German standard for particulate filtration of fuel cell air filters.

(4) DIN 71460-2 – German standard for adsorption performance testing of fuel cell air filters.

Parameters

ParameterUnitValue
Test Flow Rate m³/h 70 … 1400
Differential Pressure Measuring Range Pa ≤1000
Test Specimen Dimensions mm ≤500 × 700 (bypass option: 100 cm²)
Operating Medium (Aerosol) – ISO 12103-1 A2 fine test dust, DEHS, NaCl, KCl
Operating Medium (Gas) – Nitrogen dioxide, Hydrogen (others on request)
Operating Medium (Standard/ Pollutant Gas) – n-Butane (80 ppm), Toluene (80 ppm), Sulphur dioxide (30 ppm)
Operating Medium (Other Gas) – Nitrogen dioxide (30 ppm), Ammonia (30 ppm), Hydrogen sulphide (0.4 ppm), Acetaldehyde (30 ppm)
Power Supply – 3 × 400 VAC; 32 A; 50 Hz
Gas Sensors – FID (organic substances), UV-fluorescence (SO₂), CLD (NOx)
Other Sensors – Temperature, Relative Humidity, Barometric Pressure
System Dimensions mm 5000 × 2750 × 3400

Features

Gas-tight Test Channels: Ensure no leakage during aerosol or gas testing.

Modular Structure: Adaptable to different filter sizes and test requirements.

High Degree of Automation: Software-based test sequences minimize manual intervention.

Standardized Reporting: Automatic generation of test reports for efficiency and adsorption results.

Versatile Testing Mediums: Support for aerosols (salt, oil, dust) and gases (NO₂, H₂, n-butane, toluene, SO₂, NH₃).

Accessories

(1) Filter adapters – enable secure mounting of different filter types.

(2) Aerosol generators – for fine dust, DEHS, NaCl, KCl.

(3) Gas dosing units – for standard and pollutant gases including NO₂, H₂, n-butane, SO₂.

(4) Gas sensors – FID, UV-fluorescence, CLD for real-time detection.

Test Procedures

Preparation:

Select the filter type and confirm its dimensions fit the test duct.

Set up the appropriate aerosol or gas medium according to the test standard.

Testing:

Insert the test filter into the filter adapter and mount it in the test duct.

Define test flow rate and start the automated sequence.

The system measures differential pressure, dust loading, fractional efficiency, and adsorption/desorption curves.

Data Analysis:

Collect data via software and analyze filter performance.

Generate standardized reports for quality assurance or R&D documentation.

Maintenance Information

Ensure all gas and aerosol lines are properly sealed to prevent leaks.

Periodically calibrate sensors and aerosol generators to maintain measurement accuracy.

Keep the system clean and check mechanical components for wear.

Ensure ventilation and follow safety guidelines when working with gases.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

Textile far-infrared radiation temperature rise test instrument
Textile far-infrared radiation temperature rise test instrument Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications Textile Far Infrared Radiation Temperature Rise Tester is used for various te...
Metallic Cables and Wire Ropes Mechanical Property Testing Machine
Metallic Cables and Wire Ropes Mechanical Property Testing Machine Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0812-2010 (China): Specifies mechanical testing protocols for metal cables/c...
Safety Helmet High Temperature Pre-Treatment Chamber
Safety Helmet High Temperature Pre-Treatment Chamber Whats/App: +86 156 1874 6768 Web: standard-groups.com The safety helmet high-temperature pre-treatment chamber is manufactured according to the G...
Walk-in High-Temperature Aging Chamber,walk in stability chamber
Walk-in High-Temperature Aging Chamber,walk in stability chamber Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications Walk-in High-Temperature Aging Chambers are typically assembled on-site using...
ISO certified Photovoltaic module wire tube bending testing machine
ISO certified Photovoltaic module wire tube bending testing machine Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application (1) Bending fatigue testing of photovoltaic module output cables...
All supplier products

Same products

Filter Efficiency Testing Equipment
Filter Efficiency Testing Equipment Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com TheFilter Efficiency Testing Equipmentis ...
G506 Filter Performance Tester
G506 Filter Performance Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM F2100:Standard Specificat...
Automated Filter Integrity Tester
Automated Filter Integrity Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Testing Standards ASTM F838 – Memb...
High Efficiency Oil Water Separation Tester
High Efficiency Oil Water Separation Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO/TS 16332-2006 – Meth...
DIN 51524 hydraulic oil analyzer
DIN 51524 hydraulic oil analyzer Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards You can operate this unit in c...