EN 779 Turbine Filter Test System

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

The system complies with:

(1) ISO 29461-1 – Air filter performance, test methods.

(2) ISO 29461-2 – Air filter performance, gas adsorption test.

(3) ISO/AWI 29461-4 – Air filter performance, salt removal efficiency.

(4) ISO 16890-2 – Air filter classification and testing for general air filters.

(5) ISO 16890-3 – Gravimetric efficiency determination.

(6) ISO 16890-4 – Fractional efficiency and dust holding capacity.

(7) ISO/DIS 29461-7 – Gas turbine filter testing guidance.

(8) ISO 12103-1 – Fine test dust (A2) for filter performance.

(9) ASHRAE 52.2 – Air filter testing using airflow and particle size distribution.

(10) EN 779 – Requirements for high-efficiency air filters.

(11) GB/T 12103-1 & GB/T 12103-2 – Test dust and adsorption methods.

(12) GB/T 18801-2015 – Performance testing for air purifiers.

Principle

A defined volume flow rate is sucked through a HEPA intake filter, the attached test duct and the safety filter. The test system is composed of 6 sections: aerosol inlet, upstream measurement, specimen holder, downstream measurement, safety filter and flow rate measurement. The measurement sections are equipped with ports for the aerosol characterization and the determination of the pressure drop.

Technical data

ParameterUnitValue
Test flow rate m³/h 800 … 11.000
Test aerosol DEHS, KCl, Ashrae, ISO 12103-A2 (AC fine), ISO 12103-A4 (AC coarse)
Test specimen dimensions 610 × 610 mm (pocket and cassette filters); diameter ≤ 400 mm, length ≤ 1500 mm (filter cartridges)
Differential pressure of test specimen Pa max. 2000
Compressed air supply (operating gas/air) bar min. 5 (17 m³/h)
Water supply (operating medium) bar min. 3
Power supply 3 × 400 VAC, 125 A, 50 Hz
Air circulation % 0 … 100
Sensors (others) Temperature, relative humidity, air pressure
System dimensions (W × H × D) m 4.6 × 1.7 × 14.0

Features

Circulation mode for control of the test parameters temperature and humidity

Injection of different aerosols (NaCl, KCl, DEHS) into an optimal flow profile

Automated feeding of water droplets and supersaturated vapour

High level of automation, mostly automated test runs with complete operator guiding

Gas-Tight Test Ducts: Ensure accurate particle and gas measurements.

Modular Design: Accommodates different filter types and sizes.

High Automation: Software-driven sequences with automatic data acquisition.

Multi-Medium Testing: Supports various aerosols and pollutant gases.

Standardized Reporting: Automatically generates reports including differential pressure, dust loading, fractional efficiency, and salt removal efficiency.

Environmental Control: Built-in temperature and humidity sensors allow testing under controlled conditions.

Accessories

Accessory / ConsumableQuantityDescription
DEHS Aerosol Generator 1 For fine aerosol particle testing
ASHRAE Dust Generator 1 For standardized dust loading tests
Optical Particle Counters 2 0.2–10 μm, four channels
H14 Air Filter 1 For system calibration
Temperature & Humidity Sensors 2 Monitor test environment
Gas Analyzers 3 Measure NO₂, H₂SO₄, and other gases

Operation Instructions

Safety:

Confirm power, gas, and water connections before testing.

Avoid direct exposure to aerosols and gases; ensure good ventilation.

Wear protective equipment when handling hazardous gases.

Standard-Specific Testing:

The system adjusts test conditions automatically according to ISO 16890. ISO 29461-2. or other relevant standards.

After testing, the system generates standardized reports.

Test Procedure

Preparation:

Select the appropriate filter adapter and install the sample.

Configure test airflow, aerosol concentration, humidity, and temperature per standard requirements.

Testing:

Start the automated system; aerosol and gas injection begins.

The system continuously records differential pressure, dust loading, and fractional efficiency.

Data Analysis & Reporting:

After testing, the software analyzes collected data.

Generate reports including pressure curves, dust loading, fractional efficiency, and salt removal efficiency.

Maintenance Information

Regularly inspect aerosol and gas generators for proper operation.

Calibrate sensors periodically for accuracy.

Keep the system clean and free of particulate build-up.

Ensure all safety interlocks and ventilation systems are functional.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

ASTM ISO Vacuum Ball Disk Friction Tester Manufacturer
ASTM ISO Vacuum Ball Disk Friction Tester Manufacturer Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com A vacuum ball-on-disk friction tester is a device used to simulate friction a...
Syringe negative pressure tester device
Syringe negative pressure tester device Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The tester complies with the following domestic and international standards: (1...
ISO Sliding Bearing Fatigue Testing Machine Manufacturer China
ISO Sliding Bearing Fatigue Testing Machine Manufacturer China Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Materials Development: Evaluate the fatigue properties of new ma...
Photo fill light box
Photo fill light box Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard CE & FCC Certified: Meets electromagnetic compatibility and safety standards ...
Thermal - Cold Cycle Testing Machine
Thermal - Cold Cycle Testing Machine Whats/App: +86 156 1874 6768 Web: standard-groups.com The Thermal-Cold Cycle Testing Machine is a professional equipment used for high and low-te...
All supplier products

Same products

Filter Efficiency Testing Equipment
Filter Efficiency Testing Equipment Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com TheFilter Efficiency Testing Equipmentis ...
G506 Filter Performance Tester
G506 Filter Performance Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM F2100:Standard Specificat...
Automated Filter Integrity Tester
Automated Filter Integrity Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Testing Standards ASTM F838 – Memb...
High Efficiency Oil Water Separation Tester
High Efficiency Oil Water Separation Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO/TS 16332-2006 – Meth...
DIN 51524 hydraulic oil analyzer
DIN 51524 hydraulic oil analyzer Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards You can operate this unit in c...