Breathing simulator manikin

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

GB/T 14296-2005 — Indoor Air Quality Evaluation Methods (Air Pollutant Exposure)

GB/T 18883-2022 — Indoor Air Quality Standard

ISO 7730:2005 — Ergonomics of the Thermal Environment — Analytical Determination and Interpretation of Thermal Comfort

ASTM F1291-2020 — Test Methods Related to Breathing and Exposure Measurement

Product Specifications

ComponentTechnical Parameters
Breathing Function Sinusoidal breathing profile; breathing frequency 10 breaths/min (one cycle 6.0 s: exhalation 2.0 s, inhalation 2.0 s, pause 2.0 s); customizable on request
Exhalation Volume Adjustable: 7 L/min, 15 L/min, 23 L/min; exhalation positive pressure 10–80 kPa (0.1–0.8 bar); inhalation negative pressure approx. −0.4 bar
Breathing Mode Switchable nasal breathing or oral breathing (when used with a thermal manikin)
Exhaled Airflow Equipped with heating module; airflow temperature adjustable from 34 to 36 °C
Sampling Branch Main breathing line includes a sampling port for aerosol or pollutant sampling
Electrical Control System Integrated control cabinet with pressure gauges and flow display; power supply AC 220 V / 50 Hz; total power ≤ 350 W; exhaled air heating module power ≤ 1000 W
Operating Environment Temperature 10–40 °C; humidity 15–95%
Others Can be used with a thermal manikin or independently with a manikin head; platform with lockable caster wheels; over-pressure and short-circuit protection; total weight approx. 45 kg

Features

Accurate simulation of human breathing cycles and airflow parameters, with support for customized breathing functions;

Adjustable exhalation temperature to ensure experimental conditions are close to real human breathing;

Compatible with a 16-zone thermal manikin to simulate interaction between body heat sources and breathing airflow;

Modular design allowing independent use or localized experiments on specific body parts;

High-sensitivity airflow control components with millisecond-level response, using imported components from Germany;

Comprehensive safety protection, including over-pressure protection, short-circuit protection, and lockable caster wheels.

Accessories

Accessory / ConsumableSpecification
Breathing Control Cabinet Includes PID control, pressure gauges, and flow display modules
Heating Module Exhaled air heating module with adjustable temperature from 34 to 36 °C
Main Breathing Line Equipped with sampling branch for aerosol or pollutant sampling devices
Quick-Connect Cables For connection to thermal manikin or head model, easy installation and removal
Customizable Modules Nasal or oral breathing interfaces; child or infant manikin compatible modules

Test Principle

The manikin breathing system simulates a sinusoidal human breathing cycle by controlling inhalation and exhalation flow rate, pressure, and temperature. When combined with a thermal manikin, the system generates thermal airflow patterns and air distribution similar to those of the human body. Under controlled experimental conditions, you can collect data on particle concentration, gaseous pollutant distribution, and airflow behavior. These data are used for breathing exposure assessment, airborne disease research, and indoor airflow field analysis.

Operation Instructions

Confirm that the ambient temperature is 10–40 °C and relative humidity is 15–95% before installation;

Ensure power supply is AC 220 V / 50 Hz, with total power ≤ 350 W and exhaled air heating module power ≤ 1000 W;

Connect the breathing system to the thermal manikin using quick-connect interfaces and check cable integrity;

Set breathing parameters such as frequency, volume, inhalation/exhalation timing, and breathing mode, with PID parameters automatically tuned;

Switch between nasal and oral breathing modes during testing as required by the exposure scenario;

When using the sampling branch, ensure aerosol sampling devices are properly connected to avoid air leakage;

After testing, switch off the system, save data, and disconnect all connections.

Test Procedure

Connect the breathing system to the thermal manikin or head model and check all cables and interfaces;

Power on the control system and set breathing parameters and airflow temperature;

Activate the heating module and wait until exhaled air temperature stabilizes;

Select nasal or oral breathing mode according to test requirements;

Start the experiment and record breathing pressure, flow rate, temperature, and particle distribution data in real time;

After completion, shut down the system, disconnect power, and clean sampling ports and pipelines.


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Five Finger Scratch Resistance Testing Machine
Five Finger Scratch Resistance Testing Machine Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Five-Finger Scratch Tester is a precision instrument designed to assess t...
Nail Bed Flammable Tester
Nail Bed Flammable Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The tester is used as test fixture in the flammability test of solid and soft toys. Place t...
Femoral Stem Neck Prostheses Fatigue Tester
Femoral Stem Neck Prostheses Fatigue Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0809.6 (China): Specifies neck fatigue testing for femoral stems in partial/...
Safety Helmets Impact and Penetration Resistance Tester
Safety Helmets Impact and Penetration Resistance Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Safety Helmets Impact and Penetration Resistance Tester is a dual-function testing syst...
Cargo Compartment Liner Burn-Through Fuel Test System,Aviation fuel fire penetration test system
Cargo Compartment Liner Burn-Through Fuel Test System,Aviation fuel fire penetration test system Whats/App: +86 156 1874 6768 Web: standard-groups.com The Cargo Compartment Liner Burn-Through Fuel Test System is a specialized aviation fire te...
供应产品

Same products

不锈钢焊剂芯焊丝 E308T1-1 E308HT1-1 E308LT1-1 用于304不锈钢焊接的气保焊丝
不锈钢焊剂芯焊丝 E308T1-1 E308HT1-1 E308LT1-1 用于304不锈钢焊接的气保焊丝 卖方: 863048 产品描述 不锈钢药芯焊丝 E308T1-1 E308HT1-1 E308LT1-1,FCAW不锈钢焊丝,AWS A5.22药芯焊丝,304不锈钢焊接耗材,低碳不锈钢药芯焊丝,耐高温不锈钢焊丝,...
E317LT1-1 焊剂芯线 CO2 不锈钢焊剂含量70% 直径1.0/1.2/1.6mm OEM定制 1200℃熔点
E317LT1-1 焊剂芯线 CO2 不锈钢焊剂含量70% 直径1.0/1.2/1.6mm OEM定制 1200℃熔点 卖方: 863048 迅卓不锈钢药心焊丝 E317LT1-1 分类: Сварочная проволока E317LT1-1 с флюсом, нержавеющая сталь, 70% содержан...
Mini Jewelry Laser Repair System – 60W to 400W Precision Welding
Mini Jewelry Laser Repair System – 60W to 400W Precision Welding 卖方: Beijing JCZ Technology Co., Ltd The Mini Jewelry Laser Welding Machineis an advanced mini laser welding system engineered for pre...
Portable Backpack Laser Cleaning Machine – Lightweight, Efficient Rust & Paint Removal
Portable Backpack Laser Cleaning Machine – Lightweight, Efficient Rust & Paint Removal 卖方: Beijing JCZ Technology Co., Ltd TheHCP-M SeriesBackpack Laser Cleaning Machinefeatures a compact pulsed fiber laser system, desig...
Suitcase Pulsed Laser Cleaning Machine | Ultra-Portable Handheld Fiber Laser Cleaner
Suitcase Pulsed Laser Cleaning Machine | Ultra-Portable Handheld Fiber Laser Cleaner 卖方: Beijing JCZ Technology Co., Ltd The Suitcase Pulsed Laser Cleaning Machineis an ultra-portable, high-precision handheld fiber las...