Photovoltaic module exposure test equipment
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
(1)GB/T 6495.3 Photovoltaic Devices – Part 3: Measurement Principles for Terrestrial Photovoltaic (PV) Solar Devices with Reference Spectral Irradiance Data
(2)GB/T 9535 Crystalline Silicon Terrestrial Photovoltaic (PV) Modules – Design Qualification and Type Approval
(3)IEC 61215 Terrestrial Photovoltaic (PV) Modules – Design Qualification and Type Approval
(4)IEC 60904-1 Photovoltaic Devices – Part 1: Measurement of Photovoltaic Current-Voltage Characteristics
(5)IEC 60904-2 Photovoltaic Devices – Part 2: Requirements for Reference Solar Devices
(6)IEC 60904-6 Photovoltaic Devices – Part 6: Requirements for Reference Solar Modules
Parameters
| Item | Technical Requirement |
|---|---|
| Solar simulator classification | Class B or better |
| Spectral distribution | Compliant with GB/T 6495.3 standard solar spectrum |
| STC irradiance | 1000 W/m² |
| STC cell temperature | 25 ℃ |
| NOCT irradiance | 800 W/m² |
| Temperature measurement accuracy | ±1 ℃ |
| Temperature measurement repeatability | ±0.5 ℃ |
| Test mode | Natural sunlight or solar simulator |
| Test parameters | Photovoltaic module I-V characteristics |
Features
Solar simulator compliant with IEC 60904 series standards, ensuring high result comparability
Spectral distribution strictly aligned with GB/T 6495.3 requirements for accurate measurement
Supports both STC and NOCT testing, covering key photovoltaic performance conditions
Test samples and reference devices positioned in the same plane perpendicular to incident light, reducing systematic measurement errors
High-precision temperature monitoring meets stringent photovoltaic standard requirements
Simple and flexible system configuration suitable for both laboratory and outdoor testing environments
Accessories
(1)Class B solar simulator compliant with IEC 60904-6
(2)Reference photovoltaic device compliant with IEC 60904-2 or IEC 60904-6
(3)Test sample mounting frame ensuring coplanar alignment with reference device
(4)Temperature measurement system with ±1 ℃ accuracy
(5)Irradiance monitoring device for real-time light intensity measurement
(6)Consumable temperature sensors for periodic calibration or replacement
(7)Consumable light source lamps replaced according to service life
Test Procedures
Calibrate the solar simulator and reference photovoltaic device before testing
Install the test module and reference device on the mounting frame, ensuring both are perpendicular to the incident light and in the same plane
Attach the temperature sensor firmly to the specified position on the back of the module
Set irradiance and temperature parameters according to the selected test condition
Avoid shading, reflection, or external light interference during testing
Measure and record the current–voltage (I-V) characteristics under defined conditions
在线联系供应商
Other supplier products
| Automated Filter Integrity Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Testing Standards ASTM F838 – Membrane bubble point method ASTM D5199 – Gas a... | |
| High Low Temperature Impact Testing Machine ASTM D256 | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application (1) Impact toughness determination of rigid plastics. (2) Impac... | |
| automotive seat belt abrasion tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application This benchtop-style tester allows operators to assess the decrease in webbing ... | |
| Automotive Seat Headrest Tester | Load Bearing & Safety Strength Test Equipment | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Features (1) Dedicated design for static strength testing of automotive head... | |
| Footwear Flexing Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The finished shoe resistance zigzag tester drives the zigzag part of the finished shoe to z... |