HEPA/ULPA filter scanning test bench
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Application
(1) HEPA and ULPA Filter Classification: Determine filter grade and performance.
(2) Standard-Compliant Testing: Conduct tests according to ISO 29463 and EN 1822 standards.
(3) Quality Control and Production Assurance: Integrate into filter production lines for real-time quality verification.
Standards
The system supports and complies with the following standards:
(1) GB/T 14295-2019 – Air filters.
(2) GB/T 13554-2020 – High-efficiency air filters.
(3) GB/T 6165-2021 – HEPA filter performance: efficiency and resistance.
(4) ISO 29463-4 – HEPA filter leakage testing.
(5) ISO 29463-5 – HEPA filter overall separation efficiency testing.
(6) EN 1822-4 – HEPA filter leakage testing.
(7) EN 1822-5 – HEPA filter overall separation efficiency testing.
(8) ISO 14644-3 – Cleanroom air filtration system testing.
(9) IEST-RP-CC034 – Standardized HEPA and ULPA filter testing.
Parameters
| Parameter | Unit | Value |
|---|---|---|
| Test Flow Rate | m³/h | 120 – 5000 |
| Differential Pressure Measurement | Pa | 1000 |
| Test Sample Dimensions | mm | Max 1830 × 1220 × 300 |
| Aerosol Operating Medium | – | DEHS, PSL |
| Gas/Air Supply | bar | 6 |
| Power Supply | – | 3 × 400 VAC, 100 A |
Accessories
| Accessory | Quantity |
|---|---|
| Filter Media Test Instrument | 1 |
| Filter Adapter | 1 |
| Test System Software | 1 |
Test Principle
Sample Installation: The filter is mounted in the filter holder, ensuring a secure fit.
Aerosol Generation & Flow: Aerosol (DEHS or PSL) is introduced at a controlled volume through the filter.
Leakage and Efficiency Measurement: Downstream particle concentration is measured via particle counters to calculate overall filtration efficiency and identify potential leaks.
Automated Data Collection: PC-controlled software records upstream and downstream particle counts and calculates efficiency and leakage locations.
在线联系供应商
Other supplier products
| Building material flame retardancy test furnace,building code compliance equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com Technical parameters: Gas used: methane flow rate (35 ± 0.5)L/min, its purity is gr... | |
| G506 Mask Automatic Filter Performance Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application G506 Mask Automatic Filter Performance Tester is used fortestingprofessional f... | |
| Spinal Internal Fixation Components System Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0961-2014 (China): Specifies static/fatigue testing protocols for spinal fix... | |
| SI80 Carbon Arc Aging Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications Material Science: Accelerated aging tests for polymers, coatings, and textile... | |
| Torch and Grit particle tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application You can use the apparatus in the following scenarios: Battery housing materia... |
Same products
| G506 Filter Performance Tester | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM F2100:Standard Specificat... | |
| Automated Filter Integrity Tester | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Testing Standards ASTM F838 – Memb... | |