Nonwovens liquid loss tester

Nonwovens liquid loss tester built-in loss of Taiwan and sample fixture for non-woven fabric materials for a specific test fluid slip and loss of performance.
Standards compliant:
Edana152.0-99; ISO9073-11
The main purpose:
Non-woven fabric liquid loss tester _ liquid slip and loss of the tester built-in loss of Taiwan and sample fixture for non-woven fabric materials for a specific test fluid slip and loss of performance.
The main parameters :
1, flow control range: 6.10mL / s ~ 6.41 mL / s;
2, slip and drain Taiwan angle control: 25 °;
3, slip and loss length: 250 ± 0.2mm;
4, timing: 60s, accuracy: ± 0.1s;
5, slip and loss Weight: 30g ± 0.01g.
Main features of thefabric liquid loss tester:
1, with the requirements of the requirements of the loss of Taiwan and sample fixture;
2, having a dose device capable of giving a 25 mL test solution in a continuous stream within (4 ± 0.1) s;
3, the test liquid out of time manual operation, automatic recording, digital display;
4, can be equipped with loss of weighing system.
Test principle:
A standard absorbent medium is placed under the nonwoven fabric to be measured and the combined sample is placed on a sloping pad to measure that when a particular amount of artificial urine flows down to the combined sample, the liquid penetrates the nonwoven Standard Absorbent Media Absorption, the liquid loss performance of the nonwovens sample was obtained by weighing the weight of the mass of the media prior to testing.
Other supplier products
|
|
Solid Auto - ignition Point Tester |
Product Introduction
This tester determines the self - ignition characteristics of solid chemical products by placing the substances in a 20mm - s... |
|
|
Catheter Power Injection Flow and Pressure Tester |
Main Uses
The catheter flow and pressure tester for power injection is an important medical device, mainly used to detect key parameters such as f... |
|
|
Fiber Crimp Elasticity Tester |
Product Introduction:
YG362C fiber crimp elasticity meter is used for crimp degree of 1-22dtex fiber.
Crimp elasticity, crimp recovery rate index... |
|
|
Respirator Fit meter |
PortaCount Respirator Fit Tester:
Qinsun assesses the overall integrity of masks by performing an aerosol leak test utilizing the PortaCount Respi... |
|
|
Luggage lifting and unloading testing machine |
Product Introduction:
The luggage simulation lifting and unloading tester is a measuring instrument for the life of the handle of the luggage carr... |
供应产品
Same products