Problems That Should Be Paid Attention to when Calibrating the Touch Trigger Probe

  1. touch triggerprobe is smaller than the static diameter

After learning about the principle of probe calibration, we can easily explain why the dynamic diameter of the probe stylus is smaller than the static diameter.

(1) The touch trigger probesis equivalent to a lever structure in principle. During the touch measuring, the sensor must be able to trigger (equivalent to the switch off) in order to send out a signal. Since the probe stylus (force arm) has a certain length, after the gem ball of the probe stylus touches the standard ball, it has to travel a certain distance to trigger the sensor. The longer the probe stylus, the greater the distance. Therefore, the trigger signal is delayed, so that the diameter of the fitting ball is smaller than the sum of the diameter of the gem ball and the standard ball. When the software subtracts the diameter of the standard ball from the diameter of the fitting ball, the "dynamic diameter" of the calibrated probe stylus we get is smaller than its static diameter.

(2) The probe stylus will be slightly deformed during the touch measuring, which increases the signal delay, which is also one of the reasons for this phenomenon.

(3) The time required for the trigger signal of the sensor (probe) to reach the counter is fixed. However, the rate of change of the grating reading during this time is related to the touch speed of the measuring machine. When the touch speed is fast, the "dynamic diameter" of the probe stylus is small.

  1. measurement touch probe

The calibrated dynamic diameter is smaller than the static diameter, which will not affect the measurement accuracy of the measuring machine. On the contrary, it also compensates for the delay of the contact measurement and the deformation of the probe stylus. Because the ball radius correction of the stylus measuring software used by the measuring machine (converting the coordinates of the ball center point measured by the needle to the coordinates of the contact point) is a dynamic diameter, but not a static diameter. When calibrating the probe, please pay attention to the following issues:

(1) The measuring seat, probe (sensor), extension rod, probe stylus, and standard ball must be installed reliably and firmly, without looseness and gaps. After checking whether the installed stylus and standard ball are firm, wipe the fingerprints and stains on the stylus and standard ball to keep them clean.

(2) The measurement speed during probe calibration should be the same as that during measurement, and automatic measurement methods should be adopted. When using the manual calibration method, attention should be paid to the uniformity of the measurement speed and position. Pay attention to whether there is a big deviation between the corrected stylus diameter and the correction result of the same length before correction, and the shape deviation during correction. If there are obvious changes, find the cause or clean the standard ball and stylus, repeat the correction of 2~3 times, and observe the degree of repetition of the results. Check whether the probe, stylus, and standard ball are firmly installed, and check the working status of the instrument.

(3) When multiple probe angles, positions or different probe lengths are required for calibration, the calibration effect (accuracy) must be checked after calibration. The method is: after calibrating all the defined probes, use the functions of the probe to measure the standard sphere with all the calibrated probes in turn, and observe the change of the center coordinates of the sphere. If there is a change of 1 to 2 meters, it is normal. If it changes a lot, it is necessary to check whether the installation of the base, probe, extension rod, stylus and standard ball is firm. It is an important reason for this phenomenon.

(4) Replace the probe (different software, different methods), because the length of the probe is an important parameter for the automatic calibration of the probe. If there is a deviation in the length of the input probe, it will cause abnormal collision of the probe.

(5) Input the standard ball diameter correctly. It can be seen from the principle of calibrating the probe that the diameter of the standard ball directly affects the calibration value of the diameter of the stylus gem ball. Although this is a "low probability event", it is still possible for beginners.

In summary, it is normal that the dynamic radius of the touch trigger probe is smaller than the static radius, so the surveyor does not need to worry about this. The calibration of the probe is an important part of the measurement process, and the deviation generated will be directly added to the measurement result. When using combined probes (multiple probe angles, multiple probe positions and multiple probe lengths), its calibration accuracy is particularly important, and we must attach great importance to it. When a problem is found, rechecking the results of the probe calibration will waste valuable time and increase a lot of work.

Pioneeris a professional probe factory, we provide touch probe, tool measuring probe, cnc machine tool probeand etc. Want to know more? Please contact us.



在线联系供应商

至: Harbin Pioneer M&E Technical Development Co.,Ltd
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

EP60 Electrical Probe
EP60 Electrical Probe EP60 Electric probe includes a electric probe unit and a taper shank. There is an unclosed power circuit inside the probe. When the stylus contacts...
pioneergroup-cn./*
pioneergroup-cn./* As an important component of the probe measurement system, the probe stylus/sensor is mainly composed of three parts: the stylus ball, the stylus r...
PIONEER CNC MACHINE PROBE & SETTER SYSTEM MANUFACTURER
PIONEER CNC MACHINE PROBE & SETTER SYSTEM MANUFACTURER Harbin Pioneer M&E Technical Development Co, Ltd was founded by Mr. Baichun Sun in 1995. After several years of development, the company gradua...
Radio Probe System
Radio Probe System The radio communication probe systemis suitable for various complex working conditions in the processing of CNC machine tools, mainly suitable for ...
TP300 Cable Probe
TP300 Cable Probe TP300 cable probe system mainly used in various specifications of CNC grinding machines, CNC lathes and specially measuring devices. TP300 cable pr...
供应产品

Same products

DM-16BF DM-18BF New cover Electric Sprayer
DM-16BF DM-18BF New cover Electric Sprayer 卖方: Zhejiang Huaqi Information Technology Co., LTD The Electric Knapsack Sprayer is a game-changer in the field of agricultural pest control. Engine...
Simple Round Breeding Cage
Simple Round Breeding Cage 卖方: Zhejiang Huaqi Information Technology Co., LTD The Simple Round Breeding Cage features a durable and lightweight construction, making it suitabl...
叙利亚-细胞储存液氮容器-液氮液位报警器
叙利亚-细胞储存液氮容器-液氮液位报警器 卖方: 河南天之道生物科技有限公司 精准的实验条件与安全的样本存储是科研工作顺利开展的关键要素。其中,细胞储存液氮罐作为实验室不可缺少的一种设备,在样本保存方面发挥着重要的作用。 实验室用的细胞储存液氮罐,航空铝制造,重量轻、占用...
意大利-液氮细胞储存罐-液氮罐液位报警器
意大利-液氮细胞储存罐-液氮罐液位报警器 卖方: 河南天驰仪器设备有限公司 精准的实验条件与安全的样本存储是科研工作顺利开展的关键要素。其中,细胞储存液氮罐作为实验室不可缺少的一种设备,在样本保存方面发挥着重要的作用。 实验室用的细胞储存液氮罐,航空铝制造,重量轻、占用...
埃及-细胞储存液氮罐-液氮液位监测仪
埃及-细胞储存液氮罐-液氮液位监测仪 卖方: Henan Tianzhidao Biological Technology Co., Ltd. 精准的实验条件与安全的样本存储是科研工作顺利开展的关键要素。其中,细胞储存液氮罐作为实验室不可缺少的一种设备,在样本保存方面发挥着重要的作用。 实验室用的细胞储存液氮罐,航空铝制造,重量轻、占用...