Cone Calorimeter

A cone calorimeter is a fire test tool that is based on the amount of heat released during the combustion process directly related to the oxygen consumption in the combustion process. The amount of heat generated is directly proportional to the severity of the fire, such as the rate of fire growth. In order to obtain the flammability of the material, this is exposed to an external radiation heat source. Therefore, since this is a forced combustion test, the cone value is generally considered to reflect the second ignitability of the ignition.

Standard:

ISO5660,ASTM E 1354, ASTM E 1474, ASTM E 1740, ASTM F 1550, ASTM D 6113, NFPA 264, CAN ULC 135, BS 476

Scope of application:

The Cone Calorimeter is the most significant bench scale instrument in the field of fire testing.

Principle:

A sample is placed in a "cone" shaped radiant heater. Normally, the sample is exposed to 35 kW / m2 of external flux-forming heaters. However, for more refractories the heater is often increased to 50 kW / m2. Once sufficient pyrolysis products are produced, ignition occurs. Combustion products pass through a conical heater and through an instrument discharge tube. The measured / calculated values are usually important including, but not limited to, the maximum ignition time, the mass loss rate during combustion, the maximum amount of time and heat released during combustion, the total amount released during the test.

Heatcomplete system includes:

1. Conical heater - in a truncated cone, 230V, 5000W, with a heat output of 100 kW / m2

2. Test the sample device in horizontal or vertical.

3. 3 K-type thermocouple and 3 PID temperature controllers for temperature regulation

4. Opening and closing mechanism - to protect the sample area before testing to ensure that the initial quality measurement is stable, the operator can have additional time to test the system before testing. For flammable samples, if there is no opening and closing mechanism, it is easy to burn prematurely. This extra time is very important to the operator

5. Sample holder - sample size 100mm x 100mm, thickness of not more than 50mm, placed horizontally or vertically

6. Load cell - quality measurement by strain gage component test, the accuracy of up to 0.01g. Equipped with quick electronic weighing components, mechanical stop device can avoid damage caused by movement, given the stable results, to ensure that equipment longevity

7. Spark ignition - 10kV spark igniter, equipped with a safety cease-fire device. The igniter is automatically positioned by means of a lever connected to the closing mechanism

8. Exhaust system - made of stainless steel to extend life. (Flow rate adjustable from 0 g / s to 50 g / s with an accuracy of at least 0.1 g / s), and orifice flow testers (thermocouples and differential pressure transducers), including hoods, gas sample sockets and exhaust fans. Common tests are typically 24 l / s

9. Gas sampling - including particulate filter, freezing cold trap, pump, drying cylinder, flow controller

10. Oxygen analysis - paramagnetic oxygen analyzer, 0-25%, standard performance

11. Smoke fog - measured with a laser system using a photodiode, 0.5 mW helium-neon laser, primary and standby photodetectors. With positioning bracket and 0.3,0.8 neutral density filter for calibration

12. Heat flow meter - used to set the level of radiation on the sample surface

13. The burner calibration-calibration apparatus tested the heat release rate by using 99.5% pure methane

14. DATA ACQUISITION - An Agilent wheel data collector / converter with 3-hole card slot, 6 1/2 digit (22 bit) internal DMM for 120 single-ended measurements or 48 double-ended measurements. Scan rate up to 250 channels / s with 115k baud RS232 and PCI GPIB interface. All readings are recorded automatically, with a stable reading of 50000 in the memory

15. Cone Calorimeter Software - Operating languages include English, French, German, Spanish and Japanese. The user interface is a Windows operating system with easy-to-use button operations, standard Windows data entry, drop-down options, click-to-select, and switch

 
Feature:

1). Displays the instrument status.

2). Calibrate the instrument and store the calibration results.

3). Collect the test data.

4). Calculate the required parameters.

5) Display the results by the standard way.

6). Multiple tests are averaged.

7). Data entry to FDMS (Fire Data Management System) - Most European and US laboratories apply the system format, sharing data between different laboratory calorimeters.

8). the file output is in CSV (Comma Separated File) format, which can be quickly converted to a spreadsheet


Optional Configuration:

1, Carbon dioxide and carbon monoxide - available NDIR gas analyzer.

2, Hydrogen chloride - heating supply lines and gas analyzers.

3, Controlled gas accessories - for low-oxygen sample analysis.

4, the quality of the loss of calorimeter.

5, FTIR toxicity tester.

   
Optional Configuration:

1, Carbon dioxide and carbon monoxide - available NDIR gas analyzer.

2, Hydrogen chloride - heating supply lines and gas analyzers.

3, Controlled gas accessories - for low-oxygen sample analysis.

4, the quality of the loss of calorimeter.

5, FTIR toxicity tester.

6, the protection of door-to protect the operator does not inhale poisonous samples generated by the flue gas.The steel plate blocks the rear,the exhaust hood blocks the top,and the test door blocks the bottom, protecting the door against the last side.



在线联系供应商

至: 千实精密机电仪器检测设备有限公司
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Flammability Test Kit for Indoors
Flammability Test Kit for Indoors It is used for European standard indoor flammability test kits. Test the flammability of the upholstered seat material. Widely used in Europe and t...
Fabric thermal protection radiation performance tester
Fabric thermal protection radiation performance tester Standards:ISO 6942Product description:Thermal protection performance tester used for the determination heat radiation performance of the single or ...
Taber Linear Abraser
Taber Linear Abraser TheTABER Linear Abraser (Abrader) - Model 5750has been engineered to evaluate product’s abrasion resistance, scratch resistance (single or mu...
AATCC Viewing Board
AATCC Viewing Board Standards:American Association of Textile Chemists and Colorist (AATCC) approved test methods 88B/88C/124/128/143/178.Product Information:1. Utilis...
din abrasion resistance tester for rubber and plastic
din abrasion resistance tester for rubber and plastic DIN Abrasion tester, which is the original product & is designed to conform to the DIN 53516, DIN ISO 4649. This top quality and highly popular...
供应产品

Same products

Digital Construction
Digital Construction 卖方: FJ Dynamics Co., Ltd. We don’t do construction, but we are transforming construction. FJD machine control system ...
FJD WiFi Camera for Autosteering Kit
FJD WiFi Camera for Autosteering Kit 卖方: FJ Dynamics Co., Ltd. $299.00 Tax excluded. Taxes and Shipping calculated at checkout Quantity: 1 Compatible wi...
Nitinol Tube
Nitinol Tube 卖方: Baoji City Changsheng Titanium Co.,Ltd Nickel titanium alloy is a type of shape memory alloy, which is a special alloy that can automati...
Molybdenum Sheet & Molybdenum Foil
Molybdenum Sheet & Molybdenum Foil 卖方: Baoji City Changsheng Titanium Co.,Ltd In refractory metals, molybdenum sheet has good thermal conductivity, electrical conductivity and...
Extension Sets
Extension Sets 卖方: Shanghai Kohope Medical Devices Co., Ltd. Now, we provide various types of extension sets, including iv extension set with injection port, ...