Rubber low temperature brittleness machine - Impact & Durability

Product Introduction:

Plastic low temperature embrittlement testing machine is to determine the temperature of rubber and plastic when the sample is damaged by impact under specified conditions, that is, the brittleness temperature, which can make comparative identification of the performance of non-rigid plastics and other elastic materials under low temperature conditions, and the advantages and shortcomings of temperature and low temperature performance. Therefore, both in scientific research materials and production process control are indispensable. This instrument is according toGB/T 15256andGB/T5470standard design, the development of compressor refrigeration equipment. The equipment adopts the cascade compressor refrigeration technology, the heat balance principle and the circulation stirring method to achieve automatic uniform cooling and constant temperature of the sample. This equipment is divided into rubber low temperature brittleness testing machine and plastic low temperature brittleness testing machine two kinds.

Standards:

1. ISO812 Rubber, vulcanized or thermoplastic - Determination of low-temperature brittleness

2. GB/T 15256 Rubber, vulcanized or thermoplastic - Determination of low-temperature brittleness

3. JIS-K-6261-2006 Rubber, vulcanized or thermoplastic - Determination of low temperature properties

4. ASTM D2137 Rubber property-Brittleness point of flexible polymers and coated fabrics

Technical parameter:

1. Temperature control range: -80℃ (room temperature ≤25℃) (there are -60. -70. -80℃ three models for customers to choose)

2. Cooling speed:

0 ℃ —— 1.2 ℃ —— 20 ℃ / min

- 20 ℃ —— 1.0 ℃ to 30 ℃ / min

- 30 ℃ —— 0.7 ℃ to 60 ℃ / min

-40℃——-60℃ about 0.7℃/min

-60℃——-80℃ About 0.5℃/min

Note: The use environment is used at room temperature (23-25 ℃), and the effect of diffusing temperature of the compressor is maintained during work. The side door can be fully opened.

3. After constant temperature, the temperature fluctuation within 3 minutes of the test: < ±0.5℃

4. Overall size: 900×505×800mm (length × width × height)

5. The overall size of the studio: 220×170×70mm (length × width × height)

6. Cooling medium: ethanol or other non-freezing liquid

7. Mixing motor: 15W

8. Working power supply 220V——240V,50Hz,2kW


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

High Temperature Oxygen Index Tester
High Temperature Oxygen Index Tester Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The High-Temperature Oxygen Index (TOI) measuring instrument determines oxygen ind...
Nissan 4-Station Abrasion Tester
Nissan 4-Station Abrasion Tester Get a quote: Whats/App: +86 181 0160 0864 The Nissan 4-Station Abrasion Tester is a high-efficiency laboratory instrument designed for evaluating...
Safety Helmet Lateral Rigidity Tester
Safety Helmet Lateral Rigidity Tester Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The Safety Helmet Lateral Rigidity Tester is a precision instrument designed to ev...
Infusion Pump Flow Rate Tester
Infusion Pump Flow Rate Tester Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The Infusion Pump Flow Rate Tester is specifically designed for testing the flow r...
Gas Jet Erosion Rig/Jet Erosion Test Rig
Gas Jet Erosion Rig/Jet Erosion Test Rig Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The Gas Jet Erosion Rig is a laboratory testing system designed to evaluate erosio...
All supplier products

Same products

Briefly describe the importance, core functionality and how to implement a hospital clock system solution
Briefly describe the importance, core functionality and how to implement a hospital clock system solution Seller: California Triangle In the healthcare industry, precise time synchronization is critical. Every second can be a matte...
Sand and Dust Test Chamber for Automotive Connectors
Sand and Dust Test Chamber for Automotive Connectors Seller: Standard International Group (HK) Limited Operating Conditions: Ambient temperature: 5°C~+28°C (average temperature within 24 hour...
Automotive Connector SO₂/H₂S Corrosion Test Chamber
Automotive Connector SO₂/H₂S Corrosion Test Chamber Seller: Standard International Group (HK) Limited Standards: GB/T 10587-89. GB/T 2423.9-93: Chinese national standards for sulfur dioxide corrosio...
Automotive Connectors Salt Spray Tester
Automotive Connectors Salt Spray Tester Seller: Standard International Group (HK) Limited Applicable Standards: Neutral Salt Spray (NSS) Tests: GB/T2423.17. ISO 3768/3769/3770. ASTM B117...
Air - tightness Testing Device for Automotive Connectors
Air - tightness Testing Device for Automotive Connectors Seller: Standard International Group (HK) Limited Main Functions: 1. Provide stable positive and negative pressures for testing the air - tightnes...